Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K201140

    Validate with FDA (Live)

    Device Name
    Axeos
    Manufacturer
    Date Cleared
    2020-06-22

    (54 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The x-ray system creates data for digital exposures in the maxillofacial area and in subareas for dentistic dentistry, for hard-tissue diagnostics within ENT medicine, and carpus exposures.

    Device Description

    The proposed Axeos is a dental cone-beam CT system (CBCT), which comprises sensor units for 2D cephalometric exposures. 2D panoramic radiograph and 3D volume exposure. The combination of sensors varies depending on the installed device configuration. In the proposed device, the 2D sensor, as well as the 2D cephalometric sensor, are identical to the corresponding sensors utilized in the predicate device (K150217). However, with respect to the 3D CBCT exposures, the proposed Axeos device utilizes a new version of the flat panel sensor with a larger field of view compared to the flat panel version utilized within the predicate Orthophos SL (K150217).

    The proposed Axeos device uses an x-ray beam that rotates around the patient's head. Detectors acquire two-dimensional x-ray images at varying radiographic angles. An enhanced software algorithm generates the 2D panoramic and 2D cephalometric images and the reconstruction algorithm reconstructs the 3D volumetric image from the raw image data.

    The exposed area can be adapted to a specific region of interest to keep the radiation dose as low as possible for the patient. This is achieved by collimating the x-ray beam and the adjustment of starting and ending points of the sensor/x-ray source movement. Furthermore, the radiation dose can be adapted by various parameters such as program types and exposure technique factors. These functions are available for CBCT, cephalometric and panoramic exposures.

    A Class I laser beam is utilized to define reference lines for the correct patient position. The patient is stabilized through different bite blocks and the motor-driven forehead and temple supports.

    The obtained digital image data are processed to provide a reconstructed image to the operator/user. The reconstructed images are transferred to the currently marketed SIDEXIS 4 image processing software (K132773) and stored in the SIDEXIS 4 software database. The proposed Axeos device uses the same Sidexis 4 image processing software (K132773) as does the predicate device (K150217) and there are no changes to the SIDEXIS 4 software (K132773) introduced in this premarket notification.

    The proposed Axeos device includes metal artifact reduction software feature which automatically reduces image artifacts caused by radiopaque objects. This identical software feature is also included in the predicate Orthophos SL (K150217) device and remains unchanged in the proposed device.

    A user control panel allows user actions as: height adjustment, selection of programs, and exposure parameters and delivers information about the unit status.

    For 3D imaging, the proposed Axeos allows the user to select volume sizes within which multiple fields of view, anatomic positions, and collimation may be selected. This is intended to allow the clinician to select the field of view based on the diagnostic need and to minimize the dose exposed to the patient.

    The main components of the proposed Axeos device are:

    • . X-ray source
    • . X-ray detector (flat panel and/or PAN sensor)
    • Operator panel
    • . Laser locator
    • Cephalometric arm with detector (Ceph sensor)
    • Remote control (only by wire).
    • . Test phantoms: Exposure phantom, Constancy test phantom, Contrast element, and Ceph test phantom.
    AI/ML Overview

    The provided text describes the Dentsply Sirona Axeos, a dental cone-beam CT system, and its substantial equivalence to a predicate device (Orthophos SL, K150217). The document focuses on regulatory approval rather than clinical studies to establish diagnostic performance against specific acceptance criteria. Therefore, most of the requested information regarding acceptance criteria, specific study designs, sample sizes, expert involvement, and ground truth establishment cannot be extracted directly from this document.

    However, I can provide information on what is present in the document that relates to performance and equivalence.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not specify quantitative acceptance criteria for diagnostic performance outcomes (e.g., sensitivity, specificity, or image quality metrics for specific pathologies). Instead, it relies on demonstrating substantial equivalence to a predicate device through technological comparison and non-clinical performance testing.

    The "performance" reported primarily relates to technical specifications and the conclusion that the device's design outputs meet design input requirements and conform to intended user needs.

    CategoryAcceptance Criteria (Implicit from Equivalence)Reported Device Performance
    Intended UseSame as predicate device (Orthophos SL, K150217)Identical: "The x-ray system creates data for digital exposures in the maxillofacial area and in subareas for dentistry and pediatric dentistry, for hard-tissue diagnostics within ENT medicine, and and carpus exposures." (Page 6)
    Indications for UseSame as predicate device (Orthophos SL, K150217)Identical (Page 6)
    Functional Imaging Cap.Same as predicate deviceIdentical: "2D panoramic, 2D cephalometric, and 3D volumetric imaging" (Page 6)
    Operating PrinciplesSame as predicate deviceIdentical (Page 6)
    Patient FixationSame as predicate deviceIdentical (Page 6)
    WorkflowSame as predicate deviceIdentical (Page 6)
    3D Flat Panel Sensor SizeLarger than predicate (demonstrated as a difference, not a specific criterion)Proposed: 3D Hamamatsu 23x16 Flatpanel with active sensor area 230 mm x 160 mm. Predicate: 3D Hamamatsu 16x16 Flatpanel with active sensor area 160 mm x 160 mm. This results in a larger max FOV for the proposed device (17 cm diameter, 13 cm height vs. 11 cm diameter, 10 cm height). (Page 7)
    Image VisualizationMay differ from predicateProposed device does not include lingual-buccal exploration, while predicate does. (Page 6)
    Metal Artifact ReductionIdentical to predicateIdentical software feature also included in the predicate. (Page 5)
    IEC StandardsConformity to relevant IEC standardsTesting performed to verify conformity with various IEC standards including general safety, EMC, radiation protection, usability, and software lifecycle. (Page 11)
    Image QualitySubstantially equivalent to predicate device"Verification activities for confirmation of the image quality of the proposed device has been performed. The results of the image quality review have demonstrated that the device is substantially equivalent to the predicate device." (Page 11)

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not describe a clinical test set with a specific sample size. The performance evaluation was primarily non-clinical, focusing on technological characteristics and conformity to standards.

    Consequently, there is no information on:

    • Sample size used for a test set.
    • Country of origin of data.
    • Whether data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable, as no clinical test set requiring ground truth established by experts is described in this document.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set requiring adjudication is described.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document is for a medical imaging device (CT system), not an AI diagnostic algorithm, and thus no MRMC study or AI assistance is mentioned.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is a CT imaging system, not a standalone AI algorithm. It produces images for interpretation by a human.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    Not applicable. For the non-clinical performance, "ground truth" would be engineering specifications and measurements against physical phantoms (e.g., for resolution, dose, etc.), or compliance with regulatory standards. The document mentions "Test phantoms: Exposure phantom, Constancy test phantom, Contrast element, and Ceph test phantom" (Page 5) which are used for technical assessments.

    8. The Sample Size for the Training Set

    Not applicable. This document describes a medical imaging device, not a machine learning model that requires a training set of data.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for a machine learning model described here.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1