510(k) Data Aggregation

    K Number
    K191148

    Validate with FDA (Live)

    Date Cleared
    2019-09-12

    (135 days)

    Product Code
    Regulation Number
    888.3027
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used without cement, CD Horizon™ Fenestrated Screws (with or without Sextant™ or Longitude™ instrumentation) are intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (DDD - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, tumor and/or trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis), pseudarthrosis, and/or failed previous fusion.

    When used in conjunction with Medtronic HV-R™ Fenestrated Screw Cement or Kyphon™ Xpede™ Bone Coment, CD Horizon™ Fenestrated Screws are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic, lumbar, or sacral spine in whom life expectancy is of insufficient duration to permit achievement of fusion™ Fenestrated Screws augmented with Medtronic HV-R™ Fenestrated Screw Cement or Kyphon™ Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.

    Kyphon™ Xpede™ Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions a cementoplasty (i.e. kyphoplasty or vertebroplasty) procedure. It is also indicated for the fixation of pathological fractures of the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty. Cancer includes multiple myeloma and metastatic lesions, including those arrsing from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and giant cell tumor. Pathological fracture may include a symptomatic microfracture (as documented by appropriate imaging and/or presence of a lytic lesion) without obvious loss of vertebral body height.

    When used in conjunction with CD Horizon™ Fenestrated Screws, Kyphon™ Xpede™ Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic, lumbar, or sacral spine in whom life expectancy is of insufficient duration to permit achievement of fusion. CD Horizon™ Fenestrated Screws augmented with Kyphon™ Xpede™ Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.

    When used in conjunction with CD Horizon™ Fenestrated Screws, Medtronic HV-R™ Fenestrated Screw Cement is intended to restore the integrity of the spinal column, even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic, lumbar, or sacral spine in whom life expectancy is of insufficient duration to permit achievement of fusion. Medronic HV-R™ Fenestrated Screw Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.

    Device Description

    The CD Horizon™ Fenestrated Screw Set consists of a variety of cannulated screws. These screws contain a series of fenestrations which allows polymethylmethacrylate (PMMA) bone cement (Medtronic HV-R™ Fenestrated Screw Cement or Kyphon™ Xpede™ Bone Cement) to be injected into the treated site. This cement is used to augment screw fixation into the pedicle in patients whose life expectancy is of insufficient duration to permit achievement of fusion. These implants may also serve as traditional pedicle screws when used without bone cement in patients with non-compromised bone quality. CD Horizon™ Fenestrated Screws are specifically designed to connect to appropriate rods and associated connecting components contained within the CD Horizon™ Spinal System. CD Horizon™ Fenestrated Screw Set implant components are fabricated from medical grade titanium and/or medical grade titanium alloy and/or medical grade cobalt-chromiummolybdenum allov.

    Kyphon™ Xpede™ Bone Cement is a polymethylmethacrylate (PMMA) that contains approximately 30% barium sulfate. It is designed for delivery in a highly viscous state. Kyphon™ Xpede™ Bone Cement is provided sterile in two components: 20 grams of powder and nine grams of liquid. The powder contains methylmethacrylate-styrene copolymer, barium sulfate and benzoyl peroxide. The liquid contains methylmethacrylate (monomer), hydroquinone and N, N dimethyl-p-toluidine.

    Medtronic HV-RTM Fenestrated Screw Cement is a polymethylmethacrylate (PMMA) that contains approximately 30% barium sulfate. It is designed for delivery in a highly viscous state. Medtronic HV-RTM Fenestrated Screw Cement is provided sterile in two components: 20 grams of powder and nine grams of liquid. The powder contains methylmethacrylate-styrene co-polymer, barium sulfate, and benzovl peroxide. The liguid contains methylmethacrylate (monomer), hydroquinone and N. N dimethyl-p-toluidine.

    AI/ML Overview

    This FDA 510(k) summary does not contain any information regarding clinical performance studies, AI algorithms, or human-in-the-loop performance.

    The document is a premarket notification for medical devices (bone cement and fenestrated screws), not an AI/ML-driven diagnostic device. The application focuses on demonstrating substantial equivalence to previously cleared predicate devices based on technological characteristics and existing performance data for those predicates.

    Therefore, I cannot provide the requested information about acceptance criteria, reported device performance, sample sizes for test/training sets, expert qualifications, adjudication methods, or MRMC studies, as these types of studies were not conducted or reported in this submission for these specific medical devices.

    The document explicitly states:

    "Medtronic believes that testing is not warranted for the subject devices as they do not present a new worst case when compared to the predicate devices. All existing predicate data previously provided in the predicate 510(k)s are still applicable."

    This indicates that the submission relies on the performance data of the predicate devices rather than new performance studies for the subject devices themselves.