510(k) Data Aggregation

    K Number
    K193465

    Validate with FDA (Live)

    Date Cleared
    2020-01-10

    (25 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CorticaLINK Spinal Fusion Platform is intended for non-cervical pedicle fixation as an adjunct to fusion for the following acute and chronic instabilities or deformities for the T1-S1 vertebrae in skeletally mature patients:

    · degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)

    • · spondylolisthesis
    • · trauma (i.e., fracture or dislocation)
    • spinal stenosis
    • · curvatures (i.e., scoliosis, kyphosis, and/or lordosis)
    • tumor
    • · pseudoarthrosis
    • · and failed previous fusion
    Device Description

    The CorticaLINK Spinal Fusion Platform is a comprehensive suite of fixation implants which can be used to stabilize the spine as an adjunct to fusion following surgical decompression. The platform is comprised of screws, rods and locking caps, with the screws being offered in several different lengths, diameters, and thread pitches to accommodate varying anatomies, pathologies, and surgeon preferences.

    The systems components are manufactured using standard manufacturing processes. The tulips, screw shafts, and one family of rods are of medical grade Titanium alloy (Ti6Al4V) that complies with ASTM F136 - Standard Specifications for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Allov for Surgical Implant Applications. The push plate is of Commercially Pure Grade4 Titanium (CP Ti) per ASTM F67 - Standard Specifications for Unalloyed Titanium for Surgical Implant Applications (UNS R50250, R50400, R50500 & R50700). The other family of rods are of Cobalt Chromium (CoCr) per ASTM F75 - Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants (UNS R30075).

    The implant components are provided clean and non-sterile. These devices are supplied in a rigid sterilization tray and are to be sterilized by a healthcare professional using a Steam Autoclave in accordance with the instructions for use provided by Spine Smith Holdings. LLC, as well as the instructions provided by the Autoclave manufacturer.

    AI/ML Overview

    The provided text is a 510(k) summary for the CorticaLINK Spinal Fusion Platform, a medical device for spinal fixation. It describes the device's intended use, materials, and equivalence to a predicate device. However, it does not contain information regarding the acceptance criteria, specific device performance metrics, or study details (like sample size, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies) typically associated with software or AI device clearances.

    The document primarily discusses mechanical performance testing against a predicate device, as per ASTM F1717, which is standard for orthopedic implants.

    Therefore, many of the requested categories cannot be populated from the provided text.

    Here's a breakdown of what can and cannot be answered:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Not Explicitly Stated for the Device)Reported Device Performance
    Implicit: Substantial equivalence in mechanical performance to predicate device per ASTM F1717 for static and dynamic compression bending and static torsion.The worst-case subject device was tested according to ASTM F1717 in static and dynamic compression bending as well as static torsion. Results demonstrate substantially equivalent mechanical performance as compared to the predicate(s).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size (Test Set): Not specified. The document refers to "the worst-case subject device" being tested, implying a single or limited number of devices for mechanical testing, not a data sample.
    • Data Provenance: Not applicable, as this refers to mechanical testing of physical orthopedic implants, not data analysis for software/AI.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth and expert review are typically associated with performance studies for diagnostic or AI-driven devices, not mechanical testing of implants.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This refers to adjudication of expert opinions for ground truth establishment, which is not relevant to mechanical testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC study was not done. This device is an orthopedic implant, not an AI or diagnostic tool that involves human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is an orthopedic implant, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable. For mechanical testing, the "ground truth" is defined by the physical properties and performance benchmarks specified in the ASTM standard.

    8. The sample size for the training set

    • Not applicable. This device does not involve a "training set" as it is not a machine learning or AI-driven device.

    9. How the ground truth for the training set was established

    • Not applicable. This device does not involve a "training set."

    Summary of the Study that Proves the Device Meets the Acceptance Criteria:

    The device's acceptability is proven through performance testing conducted in accordance with ASTM F1717. This standard evaluates the mechanical performance of spinal implant systems. The testing included static and dynamic compression bending, as well as static torsion. The conclusion states that "Results of the testing demonstrate substantially equivalent mechanical performance as compared to the predicate(s)." This means the CorticaLINK Spinal Fusion Platform met the implicit acceptance criterion of performing mechanically equivalently to legally marketed predicate devices.