510(k) Data Aggregation
(143 days)
EyeArt is indicated for use by healthcare providers to automatically detect more than mild diabetic retinopathy and visionthreatening diabetic retinopathy (severe non-proliferative diabetic retinopathy or proliferative diabetic retinopathy and/or diabetic macular edema) in eyes of adults diagnosed with diabetes who have not been previously diagnosed with more than mild diabetic retinopathy. EyeArt is indicated for use with Canon CR-2 Plus AF cameras in both primary care and eye care settings.
EyeArt is a software as a medical device that consists of several components - Client, Server, and Analysis Computation Engine. A retinal fundus camera, used to capture retinal fundus images of the patient, is connected to a computer where the EyeArt Client software is installed. The EyeArt Client software provides a graphical user interface (GUI) that allows the EyeArt operator to transfer the appropriate fundus images to and receive results from the remote EyeArt Analysis Computation Engine through the EyeArt Server. The EyeArt Analysis Computation Engine is installed on remote computer(s) in a secure data center and uses artificial intelligence algorithms to analyze the fundus images and return results. EyeArt is intended to be used with color fundus images of resolution 1.69 megapixels or higher captured using one of the indicated color fundus cameras (Canon CR-2 AF and Canon CR-2 Plus AF) with 45 degrees field of view. EyeArt is specified for use with two color fundus images per eye: optic nerve head (ONH) centered and macula centered.
For each patient eye, the EyeArt results separately indicate whether more than mild diabetic retinopathy (mtmDR) and vision-threatening diabetic retinopathy (vtDR) are detected. More than mild diabetic retinopathy is defined as the presence of moderate non-proliferative diabetic retinopathy or worse on the International Clinical Diabetic Retinopathy (ICDR) severity scale and/or the presence of diabetic macular edema. Vision-threatening diabetic retinopathy is defined as the presence of severe non-proliferative diabetic retinopathy or proliferative diabetic retinopathy on the ICDR severity scale and/or the presence of diabetic macular edema.
Here's a breakdown of the EyeArt device's acceptance criteria and the study that proves it meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
Device: EyeArt (v2.1.0)
Indication for Use: To automatically detect more than mild diabetic retinopathy (mtmDR) and vision-threatening diabetic retinopathy (vtDR) in adults diagnosed with diabetes who have not been previously diagnosed with more than mild diabetic retinopathy.
| Metric | Acceptance Criteria (Implicit by achieving high performance) | Reported Device Performance (Worst Case Across Cohorts/Outcomes) |
|---|---|---|
| Sensitivity (mtmDR) | High (e.g., above 90%) | 92.9% (Enrichment-permitted, Primary Care) |
| Specificity (mtmDR) | High (e.g., above 85%) | 85.2% (Enrichment-permitted, Ophthalmology) |
| Sensitivity (vtDR) | High (e.g., near 90% or 100% for smaller groups) | 88.9% (Sequential, Ophthalmology) |
| Specificity (vtDR) | High (e.g., above 89%) | 89.8% (Enrichment-permitted, Ophthalmology) |
| Imageability | High (e.g., above 95%) | 96.5% (Enrichment-permitted, Ophthalmology, Sequence P1/P2/P3) |
| Intra-operator Repeatability (OA - mtmDR) | High (e.g., above 90%) | 93.5% (Cohort P2) |
| Intra-operator Repeatability (OA - vtDR) | High (e.g., above 96%) | 96.8% (Cohort P2) |
| Inter-operator Reproducibility (OA - mtmDR) | High (e.g., above 90%) | 90.3% (Cohort P1) |
| Inter-operator Reproducibility (OA - vtDR) | High (e.g., above 96%) | 96.8% (Cohort P1) |
Note: The document does not explicitly state numerical acceptance criteria thresholds. The reported performance values are the actual outcomes from the clinical study, which are implicitly considered acceptable for substantial equivalence based on the FDA's clearance. The "worst case" reported here refers to the lowest performance observed across the different cohorts for each metric. Confidence intervals are provided in the tables and are generally tight, indicating reliability.
Study Details Proving Device Meets Acceptance Criteria
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Size:
- Clinical Efficacy Study: 655 participants (after exclusions from an initial 942 screened), comprising 1290 eyes (assuming 2 eyes per subject, though analyses are eye-level). These 655 participants were divided into two main cohorts:
- Sequential Enrollment: 235 subjects (45 in primary care, 190 in ophthalmology)
- Enrichment-Permitted Enrollment: 420 subjects (335 in primary care, 85 in ophthalmology)
- Precision (Repeatability/Reproducibility) Study: 62 subjects (31 subjects each at 2 US primary care sites), resulting in 186 pairs of images for repeatability analysis (Cohort P1) and 62 subjects for Cohort P2 (3 repeats each).
- Clinical Efficacy Study: 655 participants (after exclusions from an initial 942 screened), comprising 1290 eyes (assuming 2 eyes per subject, though analyses are eye-level). These 655 participants were divided into two main cohorts:
- Data Provenance: Prospective, multi-center pivotal clinical trial conducted across 11 US study sites (primary care centers and general ophthalmology centers).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: At least 2 independent graders and an additional adjudication grader (more experienced) were used for each subject's images.
- Qualifications of Experts: Experienced and certified graders at the Fundus Photography Reading Center (FPRC). They were certified to grade according to the Early Treatment for Diabetic Retinopathy Study severity (ETDRS) scale. Specific experience levels (e.g., "10 years of experience") are not detailed beyond "experienced and certified."
4. Adjudication Method for the Test Set
- Method: 2+1 adjudication. Each subject's images were independently graded by 2 experienced and certified graders. In case of significant differences (determined using prespecified significance levels) between the two independent gradings, a more experienced adjudication grader graded the same images to establish the final ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was it done? No, a comparative effectiveness study evaluating how much human readers improve with AI vs. without AI assistance was not reported in this document. The study primarily evaluated the standalone performance of the EyeArt device against expert human grading (FPRC reference standard).
6. Standalone (i.e., algorithm only without human-in-the-loop performance)
- Was it done? Yes, the entire clinical testing section (Section D) describes the standalone performance of the EyeArt algorithm. The EyeArt results (positive, negative, or ungradable for mtmDR and vtDR) for each eye were compared directly to the clinical reference standard established by FPRC grading.
7. Type of Ground Truth Used
- Type: Expert Consensus Grading (adjudicated) from the Fundus Photography Reading Center (FPRC). This grading was based on dilated 4-wide field stereo fundus imaging and applied the Early Treatment for Diabetic Retinopathy Study (ETDRS) severity scale.
- mtmDR Ground Truth: Positive if ETDRS level was 35 or greater (but not equal to 90) or clinically significant macular edema (CSME) grade was CSME present. Negative if ETDRS levels were 10-20 and CSME grade was CSME absent.
- vtDR Ground Truth: Positive if ETDRS level was 53 or greater (but not equal to 90) or CSME grade was CSME present. Negative if ETDRS levels were 10-47 and CSME grade was CSME absent.
8. Sample Size for the Training Set
- The document does not specify the sample size for the training set. It mentions that the EyeArt Analysis Computation Engine uses "an ensemble of clinically aligned machine learning (deep learning) algorithms" but provides no details on their training data.
9. How the Ground Truth for the Training Set Was Established
- The document does not specify how the ground truth for the training set was established. While the "clinically aligned framework" is mentioned, the specific methodology for annotating or establishing ground truth for the training data is not detailed in this submission summary.