Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192347

    Validate with FDA (Live)

    Date Cleared
    2019-11-14

    (77 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ST Internal Implant System is intended to be surgically placed in the maxillary or mandibular arches for the purpose providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals. It is used to restore a patient's chewing function. Smaller implants (less than 6.0 mm) are dedicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Larger implants are dedicated for the molar region and are indicated for delayed loading.

    Device Description

    The ST Internal Fixture is a substructure of a dental implant system made of CP Ti Grade 4 with the surface treated by SLA method. The abutment is a superstructure of a dental implant system which is attached to the implants made of Ti-6A1-4V ELI with the surface treated by anodizing method(except healing and temporary abutment). The ST Internal Implant System is intended to be placed in the upper or lower jaw to support prosthetic devices, such as artificial teeth, and to restore a patient's chewing function. The system is consisted of the following components: ST Internal Fixture, EZ Post Abutment, Solid Abutment, Angled Abutment, Milling Abutment, Cover Screw, Healing Abutment, Temporary Abutment, and Abutment Screw.

    AI/ML Overview

    The provided text is a 510(k) summary for the ST Internal Implant System. It does not describe a study to prove the device meets acceptance criteria in the traditional sense of a clinical trial or performance study with defined acceptance criteria for accuracy, sensitivity, specificity, etc.

    Instead, this document focuses on demonstrating substantial equivalence to predicate devices. Substantial equivalence for a medical device is established by showing that the new device (subject device) is as safe and effective as a legally marketed device (predicate device) that is already cleared by the FDA. This is typically achieved by demonstrating that the new device has the same intended use and the same technological characteristics as the predicate, or if there are differences, that those differences do not raise new questions of safety or effectiveness.

    The "acceptance criteria" here are not performance metrics like sensitivity or specificity. Rather, the implicit acceptance criteria are that the differences between the subject device and the predicate(s) are minor and do not negatively impact safety or effectiveness, as demonstrated through non-clinical (bench) testing.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of (implicit) acceptance criteria and the reported device performance:

    Implicit Acceptance Criteria (for Substantial Equivalence)Reported Device Performance
    Biocompatibility: Same material composition, manufacturing process, and patient-contacting parts as predicate.The subject device (ST Internal Implant System) has the same material composition, manufacturing process, and patient-contacting parts as the predicate device (AnyRidge Octa 1 Implant System, K182448) for fixtures (ASTM F67) and abutments (ASTM F136). Therefore, no additional biocompatibility testing was required.Implied Pass
    Surface Treatment: Same surface treatment and manufacturing process as predicate.The ST Internal Implant System has the same surface treatment (S.L.A.) and manufacturing process as the predicate device (AnyRidge Octa 1 Implant System, K182448) for fixtures. Evaluated in accordance with 'Section 11 of Class II Special Controls Guidance Document'. Implied Pass
    Pyrogen and Endotoxin: Meet regulatory requirements.Will be conducted on every batch in accordance with 'Guidance for Industry Pyrogen and Endotoxins Testing: Questions and Answers'. (This is a commitment, not a prior test result for this submission).
    Sterilization Validation & Shelf Life: Achieve sterility assurance level (SAL 10^-6) and validate proposed shelf life.Sterilization validating testing performed according to ISO 11137 and ISO 17665-1, 2, verifying SAL 10^-6. Shelf-life tests conducted using accelerated aging (ASTM F1980) validated a 5-year shelf life. Implied Pass
    Performance (Physical Properties): Withstand fatigue and loading under worst-case conditions.Bench tests (static compression-strength test and fatigue test) performed in accordance with ISO 14801 and Class II Special Controls Guidance Document. Test results show the subject device is "better able to withstand the fatigue and loading" compared to the predicate device, despite minor size and feature differences. Implied Pass

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: The document does not specify exact sample sizes (number of units tested) for the non-clinical bench tests (static compression-strength and fatigue tests). It mentions that tests were performed "with combination of the fixture and the angled abutment, in consideration of the worst case" and refers to ISO 14801. ISO 14801 typically involves testing a certain number of samples (e.g., n=10 or more) for fatigue and static strength for statistical validity, but the exact count is not provided in this summary.
    • Data Provenance: The tests were performed by MegaGen Implant Co., Ltd. which is based in Republic of Korea. The data is non-clinical bench test data, not human subject data, so "retrospective or prospective" doesn't directly apply in the same way it would for a clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This information is not applicable as there was no study involving human expert interpretation or establishing ground truth based on expert consensus. The "ground truth" for non-clinical tests like fatigue and strength is based on physical scientific principles and engineering standards (e.g., ISO 14801).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not applicable as there was no study involving human interpretation or adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This device is a dental implant system (physical product), not an AI diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • No, a standalone algorithm performance study was not done. This device is a physical dental implant system, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • For the non-clinical tests, the "ground truth" is defined by adherence to international standards and guidance documents like ISO 14801 and specific FDA guidance for root-form endosseous dental implants. This involves objective physical measurements (e.g., fatigue limit, static compression strength values) verified against predefined acceptable ranges or performance relative to a predicate.

    8. The sample size for the training set

    • This information is not applicable. The context is a 510(k) submission for a physical medical device (dental implant), not an AI/ML algorithm that requires training data.

    9. How the ground truth for the training set was established

    • This information is not applicable for the reasons stated above.