Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192208

    Validate with FDA (Live)

    Date Cleared
    2019-10-09

    (56 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    18 - 65
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LEGEND Spine CORNICE Cervical Spacer System is indicated for use as an intervertebral body fusion device in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at one or two contiguous levels from C2 to T1. These patients should have had six weeks of non-operative treatment. The LEGEND Spine CORNICE Cervical Spacer System is also to be used with supplemental fixation system that have been cleared for use in the cervical spine. The LEGEND Spine CORNICE Cervical Spacer System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.

    Device Description

    The LEGEND Spine CORNICE Cervical Spacer System is an Anterior Cervical Interbody Fusion Device that has a hollow chamber to permit packing with bone graft to facilitate fusion. The superior and inferior surfaces of the device have a pattern of teeth to provide increased stability and to help prevent movement of the device. The Legend Spine CORNICE Spacers are made from either PEEK (VESTAKEEP i4 R per ASTM F2026) radiolucent material with embedded tantalum (ASTM F560) x-ray markers or Titanium Alloy conforming to ASTM F136.

    AI/ML Overview

    The provided document is a 510(k) premarket notification letter and summary for the LEGEND Spine CORNICE Cervical Spacer System. This document describes a medical device, which is a physical implant, not an AI/ML powered device. As such, it does not include information typically sought for AI/ML devices regarding acceptance criteria, performance studies involving AI, human experts, and ground truth establishment in the way an AI/ML product would.

    The performance testing described is mechanical testing for the physical implant to ensure its structural integrity and safety. Therefore, I cannot extract the information required for an AI/ML device from this document.

    For completeness, I can describe the type of acceptance criteria and performance testing that were included for this physical device:

    Acceptance Criteria and Reported Device Performance (Physical Device - Mechanical Testing)

    Acceptance Criteria TypePerformance Test/Standard
    Structural IntegrityStatic and dynamic axial compression
    Stability/Resistance to ShearStatic and dynamic compression shear
    Resistance to DislodgementExpulsion testing
    Resistance to SinkageSubsidence testing
    ResultThe device meets or exceeds the performance of predicate devices according to ASTM F2077-14, ASTM F2267-04, and the Class II Special Controls Guidance Document.

    Missing Information (as this is not an AI/ML device):

    • Sample size used for the test set (no test set in the AI/ML sense)
    • Data provenance (not applicable for mechanical testing)
    • Number of experts used to establish ground truth (not applicable)
    • Qualifications of those experts (not applicable)
    • Adjudication method (not applicable)
    • Multi-reader multi-case (MRMC) comparative effectiveness study (not applicable)
    • Effect size (not applicable)
    • Standalone performance (not applicable)
    • Type of ground truth (not applicable)
    • Sample size for the training set (not applicable)
    • How ground truth for the training set was established (not applicable)

    The document focuses on demonstrating substantial equivalence to predicate devices through mechanical testing, which is standard for intervertebral body fusion devices. It does not involve any AI/ML components where ground truth and human-in-the-loop performance would be relevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1