510(k) Data Aggregation

    K Number
    K191385

    Validate with FDA (Live)

    Date Cleared
    2019-10-29

    (159 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Clinical indications

    • Restoration of caries defects
    • Reconstruction of lost tooth substance due to attrition, abrasion, erosion
    • Restoration of tooth fractures
    • Correction of anatomical malformation

    Cosmetic indications

    • Optimization of old composite restorations
    • Extending incisors
    • Correction of malpositioned teeth
    • Cosmetic corrections for tooth discoloration or incorrect shading
    • Closure of diastema
    Device Description

    The subject device BRILLIANT COMPONEER is a polymerized, highly filled, submicron hybrid composite enamel shell(s) for direct veneering. It simplifies the direct restoration of front teeth and premolars without having to model the basic shape of the top enamel layer. The thin, anatomical composite shells are available in different shades and shapes allowing a natural, high aesthetic restoration.

    AI/ML Overview

    The provided text describes a medical device called "BRILLIANT COMPONEER," which is a tooth shade resin material. The FDA 510(k) premarket notification K191385 outlines the substantial equivalence of this device to a predicate device, COMPONEER.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for performance are derived from the predicate device's performance and the general standard for this type of material (ISO 4049). The device is deemed substantially equivalent if its performance is comparable.

    AttributeAcceptance Criteria (Predicate Device K112168 - COMPONEER)Reported Device Performance (Subject Device K191385 - BRILLIANT COMPONEER)
    Compressive Strength392 MPa285 MPa
    Flexural Strength127 MPa125.4 MPa
    Elastic Modulus9.0 GPa9.8 GPa
    Radiopacity2 mm of Al3.24 mm of Al
    Water sorption16 µg / mm³18.2 µg / mm³
    Solubility0.9 µg / mm³0.3 µg / mm³
    BiocompatibilityConforms with ISO 10993-1Conforms with ISO 10993-1
    Performance (General Standard)Conforms with ISO 4049Conforms with ISO 4049

    Note: While some values are different, the submission asserts substantial equivalence based on extensive performance and biocompatibility testing according to applicable standards.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each non-clinical performance test. It mentions that extensive testing was performed. The data provenance is stated as testing performed by Coltène/Whaledent AG in Switzerland. The studies are retrospective in the sense that they are laboratory bench tests conducted to demonstrate compliance with standards and comparison to a predicate, not prospective clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The ground truth for the performance characteristics is established by objective measurements based on international standards (e.g., ISO 4049). For biocompatibility, it's based on objective assays and assessments according to ISO 10993 and ISO 7405. There is no indication of human expert interpretation being required to establish the "ground truth" for the test set results.

    4. Adjudication method for the test set

    This information is not applicable as the document describes non-clinical performance testing (bench tests and biocompatibility tests), which do not typically involve human adjudication in the way clinical studies or expert consensus for image interpretation would. The results are based on objective measurements and adherence to specified procedures within the standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a dental material, not an AI-assisted diagnostic tool or an imaging device that would typically involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical dental material, not an algorithm or software. The performance tests are standalone in the sense that they measure the material's inherent properties, without human interpretation in the loop.

    7. The type of ground truth used

    The ground truth used for the performance tests consists of objective measurements against established international standards. For chemical and physical properties (e.g., compressive strength, water sorption), the ground truth is the measured value itself, compared to the predicate and standard requirements. For biocompatibility, the truth is whether the material elicits a specific biological response based on standardized assays.

    8. The sample size for the training set

    Not applicable. This device is a physical dental material, not a machine learning model. Therefore, there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for a machine learning model, this question is not relevant to the described device.