510(k) Data Aggregation

    K Number
    K193006

    Validate with FDA (Live)

    Date Cleared
    2020-08-09

    (286 days)

    Product Code
    Regulation Number
    882.5802
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MagVenture TMS Therapy System is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD).

    Device Description

    The MagVenture TMS Therapy system - for adjunctive treatment of OCD is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic fields to induce electrical currents targeting specific regions of the cerebral cortex. The MagVenture TMS Therapy system - for treatment of OCD is indicated as an adjunct for the treatment of adult patients who are suffering from Obsessive-Compulsive Disorder (OCD). MagVenture TMS Therapy system has previously obtained FDA clearance for treatment of major depressive disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode (K150641, K171481, K172667 and K173620).

    Transcranial magnetic stimulation (TMS) is a non-invasive technique for stimulating brain and neural tissue. The principle of magnetic stimulation is implicit in Faraday's law. The pulses of current are generated with a circuit containing a capacitor connected to the stimulating coil. With the capacitor charged to a certain level, the conducting state will cause the discharging of the capacitor through the coil. A magnetic field is generated proportional to this current. The rapid change in the magnetic field induces a current in conducting materials e.g. the body tissue. If the current induced in the human body is of sufficient amplitude and duration, it will excite neurons. The standard-of-care FDA cleared TMS protocol for treatment of OCD uses repetitive transcranial magnetic pulses applied at a frequency of 20 Hz. The safety and effectiveness for treatment of OCD have been established in a clinical trial that lead to an FDA De Novo clearance of the primary predicate device, the Brainsway DTMS system. The present 510(k) does not include new pivotal data, but includes clinical trial data on more than 500 subjects treated with the Mag Venture TMS Therapy System, in order to demonstrate performance and safety. Treatment of OCD is applied to the human brain of the bilateral dorsomedial prefrontal cortex (DMPFC) using 20 Hz TMS for 18 min. The treatment parameters are identical to those recommended by the Primary Predicate Device.

    The MagVenture TMS Therapy system – for adjunctive treatment of OCD is an integrated system consisting of the following components:

    • MagPro Stimulator and Trolley .
    • MagPro Family (R30, R30 w. MagOption, X100, X100 w. MagOption) o
    • Trolley with holding arrangements o
    • Coil for MT determination and OCD treatment .
    • Coil Cool D-B80 with Coil Cooler Unit o
    • Marking apparatus for locating treatment area .
    • Pen for marking, Cap, Ruler o
    • Patient head fixation .
    • Treatment Chair o
    • Vacuum Pump and Vacuum pillow o
    • Super Flexible Arm mounted on the trolley o
    • Isolation Transformer .
    AI/ML Overview

    The provided text is a 510(k) summary for the MagVenture TMS Therapy System for adjunctive treatment of OCD. It describes the device, its intended use, and argues for its substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    No specific acceptance criteria table for effectiveness is provided in the document. The document establishes substantial equivalence based on prior clearances, identical intended use and treatment parameters with a primary predicate device, and non-clinical and clinical data supporting safety and performance.

    The acceptance of this device seems to be primarily based on demonstrating a substantial equivalence to an already cleared predicate device (Brainsway Deep Transcranial Magnetic Stimulation (DTMS) system, HAC – H7 coil (DEN170078, K183303)) in terms of safety and performance for the adjunctive treatment of OCD.

    The document states: "The new device is substantially equivalent to the primary predicate device in terms of performance, effectiveness and safety."

    Here's a breakdown of the requested information based on the provided text:


    1. A table of acceptance criteria and the reported device performance

    As mentioned, no explicit table of acceptance criteria for effectiveness is present. The primary "acceptance criterion" for this 510(k) is demonstrating substantial equivalence to the primary predicate device, particularly regarding its ability to treat OCD. The reported device performance is discussed in terms of its similarity to the predicate and general safety/tolerability demonstrated in other studies.

