    Intended Use

    The Sterilucent SCBI is intended to be used as a standard method for frequent monitoring of the Sterilucent HC 80TT Sterilizer Cycles, when incorporated into a test pack.

    The Sterilucent Process Challenge Device (PCD) test pack is used for routine monitoring of the Sterilucent HC 80TT Sterilizer Lumen and Flexible Cycles. The Sterilucent PCD may also be used for performance qualification of the Sterilucent HC 80TT sterilizer Lumen and Flexible Cycles during initial installation, after relocation, major repairs or mal functions, or after sterilization process failures.

    Both the Sterilucent Lumen Cycle PCD and the Sterilucent Flexible Cycle PCD are intended to have greater resistance than the stand alone SCBI. Both devices are designed to have increased resistance beyond the sterilization half-cycle, but complete inactivation upon exposure to the full cycle.

    Device Description

    The Sterilucent Self-Contained Biological Indicator (SCBI) is a sterilization process indicator that conforms to ANSI/AAMI/ISO 11138-1, and is intended for use only with the Sterilucent HC 80TT Hydrogen Peroxide Sterilizer. The HC 80TT Hydrogen Peroxide Sterilizer has two pre-programmed sterilization cycles ("Lumen" and “Flexible") which both utilize vaporized hydrogen peroxide (VHP) to rapidly sterilize a variety of reusable medical devices.

    The Sterilucent SCBI is to be used in a test pack configuration, with a different test pack (Process Challenge Device) used for each sterilization cycle. Each Process Challenge Device (PCD) is designed to have greater resistance than the worst-case sterilization load, allowing for survival beyond the sterilization half-cycle, but complete inactivation upon exposure to the full cycle.

    The Sterilucent SCBI provides information on whether necessary conditions were met to kill a specified number of microorganisms upon exposure to either cycle of the Sterilucent HC 80TT Hydrogen Peroxide Sterilizer.

    AI/ML Overview

    The document describes the acceptance criteria and study results for the Sterilucent Self-Contained Biological Indicator (SCBI) and Process Challenge Device (PCD).

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    SCBI Resistance Evaluation (D-Value & Survival/Kill Window)Meaningful valuePassed
    SCBI Holding Time Assessment (24 hours)Meaningful valuePassed
    SCBI Reduced Incubation Time (24 hours)Value less than seven (7) days based on FDA RIT protocolPassed
    SCBI Growth Inhibition (none allowed)Spore growthPassed
    SCBI Population Stability & 12-month Shelf LifeSpore population remains between 1.0-4.0 x 10^6 throughoutPassed
    SCBI Integrated Chemical Indicator VerificationISO 11140-1 Section 8.7, Table 6Passed
    PCD Resistance CharacterizationGreater resistance as demonstrated by fractional killPassed
    PCD FunctionalityFractional kill at half-cycle; all kill at full cyclePassed

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each individual test. The provenance of the data is not explicitly stated in terms of country of origin, but the submission is for the U.S. Food and Drug Administration (FDA), implying the testing was conducted to meet U.S. regulatory standards. The testing appears to be prospective as it was conducted to demonstrate the functionality of the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. The device is a biological indicator and process challenge device for sterilization. Its performance is evaluated through biological and physical tests based on established standards, not through expert interpretation of data like in diagnostic imaging.

    4. Adjudication method for the test set

    This information is not applicable for this type of device and testing. The evaluation relies on quantitative measurements and adherence to predefined performance metrics rather than expert consensus on subjective outcomes.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This device is a sterilization indicator, not an AI-powered diagnostic tool, so MRMC studies involving human readers and AI are irrelevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device itself is non-algorithmic and does not involve AI or human-in-the-loop performance in the conventional sense of software-as-a-medical-device. Its performance is inherent to its design and material properties.

    7. The type of ground truth used

    The ground truth for the SCBI and PCD testing is established by recognized international standards and methodologies for sterilization indicators. This includes:

    • Biological inactivation: The absence or presence of viable Geobacillus stearothermophilus spores after exposure to sterilization conditions.
    • Physical/chemical parameters: Measurements of D-value, survival/kill windows, spore population counts, pH changes, and color changes in chemical indicators.
    • Compliance with standards: Adherence to ANSI/AAMI/ISO 11138-1, ANSI/AAMI/ISO 11140-1, and FDA Guidance documents.

    8. The sample size for the training set

    This information is not applicable. The Sterilucent SCBI and PCD are not AI/ML-based devices; therefore, there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    This information is not applicable as the device does not utilize a training set in the AI/ML sense.