    Intended Use

    The Disposable Endoscope Valves Set is a collection of several sterile units. It is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials. It includes an air/water valve, a suction valve, a biopsy valve and a water connector.

    • Disposable Air/Water Valves: This unit is intended to be fitted to an endoscope air/water channel to control the inflow of medical gases and water, whilst preventing back-flow.

    • Disposable Suction Valves: This unit is intended to be fitted to an endoscope suction channel to control the operations of suction, whilst preventing inflow of air.

    • Disposable Biopsy Valves: This unit is intended to be fitted to an endoscope biopsy port to prevent leakage of gases and body fluids during an endoscopic procedure.

    • Disposable Endoscope Water Connectors: This unit is intended to provide irrigation via sterile water supply during GI endoscopic procedures when used in conjunction with an irrigation pump.

    The Disposable Endoscope Valves Set B is a collection of sterile device intended to be fitted to multiple endoscope working channels/ports to enable an endoscope operator control the working channels/ports and prevent retrograde flow of fluids, gases, and other materials. It includes a suction valve. This is a single-use device.

    • Air/Water Valves: This device is intended to be fitted to an endoscope air/water channel to enable the operator to control inflow of medical gases and water, whilst preventing backflow.

    • Suction Valves: The device is intended to be fitted to an endoscope suction channel to enable the operator to control suction whilst preventing inflow of air.

    Device Description

    The Disposable Endoscope Valves Set is a collection of several sterile units. It is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials. It includes an air/water valve, a suction valve, a biopsy valve and a water connector.

    The Disposable Endoscope Valves Set B is a collection of sterile device intended to be fitted to multiple endoscope working channels/ports of Pentax GI Video Endoscopes to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of fluids, gases, and other materials. It includes a suction valve, an air/water valve.

    AI/ML Overview

    The provided text describes the regulatory clearance for "Disposable Endoscope Valves Set" and "Disposable Endoscope Valves Set B" and compares them to predicate devices, but does not contain information related to software or AI performance, or any studies involving human readers or ground truth determination in the context of diagnostic accuracy.

    Therefore, I cannot fulfill the request for information regarding AI acceptance criteria, device performance tables, sample sizes, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set details as these concepts are not addressed in the provided document.

    The document focuses on the substantial equivalence of the physical medical device to legally marketed predicate devices, primarily based on:

    • Indications for use: Similar intended purposes.
    • Technological characteristics: Similar design, sterilization method, single-use nature, and environment of use. Differences in compatible endoscopes and materials are discussed but deemed not to alter suitability, supported by performance testing and biocompatibility.
    • Performance data: Bench testing to demonstrate functionality and biocompatibility testing.

    However, I can extract information related to the general performance testing conducted:

    1. A table of acceptance criteria and the reported device performance:

    The document lists performance tests but does not explicitly provide a table of acceptance criteria with corresponding reported performance values in a structured quantitative comparison format. Instead, it states that "All evaluation acceptance criteria were met" for biocompatibility and that "bench testing to support substantial equivalence" was performed for various functional aspects.

    Device ComponentTest PerformedAcceptance Criteria (Implied: Met)Device Performance (Details not provided beyond "met acceptance criteria")
    All componentsBiocompatibility (Cytotoxicity, Irritation, Sensitization)In accordance with ISO 10993-1Met
    Disposable Air/Water Valves WS-VO-01Time of inflation of 8KPa airNot explicitly statedMet (supporting substantial equivalence)
    Time of injection of 10g waterNot explicitly statedMet (supporting substantial equivalence)
    Disposable Suction Valves WS-VO-02Time of suction of 200g waterNot explicitly statedMet (supporting substantial equivalence)
    Disposable Biopsy Valves WS-M-01SPlug-in performance: Depression force, pull out forceNot explicitly statedMet (supporting substantial equivalence)
    Sealing performance: Leakage test, flush open forceNot explicitly statedMet (supporting substantial equivalence)
    Fitting testNot explicitly statedMet (supporting substantial equivalence)
    Disposable Endoscope Water connectors WS-VO-03Leakage test under 80Kpa forward pressureNot explicitly statedMet (supporting substantial equivalence)
    Leakage test under 30Kpa reverse pressureNot explicitly statedMet (supporting substantial equivalence)
    Disposable Air/Water Valves WS-VP-01Amount of water feedingNot explicitly statedMet (supporting substantial equivalence)
    Amount of air feedingNot explicitly statedMet (supporting substantial equivalence)
    Depression forceNot explicitly statedMet (supporting substantial equivalence)
    Sealing performanceNot explicitly statedMet (supporting substantial equivalence)
    Disposable Suction Valves WS-VP-02Amount of suctionNot explicitly statedMet (supporting substantial equivalence)
    Depression forceNot explicitly statedMet (supporting substantial equivalence)
    All componentsSterilization validation (ISO11135)In accordance with ISO 11135Validated
    All componentsShelf-life (accelerated aging test)3-year shelf-lifeEvaluated (and presumably met, as it supports substantial equivalence)

    The remaining points (2-9) cannot be answered. The document does not describe a study involving algorithms, AI, human readers, or diagnostic accuracy. It's a clearance document for physical medical devices (endoscope valves) based on substantial equivalence to existing devices.