Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K191402

    Validate with FDA (Live)

    Date Cleared
    2020-02-25

    (277 days)

    Product Code
    Regulation Number
    870.4250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Heater-Cooler System 3T is used to circulate water through heat exchangers to warm or cool a patient during cardiopulmonary bypass procedures lasting 6 hours or less.

    Device Description

    The 3T System is an independent (i.e., independent of the water supply) 3-circuit heating/cooling base unit that includes three water circuits (two circuits for the patient supplied by one tank (for the heating/cooling blanket, the oxygenator) and one interchangeable heating/cooling circuit for cardioplegia. If required, patient and cardioplegia circuits can be switched off separately, in order to increase the activated functional group's heating and/or cooling performance.

    The following optional components and accessories are also available for the 3T System:

      1. Heating-cooling blankets;
      1. Water circuit tubing; and
      1. Various cables.
        The 3T System includes a disposable aerosol collection set that consists of a canister connected to the 3T System and the user facility's vacuum source that captures emissions that are drawn from the tank using negative pressure created by the user facility's vacuum.
    AI/ML Overview

    The Heater-Cooler System 3T is used to circulate water through heat exchangers to warm or cool a patient during cardiopulmonary bypass procedures lasting 6 hours or less.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Performance AspectReported Device Performance
    Reduction in Emission (Aerosolized Mycobacteria)- Microbiological and orthogonal assay testing during all phases of device operation to evaluate device emissions verified reduction. - Microbiological testing verified the hydrophobic filter of the disposable aerosol collection set's canister adequately prevents contamination of the user facility's vacuum source. - Microbiological and orthogonal assay testing verified the seven-day use period of the disposable aerosol collection set.
    Aerosol Collection Set Integrity and EfficacyTesting verified the chemical resistance of materials, integrity, and efficacy of the disposable Aerosol Collection Set, as well as the integrity of its packaging design.
    Heating and Cooling FunctionsTesting verified the performance of the heating and cooling functions of the 3T System using a patient simulator circuit.
    Corrosion and Erosion ResistanceTesting evaluated the chemical resistance of the metallic and plastic components and materials of the 3T System.
    Cleaning, Disinfection, and Preservation of Water Circuit (Intermediate Level)Validation testing demonstrated an intermediate level of disinfection. Long-term testing (58 weeks) verified that the instructions adequately prevent contamination of the device.
    Cleaning and Disinfection of 3T System SurfacesValidation testing of the 3T System surface cleaning instructions was performed.
    Transportation and Storage PackagingTesting of the transportation and shipping packaging of the 3T System was performed.
    Production Disinfection and Drying Process QualificationProcess validation testing of the production disinfection and drying processes was conducted.
    Human FactorsHuman factors validation testing of the 3T System Operating Instructions was performed.
    Substantial Equivalence to Predicate Device (Stöckert Heater-Cooler System 3T)The Final Configuration 3T System has very similar indications for use, the same fundamental scientific technology, and similar technological characteristics. Performance and validation testing demonstrate that the modified 3T System is as safe and effective as the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not explicitly state the specific, quantitative sample sizes for each of the performance tests. It describes the types of testing performed (e.g., microbiological, orthogonal assay testing, long-term testing for 58 weeks).

    The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective/prospective. The testing appears to be laboratory-based performance testing conducted by the company.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The studies appear to be laboratory performance tests rather than clinical studies requiring expert ground truth for interpretation. For example, microbiological testing determines contamination levels, and heating/cooling tests measure temperature changes against predefined specifications.

    4. Adjudication Method for the Test Set

    This information is not applicable and thus not provided. Since the tests are described as performance and validation testing (e.g., measuring emissions, verifying filter efficacy, testing heating/cooling functions), there isn't a "ground truth" established by multiple experts that would require an adjudication method. The results are likely binary (pass/fail) or quantitative measurements compared against specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not conducted. This device is a medical apparatus (heater-cooler system) and not an AI-powered diagnostic or assistive technology that would involve human readers interpreting cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This concept is not applicable to the Heater-Cooler System 3T. It is a physical medical device, not an algorithm, so standalone algorithm performance is not relevant.

    7. The Type of Ground Truth Used

    The "ground truth" for the various performance tests appears to be based on:

    • Established industry standards/specifications: For performance aspects like heating/cooling efficiency, corrosion resistance, transportation, and surface cleaning.
    • Microbiological control principles: For assessing the reduction in emissions, filter efficacy, and prevention of contamination. This involves measuring microbial counts against acceptable limits.
    • Chemical and material science principles: For testing chemical resistance and material integrity.
    • FDA guidance: Specifically referenced for "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" for cleaning, disinfection, and water preservation.

    8. The Sample Size for the Training Set

    This device does not involve a "training set" in the context of machine learning or AI. The performance testing focuses on the physical and functional aspects of the hardware and associated procedures.

    9. How the Ground Truth for the Training Set was Established

    As there is no training set for this device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1