Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K201178

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-05-29

    (28 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IVUS modality of the iLab™ Polaris Multi-Modality Guidance System is intended for ultrasound examinations of intravascular pathology. Intravascular ultrasound is indicated in patients who are candidates for transluminal interventional procedures such as angioplasty and atherectomy.

    FFR and DFR™ are intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices.

    FFR and DFR are indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters.

    The Imaging Catheters generate ultrasound images and are intended for ultrasound examination of vascular and cardiac pathology. Boston Scientific manufactures a wide variety of catheters for different applications. The recommended use of each of these catheters may vary depending on the size and type of the catheter. Please refer to the Imaging Catheter Directions for Use, packaged with each catheter.

    Indications for Auto Pullback Use (IVUS Only)

    Automatic Pullback is indicated when the following occurs:

    • The physician/operator wants to standardize the method in which intravascular ultrasound images are obtained and documented: procedure-to-procedure, operator-to-operator.

    · The physician/operator wants to make linear determinations post-procedurally, which requires the imaging core of a catheter to be pulled back at a known uniform speed.

    · Two-dimensional, longitudinal reconstruction of the anatomy is desired.

    Device Description

    The iLab™ Polaris Multi-Modality Guidance System is a non-patient contacting, diagnostic device designed to provide three (3) primary modalities: IVUS (Intravascular Ultrasound), FFR (Fractional Flow Reserve) and Diastolic hyperemia-Free Ratio (DFR).

    The Polaris 2.12 (iLab 3.12) software update supports the iLab™ Polaris Multi-Modality Guidance System and supports existing device IVUS, FFR and DFR™ functionality as well as includes scaling updates to DFR™ equalization values, enhanced imaging modes and OS patches for additional security.

    Diastolic hyperemia-Free Ratio™ (DFR) is a resting index that measures multiple diastolic portions during the cardiac cycle. DFR™ calculates the diastolic portion of the cardiac cycle averaged over five beats, using two criteria for the measurement windows: 1) Pa less than mean Pa and 2) down sloping Pa values. No hyperemic agent is required for DFR™ calculation. The Polaris 2.12 software update adds an automatic scaling feature for DFR™ equalization.

    AI/ML Overview

    This looks like an FDA 510(k) clearance letter and summary for a software update to an existing medical device, the iLab Polaris Multi-Modality Guidance System. Based on the provided text, the submission focuses on non-clinical performance data for a software update, not a new AI/ML-based diagnostic algorithm requiring extensive clinical validation.

    Therefore, many of the requested points regarding the study that proves the device meets acceptance criteria (especially those related to clinical performance, ground truth, expert adjudication, and MRMC studies) are explicitly stated as "Not applicable" in the document.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria with corresponding performance metrics (e.g., sensitivity, specificity, accuracy) for a diagnostic AI/ML algorithm. Instead, it states that
    "Determination of substantial equivalence is based on an assessment of non-clinical performance data which includes software verification and validation carried out on Polaris 2.12 (iLab 3.12) software. Testing was conducted according to applicable international standards, FDA recognized consensus standards, and the same well-established test methods and criteria applied to the predicate device."

    Device Performance Reported: The core performance claim is "substantial equivalence" to the predicate device (iLab Polaris Multi-Modality Guidance System K191008). This is based on:

    • Non-clinical performance data: software verification and validation.
    • Compliance with standards: IEC 62304 and FDA Guidance for Premarket Submissions for Software Contained in Medical Devices.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: This information is not provided in the document. As the submission is based on non-clinical software verification and validation, it's unlikely to involve traditional clinical "test sets" with patient data in the same way an AI diagnostic algorithm would.
    • Data Provenance: This information is not provided and is likely not relevant given the non-clinical nature of the submission (software V&V).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. The submission explicitly states "Clinical Performance Data: Not applicable. A determination of Substantial Equivalence for this modification is not based on clinical data." Therefore, there was no ground truth derived from expert review of patient cases in the context of this 510(k).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not Applicable. As no clinical "test set" with expert review was used for performance evaluation, no adjudication method was employed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The document clearly states: "Clinical Performance Data: Not applicable." Therefore, no MRMC study was conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No, not in the context of a new diagnostic algorithm. The device is a "Multi-Modality Guidance System" with IVUS, FFR, and DFR functionalities, intended for use by clinicians. The software update (Polaris 2.12 / iLab 3.12) supports these existing functionalities and includes "scaling updates to DFR™ equalization values, enhanced imaging modes and OS patches for additional security." This is a software update to a guidance system, not a standalone diagnostic algorithm that makes a diagnosis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not Applicable. As no clinical performance data was submitted, there was no ground truth established from patient data. The "ground truth" for software verification and validation would be adherence to specified software requirements and functionality, confirmed through various engineering tests.

    8. The sample size for the training set

    • Not Applicable. This submission is for a software update to an existing device, not a de novo AI/ML algorithm that requires a "training set" of data.

    9. How the ground truth for the training set was established

    • Not Applicable. Since there was no "training set" of data for an AI/ML algorithm, no ground truth was established for it.

    In summary: This FDA 510(k) submission primarily relies on non-clinical software verification and validation to demonstrate substantial equivalence for a software update to an existing medical device. It does not introduce a new AI/ML diagnostic algorithm that would typically require the extensive clinical study data (including test sets, ground truth, expert review, and MRMC studies) you've inquired about.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1