Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192563

    Validate with FDA (Live)

    Date Cleared
    2020-01-17

    (122 days)

    Product Code
    Regulation Number
    868.1400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Altera Gas Sampling Lines are intended to connect from a port in the breathing circuit to the expired gas monitor.

    Device Description

    The Altera Gas Sampling Lines consists of a disposable single patient use coextruded gas sampling line which is provided in 3 different lengths: 2, 2.5, and 3 meters. The disposable single patient use gas sampling lines are smooth narrow diameter tubes that have standardized male luer connectors at both ends.

    The gas sampling line connects from a port in a circuit to an expired gas monitor. The gas sampling line provides a conduit for drawing gas samples from the sampling port to the gas monitor to analyze respiratory gases. These disposable single patient use gas sampling lines are used to transmit one directional flow of gas sample from the patient breathing circuit port to the gas module host device. A vacuum source and gas measurement sensors are in the host device, which pull the gas from the sampling port to the host device for gas monitoring.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device, specifically "Altera Gas Sampling Lines." It's important to understand that a 510(k) submission generally focuses on demonstrating substantial equivalence to a predicate device, rather than proving independent effectiveness through extensive clinical trials as might be seen for novel technologies or higher-risk devices.

    Based on the provided text, the device is a simple, single-use, disposable component (a gas sampling line) that connects a breathing circuit to an expired gas monitor. There is no evidence in this document of AI or machine learning components, nor any mention of a study involving human readers improving with AI assistance. Therefore, the questions related to AI performance, expert ground truth establishment for AI, MRMC studies, or standalone algorithm performance are not applicable to this submission.

    Here's an analysis of the provided information concerning the acceptance criteria and the study that proves the device meets them, primarily focusing on the device's physical and functional performance, as that is the scope of this 510(k).


    Acceptance Criteria and Device Performance for Altera Gas Sampling Lines

    Given the nature of this device (a gas sampling line), the "acceptance criteria" are primarily related to its physical and functional performance characteristics, ensuring it performs its intended function (transporting gas samples) safely and effectively, and equivalently to the predicate device. The "study" proving this involves various non-clinical performance tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    The table below summarizes the key performance attributes tested and the results presented in the document. The general acceptance criterion for these tests is that the proposed device (Altera GSL) performs comparably or better than the predicate device (Vyaire Vital Signs) and meets relevant industry standards.

    Acceptance Criteria (Attribute)Predicate (Vyaire Vital Signs) Performance (Reference/Value)Proposed (Altera GSL) Performance (Reference/Value)Comparison
    ClassificationCCK - Carbon Dioxide Gas Analyzer, 21 CFR 868.1400CCK - Carbon Dioxide Gas Analyzer, 21 CFR 868.1400Same
    Indications for UseConnects from a port in the breathing circuit to the expired gas monitor. Used with specific GE Healthcare monitors.Connects from a port in the breathing circuit to the expired gas monitor.Similar
    Principle of OperationGas pulled from one end of the tube to the other by a pump in the gas sampling device.Gas pulled from one end of the tube to the other by a pump in the gas sampling device.Similar
    Patient PopulationAny patient population requiring gas monitoring.Any patient population requiring gas monitoring.Similar
    Environments of UseHospitals.Hospitals, sub-acute, pre-hospital.Similar
    CompatibilityDesigned for specific GE Healthcare modules.Designed for use with any gas monitoring device with luer connections.Similar
    Single patient use, disposableYes.Yes.Similar
    MaterialsISO 10993 tested, Co-extruded PE/EVA/PVC.ISO 10993 tested, Co-extruded PE/EVA/PVC.Similar
    Flow Resistance~26 cmH2O @ 150 ml/min.~12 - 17.5 cmH2O @ 150 ml/min.Similar
    Leakage (per ISO 80601-2-55)< 10 ml/min.< 10 ml/min.Similar
    Tubing ID / ODID - 1.2 mm, OD - 2.8 mm.ID - 1.2 mm, OD - 2.8 mm.Similar
    Lengths AvailableUp to 3 meters.2, 2.5 and 3 meters.Similar
    ConnectorsStandard ISO 80369 small bore luer fittings – male / female (referenced ISO 594-2).Standard ISO 80369 small bore luer fittings – male / female (referenced ISO 80369-7).Similar
    Separation Force of Connectors>35 Newtons.>35 Newtons.Similar
    Leakage of Connectors0.02-0.05 ml/min.< 0.005 Pa*m3/s (units per ISO 80369-7).Similar
    ISO Luer Standard ComplianceISO 594-2 (Unscrewing torque, Resistance to overriding, Stress cracking).ISO 80369-7 (Unscrewing torque, Resistance to overriding, Stress cracking).Similar
    Performance post agingN/A (for predicate, not explicitly stated).After 3 years accelerated aging, performance testing still met pre-established acceptance criteria.Improved/Additional

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes for each physical and functional test (e.g., how many gas sampling lines were tested for flow resistance, leakage, or mechanical properties). It broadly states "Performance Testing" was completed. For a 510(k) of this type, standard engineering and quality control practices would dictate appropriate sample sizes for manufacturing verification, but these details are not provided in the summary document.
    • Data Provenance: The tests would have been performed by the manufacturer, Meditera Tibbi Malzeme San. ve Tic. A.S., or their designated testing facilities. The country of origin for the manufacturing is Turkey (Izmir). The data provenance is retrospective in the context of the 510(k) submission, meaning the tests were completed and the data collected prior to the submission for regulatory review. There is no mention of prospective clinical data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not Applicable. As previously stated, this device is a physical component without AI/ML. "Ground truth" in this context refers to engineering specifications and performance standards (e.g., ISO standards for luer fittings, flow resistance measurements). These are established through scientific principles and engineering testing, not by expert human interpretation of medical images or data.

    4. Adjudication Method for the Test Set

    • Not Applicable. Since there are no human experts evaluating data (as in clinical trials or AI performance evaluations), there is no need for an adjudication method. Test results are based on objective measurements against engineering specifications and standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. An MRMC study is relevant for evaluating the impact of AI on human performance (e.g., radiologists interpreting images). This device is a simple medical accessory, not an AI software.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not Applicable. This device does not involve an algorithm. The "standalone performance" refers to the physical and functional integrity of the gas sampling line itself, as detailed in the performance testing section.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's acceptance is based on engineering specifications, international performance standards (e.g., ISO 80601-2-55 for leakage, ISO 80369-7 for luer fittings), and comparative performance data against a legally marketed predicate device. The tests performed include:
      • Flow resistance
      • Accelerated aging (for 3-year shelf-life)
      • Mechanical testing (luer fitting properties: fluid leakage, air leakage, stress cracking, separation force, unscrewing torque, resistance to overriding)
      • Air leakage (general)

    8. The Sample Size for the Training Set

    • Not Applicable. This document is for a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. No training set exists for this type of device.

    Conclusion from the document:

    The manufacturer concluded, and the FDA agreed in the 510(k) clearance, that the "Altera Gas Sampling Lines" are substantially equivalent to the predicate device (Vyaire Vital Signs Gas sampling Lines) based on a comparison of intended use, principle of operation, technological characteristics, and non-clinical performance testing. The performance testing ensured the device met applicable standard requirements and did not raise new questions of safety or effectiveness compared to the predicate.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1