Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192503

    Validate with FDA (Live)

    Date Cleared
    2020-11-19

    (434 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GM501 Wash with Phenol Red and Gentamicin is intended for in vitro procedures involving handling and micromanipulation of human oocytes and embryos outside of a CO2 incubator. Include oocyte and embryo washing (e.g. after oocyte aspiration, after hyaluronidase treatment to remove cumulus cells, before and after cryopreservation, and before embryo transfer) and micromanipulation procedures (e.g. assisted hatching). GM501 Wash with Phenol Red and Gentamicin is not intended for use in transferring embryos into the uterine cavity.

    Device Description

    GM501 Wash with Phenol Red and Gentamicin is a ready-to-use solution providing supporting conditions for human oocytes and embryos during in vitro Assisted Reproduction Technology (ART) procedures taking place outside of a CO2 incubator, including washing and micromanipulation procedures. GM501 Wash is aseptically filled into sterilized bottles (20 and 50 ml) and has a six-month shelf-life when stored as recommended. This product can also be used for up to seven days after bottle opening.

    GM501 Wash with Phenol Red and Gentamicin is identical to the predicate GM501 Wash with the exception of the addition of gentamicin sulfate (10 mg/liter) and phenol red (3 mg/liter).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text, structured to address your specific points:

    1. A table of acceptance criteria and the reported device performance

    ParameterAcceptance Criteria (Subject Device - GM501 Wash with Phenol Red and Gentamicin)Reported Device Performance (Subject Device)Comparison to Predicate Device
    Endotoxins< 0.25 LAL, EU/ml< 0.25 LAL, EU/mlSame
    Osmolality (mOsm/Kg)270-290270-290Same
    pH7.2-7.57.2-7.5Same
    1-Cell MEA (Mouse Embryo Assay)≥ 80% blastocyst at 96h after 1h exposure to GM501 Wash with Phenol Red and Gentamicin≥ 80% blastocyst at 96h after 1h exposure to GM501 Wash with Phenol Red and GentamicinSame
    Shelf-life6 months6 months (supported by testing of MEA, sterility, pH, osmolality, and endotoxins)Same
    Use-life after bottle opening7 days (supported by stability testing at the end of the shelf-life period with MEA, sterility, pH, osmolality, and endotoxins)7 days (supported by stability testing at the end of the shelf-life period with MEA, sterility, pH, osmolality, and endotoxins)N/A (this specific testing addresses a formulation change in the subject device)
    SterilizationSterilized by sterile filtrationSterilized by sterile filtrationSame

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for the "test set" in terms of human subjects or embryos. The primary performance data presented is the 1-Cell MEA (Mouse Embryo Assay) which uses mouse embryos. The specific number of mouse embryos used in this assay is not provided.

    The data provenance (country of origin, retrospective/prospective) is also not explicitly stated in the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not present in the document. The device is a reproductive media for in vitro procedures. Its performance is assessed through laboratory assays (like MEA) and chemical/physical property measurements, not through expert interpretation of images or clinical diagnoses.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not present in the document. The device's performance is determined by objective laboratory measurements (MEA, pH, osmolality, endotoxins, sterility) rather than subjective human interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and not present in the document. The device is a reproductive media, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and "human reader improvement with AI" are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable and not present in the document. The device is a biological/chemical media, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for assessing the performance of GM501 Wash with Phenol Red and Gentamicin is based on:

    • Laboratory Assay Results: Primarily the Mouse Embryo Assay (MEA), which assesses the ability of the media to support blastocyst development from 1-cell mouse embryos.
    • Chemical/Physical Specifications: Adherence to defined ranges for pH, osmolality, and endotoxin levels.
    • Sterility Testing: Confirmation of absence of microbial contamination.
    • Stability Testing: Demonstrating that the media maintains its specifications over its shelf-life and use-life.

    8. The sample size for the training set

    This information is not applicable and not present in the document. The device is a reproductive media, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable and not present in the document. The device is a reproductive media, not an AI model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1