Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vein set is intended for vein puncture to collect blood specimens for patients. It is also indicated for intravenous administration of fluids after removing the attached luer adapter from the blood collection set connector and attaching a syringe, or other compatible/appropriate device may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy

    Safety vein set is intended for vein puncture to collect blood specimens. It is also indicated for intravenous administration of fluids after removing the attached luer adapter from the blood collection set connector and attaching a syringe, or other compatible/appropriate device may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. Additionally, after withdraw of the needle from the patient's vein, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle stick.

    Blood collection needle (vein set type) is intended for use in the blood collection or short-term infusion (up to 2 hours) of intravenous fluids.

    Blood collection needle (needle holder type) is intended for use in the blood collection or short-term infusion (up to 2 hours) of intravenous fluids.

    Safety blood collection needle (vein set type) is intended for use in the blood collection or short-term influsion (up to 2 hours) of intravenous fluids. The device is designed with a safety mechanism to help reduce the risk injury.

    Safety blood collection needle (needle holder type) is intended for use in the blood collection or short-term infusion (up to 2 hours) of intravenous fluids. The device is designed with a safety mechanism to help reduce stick injury

    Device Description

    The proposed device is a blood collection device that forms a passage between the patient's vein and a vacuum blood collection tube for collecting blood. The proposed device can also be used for short-term infusion. The proposed devices are divided into several types, some have safety devices and some do not have safety devices, and the safety mechanism helps to reduce the risk of needle stick injuries. The proposed devices are provided sterile, single use. The propose device can be used for vein collect blood/infusion as well as peripheral vein collect blood/infusion, but not central vein collect blood/infusion.

    AI/ML Overview

    The provided text details the 510(k) submission for several blood collection and intravenous administration devices. It focuses on demonstrating substantial equivalence to a predicate device, rather than defining and proving acceptance criteria for a new, innovative device with a complex algorithm. Therefore, many of the questions about acceptance criteria, specific performance metrics, expert qualifications, and study designs for algorithmic performance (like MRMC studies) are not directly applicable or available in this document.

    However, I can extract information related to the device's performance as demonstrated by the non-clinical tests conducted to support its substantial equivalence. The "acceptance criteria" here are largely defined by the requirements of the referenced ISO standards and general safety/effectiveness considerations for medical devices of this type.

    Here's an attempt to answer your questions based on the provided text, while acknowledging its limitations for certain aspects:


    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly state "acceptance criteria" as singular numerical targets for a new algorithm. Instead, it refers to conformity with established international standards for medical devices and safety features. The "reported device performance" is essentially the outcome of these tests, demonstrating compliance.

    Acceptance Criteria (Implicit from Standards)Reported Device Performance
    Biocompatibility
    In Vitro Cytotoxicity: No cytotoxicityNo Cytotoxicity (Comply with ISO 10993 requirements)
    Skin Sensitization: No sensitizationNo Sensitization (Comply with ISO 10993 requirements)
    Intracutaneous Reactivity: No intracutaneous reactivityNo Intracutaneous Reactivity (Comply with ISO 10993 requirements)
    Acute Systemic Toxicity: No systemic toxicityNo Systemic Toxicity (Comply with ISO 10993 requirements)
    Hemolytic Properties: No hemolytic propertiesNo Hemolytic (Comply with ISO 10993 requirements)
    Pyrogenicity: No pyrogenicityNo Pyrogen (Comply with ISO 10993 requirements)
    In Vivo Thromboresistance: (Implicitly, desired outcome is no)No (Comply with ISO 10993 requirements)
    Complement Activation: (Implicitly, desired outcome is no)No (Comply with ISO 10993 requirements)
    Sterilization
    Sterilization Method: EO SterilizedEO sterilized (Same as predicate)
    Sterility Assurance Level (SAL): 10^-610^-6 (Same as predicate)
    Endotoxin Limit: 20EU per device20EU per device (Same as predicate)
    Physical/Performance
    Conical fittings (Luer taper) per ISO 594-1 & 594-2Complies with ISO 594-1:1986 and ISO 594-2:1998
    Infusion sets for single use per ISO 8536-4Complies with ISO 8536-4:2010
    Sterile hypodermic needles per ISO 7864Complies with ISO 7864:2016
    Stainless steel needle tubing per ISO 9626Complies with ISO 9626:2016
    Particulate Matter testing per USP <788>Complies with USP <788> Particulate Matter testing (Method 1)
    Seal Strength per ASTM F88/F88M-15Complies with ASTM F88/F88M-15
    Detecting Seal Leaks per ASTM F1929-15Complies with ASTM F1929-15
    Safety Mechanism (for safety models)Device safety mechanism performance
    Simulated Clinical StudyMeet pre-established criteria

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document generally refers to "non clinical tests" and "simulated clinical study" without specifying the exact sample sizes (e.g., number of devices tested) for each specific test mentioned (e.g., cytotoxicity, seal strength). The provenance of the data (country of origin, retrospective/prospective) is not explicitly stated, but the submission is from Jiangsu Kangbao Medical Equipment Co., Ltd. in China, implying testing was likely conducted in China or by labs contracted by them. The studies described are non-clinical and simulated clinical, not human clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable to the provided document. The device is a physical medical instrument (vein set, blood collection needle), not an AI/algorithmic device that requires expert review to establish "ground truth" for interpretations of data like images or clinical signals. The "truth" here is determined by objective physical and biological measurements against established standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or performance evaluations where human experts are making subjective assessments and a consensus is needed. For physical medical devices tested against standards, the "ground truth" is determined by the test methodology itself (e.g., a tensile strength test yields a numerical result, a cytotoxicity test yields a "yes/no" result based on cellular response).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. MRMC studies are used to evaluate the performance of diagnostic imaging aids or similar AI-powered tools that assist human readers. The submitted device is a physical sterile device for blood collection and IV administration, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device described does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this type of device, the "ground truth" is based on:

    • Compliance with international standards: ISO standards (e.g., ISO 10993 for biocompatibility, ISO 594 for Luer fittings, ISO 7864 for needles) and US Pharmacopeia (USP <788>) for particulate matter.
    • Physical and chemical tests: Measuring attributes like seal strength, leak detection, mechanical properties, and material composition.
    • Biological tests: Assays for cytotoxicity, sensitization, systemic toxicity, pyrogenicity, hemolysis, etc.
    • Safety mechanism evaluation: Verification that the safety feature (for applicable models) functions as intended to prevent needle stick injuries, as evaluated in a "simulated clinical study" and "safety feature test".

    8. The sample size for the training set

    This question is not applicable, as there is no mention of an algorithm or training set for this physical medical device.

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no mention of an algorithm or training set for this physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1