Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K201352

    Validate with FDA (Live)

    Device Name
    3D Auto LAA
    Manufacturer
    Date Cleared
    2020-08-17

    (88 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 3D Auto LAA is a software application designed to view and quantify 3D image data acquired by Philips Ultrasound Systems for use in measuring the area, circumference, and diameter of a Left Atrial Appendage (LAA) orifice.

    Device Description

    The purpose of this Traditional 510(k) Pre-market Notification is to introduce the new 3D Auto LAA cardiac quantification application for use on the Philips EPIQ and Affiniti Diagnostic Ultrasound systems, which were most recently cleared under K201012. The 3D Auto LAA application is compatible with 3D images generated via Philips Transesophageal Echocardiogram (TEE) transducers such as the X8-2t and X7-2t. The Philips 3D Auto LAA cardiac quantification application is a semi-automated application intended to provide measurements of the Left Atrial Appendage. The 3D Auto LAA application is designed to provide automatic and editable area, circumference, and diameter measurements of the Left Atrial Appendage (LAA) orifice. When the 3D Auto LAA application is launched, going through the workflow provides the clinician with a semi-automated preliminary border of the LAA in three planes based on greyscale intensity differentiation with some shape regularization.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Philips 3D Auto LAA device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly list specific acceptance criteria with numerical targets (e.g., accuracy > 90%) or directly reported performance metrics tied to such criteria. Instead, it states that the device was tested to "meet the defined requirements and performance claims." The "performance claims" are implicitly about the functionality of providing semi-automated measurements of LAA area, circumference, and diameter.

    Here's an interpretation based on the document's content, focusing on the functionality described:

    Acceptance Criterion (Inferred from functionality)Reported Device Performance (Inferred from "successful V&V testing" and description)
    Ability to view 3D image dataDevice successfully views 3D image data.
    Ability to quantify LAA orifice measurements (area, circumference, diameter)Device provides semi-automated and editable measurements of LAA area, circumference, and diameter.
    Semi-automated border detectionDevice provides a semi-automated preliminary border of the LAA in three planes.
    Measurable LAA AreaDevice generates LAA Area (mm²).
    Measurable LAA CircumferenceDevice generates LAA Circumference (mm).
    Measurable LAA Max DiameterDevice generates LAA Max Diameter (mm).
    Measurable LAA Minimum DiameterDevice generates LAA Minimum Diameter (mm).
    Editability of generated bordersUser can modify the entire contour or individual points.
    Compatibility with Philips Ultrasound SystemsCompatible with Philips EPIQ and Affiniti Diagnostic Ultrasound systems and Philips Transesophageal Echocardiogram (TEE) transducers.

    Important Note: The document focuses on demonstrating substantial equivalence to a predicate device by comparing intended use and technological characteristics, and by conducting internal V&V testing. It does not provide a detailed performance study with quantitative results against specific, pre-defined acceptance metrics.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify a numerical sample size for the "Performance Validation Study" or any other non-clinical V&V testing. It only mentions that testing was performed.
    • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. Given it's a non-clinical performance validation study and not a clinical trial, it's likely internal engineering data or synthetic data, but this is not explicitly stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide any information regarding the number or qualifications of experts used to establish ground truth for the non-clinical test set. As this was a "Performance Validation Study" for a software application for quantification, it's more likely that the "ground truth" was derived from predefined technical specifications, comparison to manual measurements by engineers, or comparison to the predicate device's output, rather than expert clinical consensus on images.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and The Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed or reported in this document. The document explicitly states: "The subject Philips 3D Auto LAA cardiac quantification did not require clinical data in order to make a determination for substantial equivalence when compared to the predicate device(s)." This indicates no human-in-the-loop performance study for regulatory submission purposes.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance assessment was done implicitly. The "Performance Validation Study" and "Software Verification and Validation testing" address the algorithm's performance in generating semi-automated borders and measurements. However, it's important to note the device is described as "semi-automated," implying a human-in-the-loop is part of the intended use (to refine the borders). The testing reported here focuses on the algorithm's capability to generate these preliminary borders and measurements.

    7. The Type of Ground Truth Used

    The document does not explicitly state the type of ground truth used for the non-clinical V&V testing. Given the nature of a software quantification tool, it's highly probable the ground truth involved:

    • Comparison against measurements obtained manually by trained individuals (engineers or potentially clinicians) using existing validated methods.
    • Comparison against the performance of the predicate device (QLAB System's 3DQ Q-App) manual tracing options, to ensure equivalence.
    • Internal gold standards derived from expert-generated manual segmentations on a subset of images used for development and testing.

    8. The Sample Size for the Training Set

    The document does not provide any information regarding the sample size for the training set. This information is typically not included in a 510(k) summary focused on substantial equivalence and non-clinical testing.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide any information on how the ground truth for the training set was established.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1