510(k) Data Aggregation

    K Number
    K192383

    Validate with FDA (Live)

    Device Name
    BriefCase
    Date Cleared
    2019-12-20

    (112 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of head CTA images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive findings of Large Vessel Occlusion (LVO) pathologies.

    BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase is a radiological computer-assisted triage and notification software device. The software system is based on an algorithm programmed component and is comprised of a standard off-the-shelf operating system, the Microsoft Windows server 2012 64bit, and additional applications, which include PostgreSQL. DICOM module and the BriefCase Image Processing Application. The device consists of the following three modules: (1) Aidoc Hospital Server (AHS) for image acquisition; (2) Aidoc Cloud Server (ACS) for image processing; and (3) Aidoc Worklist Application for workflow integration, installed on the radiologist' desktop and provides the user interface in which notifications from the BriefCase software are received.

    DICOM images are received, saved, filtered and de-identified before processing. Series are processed chronologically by running an algorithm on each series to detect suspected findings and then notifications on flagged series are sent to the Worklist desktop application, thereby prompting preemptive triage and prioritization. The user may opt to filter out notifications by pathology, e.g. a chest radiologist may choose to filter out notifications on Large Vessel Occlusion (LVO) cases, and a neuro-radiologist would opt to divert Pulmonary Embolism (PE) notifications. In addition, where several medical centers are linked to a shared PACS, a user may read cases for a certain center but not for another, and thus may opt to filter out notification by center. Activating the filter does not impact the order in which notifications are presented in the Aidoc worklist application.

    The Worklist Application displays the pop-up text notifications of new studies with suspected findings when they come in. Notifications are in the form of a small pop-up containing patient name, accession number and the relevant pathology (e.g., LVO). A list of all incoming cases with suspected findings is also displayed. Hovering over a notification or a case in the worklist pops up a compressed, small black and white, unmarked image that is captioned "not for diagnostic use" and is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

    Presenting the radiologist with notification facilitates earlier triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

    AI/ML Overview

    Based on the provided text, here's a detailed description of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Performance Goal)Reported Device Performance
    Sensitivity> 80%88.8% (95% CI: 81.9%, 93.8%)
    Specificity> 80%87.2% (95% CI: 82.5%, 91.1%)
    Time-to-notificationComparable to predicate device3.8 min (95% CI: 3.6-4.0)
    Positive Likelihood Ratio (PLR)Not explicitly stated as an AC, but reported6.9 (95% CI: 5.0-9.6)
    Negative Likelihood Ratio (NLR)Not explicitly stated as an AC, but reported0.13 (95% CI: 0.1-0.2)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 383 cases
    • Data Provenance: The data was collected from 3 US-based clinical sites and was retrospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    The document does not explicitly state the number of experts or their specific qualifications (e.g., years of experience) used to establish the ground truth for the test set. It only mentions "reviewers" identified "True Positive cases."

    4. Adjudication Method for the Test Set

    The document does not explicitly state the adjudication method used for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. The study evaluated the standalone performance of the AI algorithm and compared its notification time to a predicate device. It did not assess human reader performance with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone study was done. The pivotal study evaluated the BriefCase software's performance in identifying LVOs, reporting its sensitivity and specificity, which implies an algorithm-only evaluation. The statement "The BriefCase time-to-notification includes the time to get the DICOM exam, de-identify it, upload it to the cloud, analyze and send a notification on a positive suspect case back to the worklist application" further supports an algorithm-only performance measurement before any human interaction.

    7. The Type of Ground Truth Used

    The ground truth was established by "reviewers." The document implies it's an expert consensus or expert interpretation of the images, as "True Positive cases (i.e., identified as positive both by the reviewers as well as the BriefCase device)" are mentioned. It does not mention pathology or outcomes data as ground truth.

    8. The Sample Size for the Training Set

    The document does not specify the sample size for the training set. It only states that the device uses an "artificial intelligence algorithm trained on medical images" and an "artificial intelligence algorithm with database of images."

    9. How the Ground Truth for the Training Set Was Established

    The document does not explicitly state how the ground truth for the training set was established. It only mentions that the algorithm was "trained on medical images."