Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192356

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-12-14

    (107 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Airlock® Centrolock® osteosynthesis implant systems are single-use devices intended for fixation and stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the hands, feet, ankle, fingers and toes.

    Examples include:

    • Mono or Bi-Cortical osteotomies in the foot or hand
    • Distal or Proximal metatarsal or metacarpal osteotomies
    • Fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc.)
    • Calcaneus/cuboid arthrodesis
    • Talar/navicular arthrodesis

    The system may be used in adult patients.

    Device Description

    Airlock® Centrolock® osteosynthesis implant systems are single-use bone fixation devices intended to be permanently implanted. implant systems are designed with different shapes and are made of Titanium (Alloy Ti-6Al-4V ELI). The system uses either 2mm cortical screws or 2,5mm locking screws. The drill holes are aligned to make sure there is no risk of conflict between the screws. The implant vary essentially through different curvatures and shapes.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Airlock® Centrolock® osteosynthesis implant system based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document (K192356) is a 510(k) summary for a medical device. For such devices, "acceptance criteria" are typically met through performance testing that demonstrates substantial equivalence to a predicate device, rather than specific numerical performance metrics for an AI algorithm. The document emphasizes mechanical and material characteristics.

    Acceptance Criterion (Established through substantially equivalent predicate)Reported Device Performance (as demonstrated by testing and comparison)
    Material CompositionWrought Titanium 6-Aluminum 4-Vanadium Alloy (ISO 5832-3 Fourth edition 2016-10-15)
    Mechanical Characteristics (Implants)Similar static and dynamic bending resistance (tested per ASTM-F382-17) to predicate Novastep Airlock® osteosynthesis Plate System.
    Mechanical Characteristics (Screws)Similar pull-out strength and torsional resistance (tested per ASTMF543-17 sections A1 A3) to predicate Novastep Airlock® osteosynthesis Plate System.
    BiocompatibilityEvaluation conducted in accordance with Blue Book Memorandum #G95-1 and ISO 10993-1.
    Intended Use / Indications for UseSame as predicate devices: Fixation and stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the hands, feet, ankle, fingers, and toes in adult patients.
    DesignSimilar technological characteristics in terms of design to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    This is not an AI/software device submission. The "test set" in this context refers to the samples used for mechanical and biocompatibility testing.

    • Sample Size: Not explicitly stated in terms of specific numbers of implants or screws tested for mechanical properties. Biocompatibility would involve various test samples, also not specified numerically.
    • Data Provenance: Not applicable in the context of clinical data. The tests are material and mechanical characterization tests conducted in a laboratory setting. No country of origin for clinical data is relevant here as no clinical studies were performed. The tests are prospective in the sense that they were designed and executed for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. This is not an AI device where expert ground truth is typically established. The "ground truth" for mechanical testing is based on established engineering standards (ASTM, ISO).

    4. Adjudication Method for the Test Set

    Not applicable. This is not an AI device with human expert interpretation or adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical implant, not an AI-powered diagnostic or assistive device. No MRMC study was performed.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical implant, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's acceptance is based on:

    • Engineering Standards: Established international standards for material composition (ISO 5832-3), mechanical testing (ASTM F382-17, ASTM F543-17), and biocompatibility (ISO 10993-1).
    • Substantial Equivalence: Comparison to legally marketed predicate devices in terms of intended use, indications for use, material, design, and function. The "truth" is that the new device is as safe and effective as the predicate.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI device that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1