Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K200564

    Validate with FDA (Live)

    Device Name
    Aria System
    Manufacturer
    Date Cleared
    2020-04-03

    (30 days)

    Product Code
    Regulation Number
    870.5800
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aria sequential circulator is a programmable sequential, pneumatic compression device intended for use by medical professionals and patients at home, for the treatment of the following conditions:

    • Chronic edema
    • Lymphedema
    • Venous insufficiency
    • Wound healing
    Device Description

    The Aria system consists of two main components: A flow generator and a garment. The garment is to be wrapped around the limb, providing a comfortable fit. The garment has seven (7) chambers that are filled with air by the flow generator to provide compression on the extremity. The Aria system uses a compressand-release massage action, similar to the predicate, in order to stimulate lymphatic vessels in the treated area and encourage fluid clearance.

    The Aria system retains similar hardware and performance features of the predicate device. Key features include flow generator, valves, A/C plug pack, lower limb garment, tubing, no-LCD User Interface and ON/OFF button. The Aria System contains a microprocessor-controlled flow generator/blower system that generates pressure from 0-45 mmHG to provide for effective treatment of the conditions described in the IFU.

    The Aria flow generator has no control settings and delivers one pre-programmed therapy mode.

    AI/ML Overview

    This document describes the Inova Labs Aria System, a programmable sequential, pneumatic compression device. It is intended for the treatment of chronic edema, lymphedema, venous insufficiency, and wound healing. The submission is a 510(k) premarket notification, indicating a claim of substantial equivalence to a predicate device.

    Here's the breakdown of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" against which the device performance is reported with specific numerical targets. Instead, it describes "bench testing" which included performance comparisons to a predicate device. The results of this testing are stated to "demonstrate that the Aria system raises no new safety or effectiveness concerns and is substantially equivalent to the predicate device."

    Here's a summary of the characteristics compared with the predicate device (Entre Model PD08-U, K143185):

    CharacteristicPredicate Device (Entre Model PD08-U, K143185)New Device (Aria System)Performance Statement (as implied from "Substantial Equivalence")
    Intended UseTreatment of Chronic edema, Lymphedema, Venous insufficiency, Wound healingTreatment of Chronic edema, Lymphedema, Venous insufficiency, Wound healingSame intended use
    Pressure Range0-45 mmHG (at 'moderate' setting)0-45 mmHGSimilar pressure range
    Cycle Time65 seconds56 secondsSimilar cycle time (minor difference)
    Total Therapy TimeApprox. 60 minutesApprox. 60 minutesSame total therapy time
    Modes of OperationSequential gradient compression therapySequential gradient compression therapySame mode of operation
    System ComponentsFlow generator, Valves, Tubing, GarmentFlow generator, Valves, Tubing, GarmentSimilar system components
    Flow Generator Operating SystemMicrocontrollerMicrocontrollerSame operating system
    Garment Air Chambers87Similar (minor difference, 7 vs 8)
    Tubing Length2m length1.8m lengthSimilar (minor difference)
    User InterfaceOn/Off Button, Pressure Low/Med/High Button, Start Therapy/PauseOn/Off (Start/Stop Therapy) ButtonSimplified UI on Aria System
    ConnectivityNoneBluetooth classic to allow export of system data to a paired appNew feature on Aria System
    Motor TypeCompressorBrush-less low voltage DCDifferent motor type
    Power Supply100-240V, 50-60Hz100-240V, 50-60HzSame power supply
    Weight2.5lb0.65lbSignificantly lighter Aria System
    Dimensions H x W x D (inches)11 x 6 x 8Flow generator unit: 2 x 3.3 x 5.3Smaller Aria System
    Tested Standards (electrical safety, EMC, usability, home medical, biocompatibility)IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, (ISO 10993-1 not referenced in 510(k) summary)IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-6, ISO 10993-1Both adhere to relevant standards, Aria includes ISO 10993-1

    The "bench testing" areas mentioned for the Aria System were:

    • Pressure stability
    • Sleeve burst test
    • Sleeve leakage test
    • Sleeve integrity test
    • Pressure accuracy test
    • Chamber filling cycle time testing
    • Total therapy time

    The report generally states that the Aria system "met the Aria System Specification when compared to the predicate device" for these tests, implying that the performance was either equivalent or acceptable relative to the predicate. Specific numerical values for acceptance criteria or detailed results are not provided in this summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily discusses non-clinical bench testing. There is no "test set" in the context of patient data described for evaluating device performance. The testing involved physical device units. The provenance of the device units used for bench testing (e.g., country of manufacture) is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable as there was no clinical test set requiring expert ground truth establishment. The evaluation was based on bench testing against engineering specifications and comparison to a predicate device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable as there was no clinical test set requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. The Aria System is a pneumatic compression device, not an AI-assisted diagnostic or interpretative device that would involve human "readers." No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not applicable. The Aria System itself is a standalone medical device (hardware) that delivers therapy; it's not an algorithm whose performance needs to be assessed independently. While it contains a microprocessor and potentially firmware/software for control, the evaluation focuses on the overall device's physical and functional performance through bench testing.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical testing, the "ground truth" or reference for evaluating performance was based on engineering specifications for the Aria System and the performance characteristics of the legally marketed predicate device (Entre Model PD08-U, K143185). For example, pressure accuracy would be compared to a calibrated standard.

    8. The sample size for the training set

    This question is not applicable. This device is not an AI/ML algorithm that requires a "training set" of data for learning.

    9. How the ground truth for the training set was established

    This question is not applicable as there was no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1