510(k) Data Aggregation

    K Number
    K200622

    Validate with FDA (Live)

    Date Cleared
    2020-04-01

    (23 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Focus 43C and TRIMAX 43C are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. These two devices are not intended for mammography, dental applications.

    Device Description

    Focus 43C Detector and TRIMAX 43C Detector (Hereinafter referred to as Focus 43C and TRIMAX 43C) are a kind of wireless digital flat panel detectors. They support the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 427mm x 427mm. Two models TRIMAX 43C and Focus 43C are totally same except for label and trademark. The sensor plate of Focus 43C and TRIMAX 43C detectors are direct-deposited with CsI scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface. The major function of the Focus 43C and TRIMAX 43C detectors are to convert the X-ray to digital image, with the application of high-resolution X-ray imaging. This kind of detector is the key component of DR system, enables to complete the digitalization of the medical X-ray imaging with the DR system software. iRay SDK(include iDetector) is intend to supply API interface for DR system manufacturers.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Focus 43C Detector" and "TRIMAX 43C Detector". The submission focuses on demonstrating substantial equivalence to a predicate device, the "Wireless Digital Flat Panel Detector (Mars1717XF-CSI, K183713)".

    Here's an analysis of the acceptance criteria and the study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a quantitative, pass/fail manner. Instead, it compares the technical characteristics of the proposed device to the predicate device to demonstrate substantial equivalence. The performance metrics are generally presented as improvements or minor changes compared to the predicate.

    CharacteristicPredicate Device (Mars1717XF-CSI)Proposed Device (Focus 43C / TRIMAX 43C)Evaluation/Comment (Implied Acceptance)
    Intended UseSame as proposedSame as predicateSubstantially equivalent. Indications for use explicitly include adult and pediatric patients.
    Classification NameStationary x-ray systemSameSubstantially equivalent.
    Product CodeMQBSameSubstantially equivalent.
    Regulation Number21 CFR 892.1680SameSubstantially equivalent.
    X-Ray AbsorberCsISameSubstantially equivalent.
    Installation TypeWireless, PortableSameSubstantially equivalent.
    Readout MechanismThin Film TransistorSameSubstantially equivalent.
    Image Matrix Size2832 × 2836 pixels3072 × 3072 pixelsImproved (higher resolution), considered substantially equivalent as improvement.
    Pixel Pitch150 µm139 µmImproved (smaller pixel size), considered substantially equivalent as improvement.
    ADC Digitization16 bitSameSubstantially equivalent.
    Effective Imaging Area424.8 mm × 425.4 mm427 mm × 427 mmSlightly larger, considered substantially equivalent.
    Spatial Resolution3.3 lp/mm3.6 lp/mmImproved (higher resolution), considered substantially equivalent as improvement.
    Modulation Transfer Function (MTF) @ 1 lp/mm0.490.61Improved, considered substantially equivalent as improvement.
    Detective Quantum Efficiency (DQE) @ 1 lp/mm (RQA5, 2.5µGy)0.400.46Improved, considered substantially equivalent as improvement.
    Power ConsumptionMax. 20WMax. 20WSubstantially equivalent.
    Communications (Wireless functionality)IEEE 802.11a/b/g/n (2.4 GHz / 5 GHz)IEEE 802.11a/b/g/n/ac (2.4 GHz / 5 GHz) + WiredImproved wireless standard (added 'ac') and wired option, considered substantially equivalent as an enhancement.
    Imaging Protect PlateN/A (not specified but presumed)Carbon Fiber PlateSpecified, likely similar or improved over predicate.
    CoolingAir coolingSameSubstantially equivalent.
    Dimensions460 mm × 460 mm × 15 mm460 mm × 460 mm × 15.2 mmMinor dimensional change, considered substantially equivalent.
    Operation Temperature+5 ~ +30 °C+5 ~ +35 °CSlightly wider acceptable range, considered substantially equivalent.
    Operation Humidity10 ~ 80% (Non-Condensing)10 ~ 90% (Non-Condensing)Slightly wider acceptable range, considered substantially equivalent.
    Storage Temperature-20 ~ +50 °C-20 ~ +55 °CSlightly wider acceptable range, considered substantially equivalent.
    Storage Humidity10 ~ 90% (Non-Condensing)5 ~ 95% (Non-Condensing)Slightly wider acceptable range, considered substantially equivalent.
    Software (iRay SDK)"intend to supply API interface for DR system manufacturers... unchanged from the predicate device."SameSubstantially equivalent. No change to the software API provided to DR system manufacturers.

    The overarching "acceptance criterion" for a 510(k) submission is demonstrating substantial equivalence to a predicate device in terms of intended use, technological characteristics, safety, and effectiveness. The document asserts that the proposed devices meet this through direct comparison and noting improvements where applicable.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" in the context of clinical images or patient data for evaluating the device's diagnostic performance. The evaluation is primarily focused on non-clinical performance testing and comparison of technical specifications to the predicate device.

    • No specific sample size for a test set of images or patient data is mentioned.
    • Data provenance (country of origin, retrospective/prospective) is not applicable or provided for such a test set.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to the provided document. The submission does not describe a study involving expert readers establishing ground truth for a diagnostic test set. The evaluation is based on technical specifications and engineering testing.

    4. Adjudication Method for the Test Set

    This information is not applicable as no diagnostic test set or adjudication process is described in the provided text.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study is mentioned. This device is a digital X-ray detector, not an AI-powered diagnostic tool, so such a study would not typically be performed or described in this type of submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not an AI algorithm. The device is a digital X-ray detector. Therefore, a standalone algorithm performance study is not applicable.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" derived from expert consensus, pathology, or outcomes data is not relevant to this submission, as it focuses on the technical performance specifications of the X-ray detector rather than its diagnostic accuracy in interpreting images. The "ground truth" for the device's performance characteristics (e.g., spatial resolution, MTF, DQE) would be established through physical measurements and standardized test phantoms according to industry standards.

    8. The Sample Size for the Training Set

    No training set is mentioned as this device is an X-ray detector, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set is mentioned or relevant for this device.

    Summary of Device Performance Demonstration:

    The submission demonstrates substantial equivalence primarily through:

    • Comparison of Technical Specifications: A detailed table comprehensively compares various technical characteristics, showing that the proposed device is either similar to or improved upon the predicate device (e.g., higher image matrix size, smaller pixel pitch, better spatial resolution, MTF, and DQE).
    • Non-Clinical Performance Testing:
      • Electrical Safety and EMC testing: Performed according to IEC/ES 60601-1 and IEC 60601-1-2, with results meeting standard requirements.
      • Biological Evaluation: Materials contacting skin evaluated per ISO 10993-1, with results assuring safety "the same as the predicate device."
      • Other Non-Clinical Considerations: These studies showed that non-clinical considerations mentioned in the "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices" are substantially equivalent to the predicate. This likely includes aspects like image quality, dose performance, and environmental testing.
    • Clinical Consideration: The document explicitly states that "clinical consideration may not necessary for changes in the pixel size and resolution with the same x-ray detection material and may not necessary for changes in the wireless functionality if non-clinical information is sufficient to support the substantial equivalence." It also states, "There was no significant difference between the images of the Focus 43C/TRIMAX 43C and those of the predicate device." This implies that while no formal clinical study is presented, the technical improvements are not expected to negatively impact clinical performance, and the output images are comparable.