Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192207

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-11-22

    (100 days)

    Product Code
    Regulation Number
    882.5600
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should start within 6 hours of symptom onset.
    2. The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
    3. The Trevo Retriever is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artely (ICA) or middle cerebral artely (MCA)-Ml segments with smaller core infarcts (0-50 cc for age < 80 years, 0-20 cc for age ≥ 80 years). Endovascular therapy with the device should start within 6-24 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
    Device Description

    The Trevo Retriever consists of a flexible, tapered core wire with a shaped section at the distal end. Platinum markers at the distal end allow fluoroscopic visualization. In addition, the shaped section is also radiopaque. Retriever dimensions are indicated on product label. The Retriever delivery wire has a hydrophilic coating on the distal 101cm length to reduce friction during use. The Retriever has a shaft marker to indicate proximity of Retriever tip relative to Microcatheter tip. A Torque Device is provided with the Retriever to facilitate manipulation and retrieval. The Retriever comes preloaded in an insertion tool to introduce the Retriever into a Microcatheter.

    AI/ML Overview

    The provided document is a 510(k) summary for the Trevo NXT ProVue Retriever, a neurovascular mechanical thrombectomy device. It describes the device, its intended use, and a comparison to predicate devices, primarily focusing on non-clinical performance and biocompatibility data to establish substantial equivalence.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and the Reported Device Performance:

    The document provides a table of various performance tests, their methods, and conclusions. The acceptance criteria themselves are explicitly stated for biocompatibility tests, but for performance tests, the conclusion is generally "meets acceptance criteria," implying that specific thresholds were defined internally and met.

