    Intended Use

    The Smart Neck Stimulator is designed to temporarily relieve pain associated with sore and aching muscles in the back of neck due to strain from exercise or normal household work activities.

    Device Description

    The subject device is intended to use on back of neck and it designed accord with human body cervical physiological curvature of streamlined ring design. The device totally has 3 modes and each mode has 15 output Intensity Levels, so the device can give certain electrical pulse through the electrode pads on the skin to help users to enjoy back neck stimulation. It can also provide a constant temperature of 37°C to 41°C to provide a warming sensation. The subject device can be controlled by a remote control which based on the radiofrequency technology. There are 5 buttons in the remote control, the "M" button for modes selection, "+" or "-" button for intensity selected to increase or decrease, the " "" button intends to turn the heating function on and the " Ho button intends to turn the heating function off. The device can be turned on when both power button on subject device has been turned on and the treatment mode can be selected by the remote control.

    AI/ML Overview

    The provided text is a 510(k) Summary for the "Smart Neck Stimulator" (K200558). This document primarily focuses on demonstrating substantial equivalence to predicate devices based on device description, intended use, and conformance to electrical safety, EMC, and biocompatibility standards through bench testing.

    Therefore, the information required to answer questions regarding acceptance criteria, study design (sample size, data provenance, expert ground truth, adjudication, MRMC, standalone performance), training set details, and ground truth establishment is not available in this document. The document details the physical and electrical safety characteristics rather than performance in a clinical, diagnostic, or AI-assisted scenario.

    The document's "Test Summary" and "Comparison to Predicate Device and Conclusion" sections refer to safety and performance by lab bench testing and adherence to standards, not clinical or AI-based performance studies.

    Here's what can be extracted from the document:

    • Device Name: Smart Neck Stimulator (Model: 4098 (RED))
    • Product Code: NUH
    • Regulation Number: 21 CFR 882.5890
    • Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief
    • Regulatory Class: Class II
    • Intended Use/Indications for Use: The Smart Neck Stimulator is designed to be used to temporarily relieve pain associated with sore and aching muscles in the back of neck due to strain from exercise or normal household work activities.
    • Type of Use: Over-The-Counter Use
    • Predicate Devices:
      • HIVOX BIOTEK INC., HIVOX OTC Electrical Stimulator (K171803)
      • Shenzhen OSTO Technology Co., Ltd., Neck Care Therapy (K172897)
    • Tests Performed (Bench Testing for Safety and Performance):
      • IEC 60601-1 (General Requirements For Basic Safety And Essential Performance)
      • IEC 60601-1-2 (Electromagnetic compatibility)
      • IEC 60601-2-10 (Particular Requirements For The Basic Safety And Essential Performance Of Nerve And Muscle Stimulators)
      • IEC 60601-1-11 (Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment)
      • IEC 62133-2 (Secondary Cells And Batteries - Safety Requirements for Lithium Systems)
      • IEC 62366-1 (Medical Devices - Application Of Usability Engineering To Medical Devices)
      • ISO 10993-5 (Biological evaluation of medical devices - Tests for In Vitro cytotoxicity)
      • ISO 10993-10 (Biological evaluation of medical devices - Tests for irritation and skin sensitization)

    Regarding the specific questions posed:

    1. A table of acceptance criteria and the reported device performance:

      • Acceptance Criteria: Implicitly, the acceptance criteria are compliance with the referenced IEC/ISO standards (e.g., electrical safety limits, biocompatibility standards, EMC limits) and demonstration that the device's technical specifications are comparable to or within acceptable limits relative to predicate devices, and would not raise new safety or effectiveness concerns.
      • Reported Device Performance: The document lists the device's specifications (e.g., output voltage, current, pulse duration, frequency, heating temperature, dimensions, weight) and states that bench tests were performed and results "complied with the test requests" and "meet the safety standards." However, it does not provide specific numerical performance results against an acceptance criterion in a tabular format beyond listing the device's characteristics and noting compliance.

      Example (Structured as requested, but data is qualitative in the document):

    Acceptance Criteria CategorySpecific Criteria (Implicitly from Standards & Predicate Comparison)Reported Device Performance (Summary from Document)
    Electrical SafetyCompliance with IEC 60601-1, IEC 60601-2-10, IEC 60601-1-11, IEC 62133-2 standards, and comparable to predicate devices for parameters like patient leakage current, average current through electrodes."results complied with the test requests" and "meet the safety standards IEC 60601-1 and IEC 60601-1-11." Specific values are provided for comparison to predicates (e.g., Patient Leakage Current NC, Average current through electrodes <10µA), deemed acceptable.
    Electromagnetic Comp.Compliance with IEC 60601-1-2 standard."results complied with the test requests" and "meet the safety standards."
    BiocompatibilityCompliance with ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and skin sensitization) standards."results complied with the test requests" and "meet the safety standards."
    Usability EngineeringCompliance with IEC 62366-1 standard."results complied with the test requests" and "meet the safety standards."
    Mechanical/PhysicalSpecifications (Weight, Dimensions, Pad dimensions) deemed not to adversely impact safety/effectiveness compared to predicates.Differences in physical dimensions (weight, dimensions, pad size) exist but are deemed not to impact safety and effectiveness, as they "meet the safety standards."
    Performance ParametersElectrical output parameters (Voltage, Current, Pulse Duration, Frequency, Net Charge, Phase Charge, Average Current, Current Density, Power Density) within acceptable range and comparable to predicates.Specific values are listed (e.g., Max Output Voltage 76V±10% @500Ω, Pulse frequency 1-200Hz). Differences from predicates are noted but "would not adversely impact the safety and effectiveness" as they "meet the safety standards."
    Operational FeaturesNumber of modes, intensity levels, heating function, timer range, automatic shut-off, etc., comparable to or within acceptable variations from predicates.Values listed (e.g., 3 modes, 15 intensity levels, 15 min timer). Differences from predicates (e.g., timer range, output intensity) are noted but deemed not to impact safety/effectiveness as they "meet the safety standards."
    1. Sample size used for the test set and the data provenance: Not applicable. This document describes bench testing and comparisons to predicate device specifications, not clinical or AI-based performance studies with patient data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for clinical or AI performance was not established. The "experts" involved would be the engineers and technicians performing the standard electrical, mechanical, and biological tests.
    3. Adjudication method: Not applicable. No clinical test set requiring adjudication.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done: No, an MRMC study was not done. This device is a Transcutaneous Electrical Nerve Stimulator for pain relief, not an imaging analysis AI device that would typically involve MRMC studies.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a physical medical device, not an AI algorithm.
    6. The type of ground truth used: Not applicable in the context of clinical or AI performance. The "ground truth" for this submission is regulatory compliance based on engineering and safety standards (e.g., a specific resistance value is 500 Ohm, a voltage reading must be within a certain tolerance as defined by the standard).
    7. The sample size for the training set: Not applicable. No AI model was trained.
    8. How the ground truth for the training set was established: Not applicable. No AI model was trained.

    In summary, this 510(k) submission establishes substantial equivalence through a detailed comparison of the subject device's design, intended use, and technical specifications against existing predicate devices and by demonstrating adherence to recognized safety and performance standards via bench testing. It does not involve AI model performance evaluations, clinical studies with patient data, or expert-adjudicated ground truth.