510(k) Data Aggregation

    K Number
    K201099

    Validate with FDA (Live)

    Date Cleared
    2020-08-12

    (110 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD PhaSeal™ Optima system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The BD PhaSeal™ Optima system also prevents microbial ingress for up to 168 hours.

    Device Description

    BD PhaSeal™ Optima Closed System Drug Transfer Devices (CSTD) are sterile, single use closed system drug transfer devices intended for the reconstitution and transfer of antineoplastic or other hazardous drugs in the healthcare setting. The BD PhaSeal™ Optima system is comprised of four devices—Protector, Injector, Connector, and Infusion Adapter.

    The closed transfer of liquid drugs takes place through a double membrane utilizing self-sealing elastomeric membranes that are tightly fitted together through the collet-style fitting on each of the BD PhaSeal™ Optima system devices. During use, the single lumen cannula of the Injector perforates the double membranes for the transfer of liquids. When the cannula is retracted the membranes seal off the transfer of environmental contaminants into the system and/or escape of drug or vapor concentrations outside the system, thereby minimizing the individual and environmental exposure to drug vapor, aerosols, leaks and spills. The BD PhaSeal™ Optima system prevents microbial ingress for up to 168 hours. Performance of the self-sealing membrane has been substantiated up to 10 penetrations.

    Device labeling includes the following statement: "The ability to prevent microbial ingress for up to 7 days should not be interpreted as modifying, extending, or superseding a manufacturer's labeling recommendations for the storage and expiration dating of the drug vial. Refer to drug manufacturer's recommendations and USP compounding guidelines for shelf life and sterility information."

    AI/ML Overview

    The provided text describes the regulatory clearance for the BD PhaSeal™ Optima Closed System Drug Transfer Device-Injector (N40-O). It does not describe an AI/ML powered medical device. Therefore, many of the requested fields are not applicable.

    Here's the information that can be extracted from the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Performance Test)Reported Device Performance
    Attachment/detachment forcePASS
    FragmentationPASS
    System flow ratePASS

    The text states: "all predetermined acceptance criteria were met."

    2. Sample sized used for the test set and the data provenance

    The sample sizes for the performance tests are not specified in the provided document.
    The data provenance is not specified beyond the tests being conducted by the manufacturer, Becton, Dickinson and Company.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as the device is a mechanical Closed System Drug Transfer Device, not an AI/ML system requiring expert ground truth for interpretation.

    4. Adjudication method for the test set

    This is not applicable as the device is a mechanical Closed System Drug Transfer Device, not an AI/ML system or a diagnostic device requiring adjudication of interpretations. The performance tests appear to be quantitative measurements against defined criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a mechanical Closed System Drug Transfer Device and not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a mechanical Closed System Drug Transfer Device and does not have an algorithm component.

    7. The type of ground truth used

    For the performance tests (Attachment/detachment force, Fragmentation, System flow rate), the ground truth would be established by objective physical measurements against predefined engineering or regulatory specifications.

    8. The sample size for the training set

    This is not applicable as there is no training set for a mechanical device.

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set for a mechanical device.