    Intended Use

    The Cure™ Lumbar Plate System is intended for use via a lateral or anterolateral surgical approach above the bifurcation of the great vessels in the treatment of thoracolumbar (T1-L5) spine instability or via the anterior surgical approach, below the bifurcation of the great vessels in the treatment of lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, lordosis, spinal stenosis, or a failed previous spine surgery. The device is intended as a temporary fixation device until fusion is achieved.

    Device Description

    The Cure™ Lumbar Plate (Cure™ OPEL-L (S)) is available in a range of sizes to coincide with the surgical approach. The Cure™ OPEL-L (S) is similar to the Cure™ Anterior Lumbar plate with the exceptions of replacing the graft window with a through hole, the inclusion of a central rib at the through hole location and convexity in the sagittal plane at the screw hole locations. All Cure™ Lumbar Plates continue to be manufactured from Grade 23 Titanium (Ti-6Al-4V ELI); manufactured according to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy (UNS R56401) for Surgical Implant Applications.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "Cure™ Lumbar Plate System." It details the device's characteristics, intended use, and a comparison to a predicate device to establish substantial equivalence.

    Based on the provided text, the device in question is a spinal implant (Cure™ Lumbar Plate System), not an AI-powered diagnostic or assistive tool. Therefore, the typical acceptance criteria and study designs associated with AI/ML medical devices (such as MRMC studies, standalone performance, training/test sets, expert adjudication of ground truth, etc.) are not applicable here.

    The document primarily discusses mechanical testing and an engineering analysis to demonstrate equivalence to a previously cleared predicate device.

    Here's how to address the request given the nature of the document:

    1. A table of acceptance criteria and the reported device performance:

    The document describes non-clinical mechanical testing rather than performance in a clinical setting against specific acceptance criteria for diagnostic accuracy. The acceptance criteria relate to mechanical properties and equivalence to the predicate device.

    Acceptance Criteria (Mechanical)Reported Device Performance
    Equivalency to predicate device (K171538) in:An engineering analysis was performed demonstrating equivalence. The Cure™ OPEL-L (S) Lumbar Plates were found to be superior in mechanical function and properties compared to the predicate device.
    - Static compression bending (per ASTM F1717)Met/Exceeded (Implied by superiority claim)
    - Static tension (per ASTM F1717)Met/Exceeded (Implied by superiority claim)
    - Dynamic compression bending (per ASTM F1717)Met/Exceeded (Implied by superiority claim)
    Material: Grade 23 Titanium (Ti-6Al-4V ELI)Confirmed: Manufactured from Grade 23 Titanium (Ti-6Al-4V ELI)
    Manufacturing standard: ASTM F136Confirmed: Manufactured according to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy (UNS R56401) for Surgical Implant Applications.
    Intended Use: (T1-L5 or L1-S1 instability due to fracture, tumor, DDD, scoliosis, lordosis, spinal stenosis, failed previous surgery; temporary fixation until fusion)Identical to Predicate Device: The identical intended use confirms that the device is suitable for the specified conditions, matching the predicate device. The document explicitly states: "The identical intended use and consistency between the fundamental scientific technology between the Cure™ OPEL-L (S) Lumbar Plate allows that it is substantially equivalent to the predicate device."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: Not applicable in the context of clinical "test sets" for AI/ML performance. The "testing" here refers to mechanical engineering tests on device prototypes/samples. The document does not specify the number of individual devices tested for mechanical properties, but it's standard engineering practice to test a representative sample.
    • Data Provenance: Not applicable. This involves lab-based mechanical testing, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. Ground truth for mechanical testing is established by engineering standards (e.g., ASTM F1717) and measurements, not expert clinical consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is relevant for clinical interpretation or subjective assessment, not mechanical properties.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a passive spinal implant, not an AI-assisted diagnostic or therapeutic tool. Therefore, no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Ground Truth: For mechanical testing, the "ground truth" is defined by the objective physical properties and performance measured against established engineering standards (e.g., ASTM F1717) and benchmarks from the predicate device.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.