    TestAcceptance CriteriaReported Device Performance
    Performance Testing
    Dimensional Verification(Implied specific measurements and tolerances)Dimensional verification meets acceptance criteria.
    Retriever Delivery Wire Kink Resistance(Implied: no damage after three retrieval attempts in tortuous model per IFU)Retriever Delivery Wire Kink Resistance meets acceptance criteria.
    Tip Flexibility(Implied: device withstands applied peak compression/flex force)Tip Flexibility meets acceptance criteria.
    Retriever Shaped Section Radial Force(Implied: meets specified force at specified diameters)Retriever Shaped Section Radial Force meets acceptance criteria.
    Corrosion ResistanceNo signs of corrosion after immersion in saline bath per EN ISO 10555-1.Corrosion Resistance meets acceptance criteria.
    Particulate CharacterizationAcceptable number of particulates in specified size ranges (≥10µm, ≥25µm, ≥50µm, ≥100µm; and ≥200µm, ≥500µm, ≥1000µm if >100µm observed).Particulate characterization was acceptable.
    Coating Integrity Characterization(Implied: no significant coating damage after simulated use in tortuous model)Coating Integrity was acceptable.
    ISO FractureNo fractures after winding around a cylindrical former per EN ISO 11070, Annex F.ISO Fracture meets acceptance criteria.
    ISO FlexureNo damage or flaking of coating after 20 cycles of reverse bending and straightening per EN ISO 11070, Annex G.ISO Flexure meets acceptance criteria.
    Reloadability into Insertion Tool(Implied: no damage after three retrieval attempts in tortuous model per IFU)Reloadability into Insertion Tool meets acceptance criteria.
    Coating Lubricity/Durability(Implied: average frictional force at 6th cycle is within acceptable limits)Coating Lubricity/Durability meets acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Retriever / Microcatheter Deliverability (Track Test-First Push)(Implied: maximum force from first 5cm push is within acceptable limits)Retriever/Microcatheter Deliverability meets acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Retriever In-Vivo Resheathability Into Microcatheter(Implied: no damage after three retrieval attempts in tortuous model per IFU)Retriever In-Vivo Resheathability into Microcatheter meets acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Retractability of Retriever into BGC(Implied: no damage after three retrieval attempts in tortuous model per IFU)Retractability of Retriever into BGC meets acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Retrievability of Retriever with Intermediate Catheter(Implied: no damage after three retrieval attempts in tortuous model per IFU)Retractability of Retriever with Intermediate Catheter meets acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Product Integrity Post RemovalNo kinks inspected on delivery wire or retriever after removal from packaging per IFU.Product Integrity Post Removal meets acceptance criteria. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Stent Containment in Insertion ToolProper position of stent retriever within insertion tool after removal from packaging per IFU.Stent Containment in Insertion Tool meets acceptance criteria.
    Design Validation(Implied: device performance successful in neurovascular model under simulated clinical conditions)Design Validation testing met acceptance criteria. All Users observed that the Trevo NXT ProVue Retrievers successfully met all the design attribute evaluations defined in the design validation protocol. No additional risks identified. Adequately met customer needs.
    Biocompatibility Testing
    ISO 10993-4: HemolysisHemolysis is ≤ 5%.Acceptance criteria met.
    ISO 10993-4: ThrombosisThrombosis is acceptable for clinical application.Acceptance criteria met.
    ISO 10993-5: CytotoxicityNot less than or equal to 50% cell viability.Acceptance criteria met.
    ISO 10993-10: SensitizationNot a sensitizer.Acceptance criteria met.
    ISO 10993-10: IrritationNo significant irritation.Acceptance criteria met.
    ISO 10993-11: Material Mediated PyrogenicityNo febrile reaction greater than 0.5°C.Acceptance criteria met. --------------------------------------------------------------------------------------------------------------------------
    ISO 10993-11: Acute Systemic ToxicityNo mortality or evidence of systemic toxicity.Acceptance criteria met. --------------------------------------------------------------------------------------------------------------------------
    ISO 10993-18: Physicochemical (USP <661>)Non-volatile Residue ≤ 15 mg; Residue on Ignition ≤ 5 mg; Heavy Metals ≤ 1 ppm; Buffering Capacity ≤ 10.0 mL.Acceptance criteria met. --------------------------------------------------------------------------------------------------------------------------
    FTIRNo unexpected readings.Acceptance criteria met. --------------------------------------------------------------------------------------------------------------------------
    Latex (LEAP - ASTM D6400)No detectable latex.Acceptance criteria met. --------------------------------------------------------------------------------------------------------------------------
    Sterilization Assurance LevelSAL of 10⁻⁶.Achieved.
    EO ResidualsMet per EN ISO 10993-7 for limited contacting, externally communicated devices.Achieved.
    Shelf LifePackaging remains functional and maintains sterility for 2 years.Met all acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: Not explicitly mentioned for each specific performance test. The design validation states "All Users observed that the Trevo NXT ProVue Retrievers successfully met all the design attribute evaluations...". For biocompatibility, standard ISO tests were performed, which usually involve specific numbers of samples or animals as per the standard.
    • Data Provenance: The studies are non-clinical (bench testing and in vitro simulated-use studies) and biocompatibility tests. They are not clinical studies involving human patient data, so concepts like country of origin or retrospective/prospective are not applicable in this context. The simulated-use studies were conducted internally for device verification and validation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not explicitly stated. The design validation study mentions "Users evaluated the device in a simulated use tortuous anatomical model". It doesn't specify if these "Users" are experts in establishing ground truth, nor their number or qualifications as one would typically see in a clinical study where expert consensus might establish ground truth for a diagnostic device. In this context of a mechanical device, performance is evaluated against defined physical and functional criteria rather than expert interpretation.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable as this is a device performance study focusing on physical and functional attributes, not a study requiring adjudication of clinical outcomes or interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a non-clinical evaluation of a mechanical thrombectomy device, not an AI-assisted diagnostic or decision support system.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a mechanical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For performance testing, the "ground truth" (or reference standard) is based on defined engineering specifications, industry standards (e.g., ISO, ASTM), and predetermined functional requirements. For example, a successful test means the device's dimension falls within a specified range, or it withstands a certain force, or it shows no damage after simulated use.
    • For biocompatibility testing, the ground truth is established by the criteria outlined in recognized international standards (ISO 10993-x), which define acceptable biological responses (e.g., hemolysis ≤ 5%, not a sensitizer).

    8. The sample size for the training set:

    • Not applicable. This is a non-clinical evaluation of a mechanical thrombectomy device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1