Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192350

    Validate with FDA (Live)

    Date Cleared
    2020-04-30

    (245 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical Non-ablative Fractional Laser Systems (Model: WFB-01) is intended for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue.

    Device Description

    The Medical Non-ablative Fractional Laser Systems (Model: WFB-01) is intended for use in dermatological procedures requiring skin resurfacing and coagulation of soft tissue.

    The Medical Non-ablative Fractional Laser Systems (Model: WFB-01) includes three main parts, Main Console, Connector and Scan Handpiece.

    The main console is used for system control, such as control of user interface, power on/off and other functions, and also used as holder for other components.

    The Connector is used to connect with accessories or power supplier, such as foot switch, pipeline holder, cold air device etc.

    The Scan Handpiece is used to provide laser emission and graphic scanning functions. It provides three scan heads (AccuTip, EffiTip, GrowTip) for treatment areas.

    AI/ML Overview

    This document is a 510(k) Summary for a medical device called "Medical Non-Ablative Fractional Laser Systems (Model: WFB-01)" by Wingderm Electro-Optics Ltd. It states that no clinical tests were conducted. Therefore, there is no information available in the provided text regarding acceptance criteria, device performance, sample sizes, ground truth, expert qualifications, or MRMC studies that would typically be associated with a clinical study.

    The submission is based on substantial equivalence to a predicate device (Lumenis, Ltd.'s M22 And Resurfx Systems, K170060) through non-clinical testing.

    Here's a breakdown of the information that is available, specifically highlighting the absence of clinical study data:

    1. A table of acceptance criteria and the reported device performance

    Since no clinical study was performed, there's no table of clinical acceptance criteria or reported clinical device performance based on human subjects. The table provided (Table 2 Performance Comparison) focuses on technical specifications and shows the proposed device's specifications are "SAME" or "Analysis" (meaning there's a difference but it's deemed not to affect safety/effectiveness) compared to the predicate device.

    ITEMProposed DevicePredicate DeviceK170060Remark
    Wavelengths1550 nm1565 nmSAME
    FluenceUp to 70 mJ per Micro-beamUp to 70 mJ per Micro-beamSAME
    Laser Power15 W15 WSAME
    Dot Density(Dot/cm²)500Up to 500SAME
    Laser BeamDiameter110 um110 umSAME
    Inter-beamspacing0.4 mm - 1.0 mm0.4 mm - 1.0 mmSAME
    Beam diametervariabilityFixedFixedSAME
    Pulse Duration10 ms10 msSAME
    Type of laserEr:GlassEr:GlassSAME
    Tip treatmentAreaAccuTip: 10mmX10mm,EffiTip: 20mmX20mmGrowTip: 10mmX20mm18mm SapphireCool Tip18mm Precision TipAnalysis
    Scanning shapesRectangle, Hexagon, Ellipse, Triangle,CircleLine, square, rectangle, circle, donut,hexagon, vertical line, and vertical rectangleAnalysis

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable, as no clinical study was conducted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical study was conducted with a test set requiring expert ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical study was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a laser system, not an AI-assisted diagnostic tool. No MRMC study was mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a laser system, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no clinical study was conducted. The "ground truth" for demonstrating substantial equivalence relied on non-clinical engineering and performance testing against recognized standards and comparison to a predicate device's specifications.

    8. The sample size for the training set

    Not applicable, as no clinical study was conducted or AI algorithm trained.

    9. How the ground truth for the training set was established

    Not applicable, as no clinical study was conducted or AI algorithm trained.

    Summary based on the provided document:

    The device, "Medical Non-Ablative Fractional Laser Systems (Model: WFB-01)," obtained 510(k) clearance by demonstrating substantial equivalence to a predicate device based on non-clinical testing. This included adherence to international standards for medical electrical equipment safety, electromagnetic compatibility, laser safety, and biocompatibility, as well as performance testing for energy output accuracy. No clinical studies were performed, and therefore, no clinical performance data, patient sample sizes, expert-established ground truth, or MRMC studies are discussed in this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1