510(k) Data Aggregation

    K Number
    K201543

    Validate with FDA (Live)

    Device Name
    MAGEC® System
    Date Cleared
    2020-07-30

    (51 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGEC® System is intended for skeletally immature patients less than 10 years of age with severe progressive spinal deformities (e.g., Cobb andle of 30 dearees or more: thoracic spine height less than 22 cm) associated with or at risk of Thoracic Insufficiency Syndrome (TIS). TIS is defined as the inability of the thorax to support normal respiration or lung growth.

    Device Description

    The MAGEC® System is comprised of a sterile single use spinal rod that can be surgically implanted using appropriate NuVasive® Reline®, Reline 4.5-5.0 (Reline Small Stature) or Armada® fixation components (i.e. pedicle screws, hooks and/or connectors). The implanted MAGEC Rod is used to brace the spine during growth to minimize the progression of scoliosis. The system includes a non-sterile hand held External Remote Controller (ERC) 1 or 2 that is used periodically after implantation to non-invasively distract the implanted spinal rod. The titanium MAGEC Rod includes an actuator portion that holds a small internal magnet. The magnet in the actuator can be turned non-invasively by use of the ERC, which is electrically powered. The hand held non-invasive ERC is placed over the patient's spine and then manually activated, which causes the implanted magnet to rotate and either distract or retract the rod. Once the physician determines that the implant has achieved its intended use and is no longer required, the implant is explanted. Additional accessories for the MAGEC® System include the MAGEC Manual Distractor, MAGEC Rod Template, and the MAGEC Wand Magnet Locator. The MAGEC Manual Distractor is a sterile, single use device, which is used in the operating room to test the MAGEC Rod prior to implantation. The MAGEC Rod Template is a sterile, single-use accessory used for intraoperative planning of the surgical procedure. The MAGEC Wand Magnet Locator is a non-sterile device which is used during the distraction procedure to locate the magnet within the MAGEC Rod previously implanted in a patient.

    The purpose of this premarket notification is to implement assembly process changes and a design change to the MAGEC Rod cleared in K171791 to address the Field Safety Notice and Product Recall issued on February 13, 2020 for the MAGEC X device (FDA Recall Number Z-1898-2020 and Recall Event ID 85495).

    AI/ML Overview

    The provided text is a 510(k) summary for the NuVasive MAGEC® System, which is a spinal rod system designed for skeletally immature patients with severe progressive spinal deformities. The submission aims to demonstrate substantial equivalence to previously cleared MAGEC systems.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative "acceptance criteria" for device performance in terms of specific thresholds for outcomes (e.g., minimum distraction force, maximum bending, etc.) that the device needed to meet. Instead, it refers to performance testing against established ASTM standards and internal verification for distraction/retraction. The general "acceptance criteria" can be inferred as demonstrating substantial equivalence to the predicate devices through successful completion of the listed nonclinical tests.

    Test DescriptionApplicable StandardReported Device Performance
    Static Compression BendingASTM F1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy ModelDemonstrated substantial equivalence to predicate devices
    Dynamic Compression BendingASTM F1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy ModelDemonstrated substantial equivalence to predicate devices
    Static TorsionASTM F1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy ModelDemonstrated substantial equivalence to predicate devices
    Distraction and Retraction VerificationN/ADemonstrated substantial equivalence to predicate devices (verified function)
    Torsional StrengthASTM F1798: Standard Guide For Evaluating The Static And Fatigue Properties Of Interconnection Mechanisms And Subassemblies Used In Spinal Arthrodesis ImplantsDemonstrated substantial equivalence to predicate devices

    2. Sample Size Used for the Test Set and Data Provenance

    The document only mentions "nonclinical testing" and lists standards for mechanical testing. It does not provide details on the sample size for these tests (e.g., number of rods tested for bending, torsion, etc.). The data provenance is also not specified beyond being "nonclinical testing." It does not involve human data, so terms like "country of origin" or "retrospective/prospective" are not applicable in this context.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable. The testing described is nonclinical mechanical testing, not clinical studies requiring expert review or ground truth establishment in a diagnostic context.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not applicable as the testing is nonclinical mechanical testing and does not involve human assessment or adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study was not done. The submission focuses on nonclinical testing for substantial equivalence of a medical device (spinal rod system), not on diagnostic performance or human reader improvement with AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This question is not applicable. The MAGEC System is a physical medical device (spinal rod and external controller), not an algorithm or AI system. Its performance is related to mechanical properties and functional reliability, not algorithmic accuracy.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The "ground truth" for the nonclinical tests is established by the specifications and acceptance limits defined within the referenced ASTM standards (e.g., specific forces, deflections, cycles, or distraction distances). The device is expected to meet or exceed the performance of the predicate and comply with these mechanical standards.

    8. The Sample Size for the Training Set

    This information is not applicable. The MAGEC System is a physical device, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for a physical medical device.