    Acceptance Criterion (Implicit for Substantial Equivalence to Predicate)Reported Device Performance
    Identical Indication for Use"The MagVenture TMS Therapy System is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD)." - Identical to Primary Predicate.
    Identical Treatment Protocol & Target"The proposed indication for use and treatment protocol are identical to that of the primary predicate device." "Treatment of OCD is applied to the human brain of the bilateral dorsomedial prefrontal cortex (DMPFC) using 20 Hz TMS for 18 min. The treatment parameters are identical to those recommended by the Primary Predicate Device."
    Substantially Equivalent Magnetic Field Properties & Spatial DistributionNon-clinical testing and e-field modeling demonstrated: "the magnetic field strength in the cortex for the two coils," showing "that the Cool D-B80 coil is able to reach the deeper underlying cortical layers similarly to the HAC - H7 coil." "the magnetic field properties of the Cool D-B80 coil is substantially equivalent to the primary predicate device, HAC - H7 in terms of magnetic field properties and realized magnetic field properties."
    Safety and Tolerability"clinical trial data on more than 500 subjects treated with the MagVenture TMS Therapy System, in order to demonstrate performance and safety." "treatment with the new device is safe and provides equivalent performance to the primary predicate device in terms of safety and performance." "The most common side effects reported for both devices are headache and/or pain at stimulation site." "The treatment with the new device does not introduce any new adverse or serious adverse events."
    Effectiveness (Supported by Predicate effectiveness)"The clinical evidence submitted also supports the safety and use of the new device as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder (OCD)." "A small pilot trial using the new device also demonstrated a clinically and statistically significant effect of treatment of OCD."

    2. Sample size used for the test set and the data provenance

    • Test Set for Effectiveness (Implicitly for OCD): The primary predicate device's pivotal trial (Carmi et al. (2)) included a total of n=99 subjects, of which n=50 received active TMS treatment for OCD. The provenance of this data is a "Prospective Multicenter Randomized Double-Blind Placebo-Controlled Trial." (Page 12, reference 2). This study's country of origin is not explicitly stated in this document but is typically associated with the predicate manufacturer (Brainsway Ltd., Israel).
    • Test Set for Safety and Performance of the New Device: "clinical trial data on more than 500 subjects treated with the MagVenture TMS Therapy System" (Cool D-B80 coil). This data focuses on safety and general performance (e.g., determining Motor Threshold). "Most of the clinical research performed using the new device has been conducted outside the USA." (Page 7). Most research focused on treatment-resistant depression (TRD) or major depressive disorder (MDD), but the document argues for relevance to OCD due to shared underlying deficits and similar treatment targeting. A "small pilot trial using the new device also demonstrated a clinically and statistically significant effect of treatment of OCD," but the number of subjects for this specific OCD trial is described as "low number of subjects" and not specified further.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not provide details on the number or qualifications of experts for establishing ground truth specifically for the effectiveness test set. The effectiveness relies on the previously published pivotal trial for the primary predicate device (Carmi et al.), which would have utilized clinical diagnoses and outcome measures as ground truth. For the predicate study, the "ground truth" for OCD diagnosis would have likely been established by qualified psychiatrists or clinicians using standardized diagnostic criteria (e.g., DSM-V).

    4. Adjudication method for the test set

    The document states that the primary predicate device's pivotal trial was a "Prospective Multicenter Randomized Double-Blind Placebo-Controlled Trial." (Page 12, reference 2). This implies a rigorous study design where treatment assignments and outcome assessments would have been blinded, contributing to the establishment of reliable "ground truth" regarding treatment efficacy. However, a specific "adjudication method" beyond standard blinded clinical trials is not explicitly detailed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. This device is not an AI/CAD system evaluated by human readers interpreting images. It is a medical device for direct therapeutic intervention. The "effectiveness" is assessed by measuring clinical outcomes in patients.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a TMS therapy system, not an algorithm or AI. It is a physical device that delivers magnetic stimulation to the brain for therapeutic purposes. Human operators are involved in setting up and operating the device for patient treatment.

    7. The type of ground truth used

    • For effectiveness (via predicate study): Clinical diagnosis of Obsessive-Compulsive Disorder (OCD) and clinical outcome measures (e.g., symptom severity scales like Y-BOCS - Yale-Brown Obsessive Compulsive Scale, though not explicitly mentioned here, this is standard for OCD trials) in a controlled clinical trial setting.
    • For safety/performance of the new device: Clinical observation of adverse events, tolerability, and physiological measurements (e.g., Motor Threshold determination).

    8. The sample size for the training set

    This document does not describe a "training set" in the context of machine learning or AI. The device is a physical medical device. The closest analogy might be the clinical data supporting the device's original clearances for MDD, which forms the basis of its design and operational parameters. The document mentions "clinical trial data on more than 500 subjects treated with the MagVenture TMS Therapy System," mostly for MDD, which informed safety and general operational performance, effectively acting as "training" for understanding the device's characteristics in a clinical setting.

    9. How the ground truth for the training set was established

    Again, this is not a machine learning context. For the clinical data on >500 subjects (mostly MDD), the "ground truth" for the disease state would have been clinical diagnoses of MDD or TRD by qualified medical professionals using standard diagnostic criteria. For safety aspects, reported adverse events and clinical assessments would have constituted the "ground truth."