510(k) Data Aggregation

    K Number
    K191109

    Validate with FDA (Live)

    Device Name
    Spectra Q
    Manufacturer
    Date Cleared
    2020-01-22

    (271 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spectra Q is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Spectra Q is intended for home use by a single user.

    Device Description

    The Spectra Q breast pump is comprised of a Spectra Q pump, breast shield (20, 24, 28, 32 mm), backflow protector, silicone valve, tubing, bottles (including the nipple, cap, cover, and disk), 2-wav connector (for double pumping), and power adapters. The breast pump is provided with one 24 mm and one 28 mm breast shield. Optional components include a small cap converter, assembled with the feeding bottle, the 20 and 32 mm breast shields, and a wide bottle stand. The Spectra Q breast pump is used by lactating women to express and collect milk from their breasts. The pump is intended for home use by a single user. Pumping with the Spectra Q breast pump can be performed on one breast (single pumping) or on both breasts at the same time (double pumping). The Spectra O breast pump allows the user to adjust the vacuum levels. Two suction patterns, massage and expression mode, are pre-programmed with variable vacuum levels and cycle rates (pump speed). The powered breast pump is capable of providing vacuum levels from -50 to -270 mmHq with cycle rates up to 110 cycles per minute. The device is powered by a 6 V AC/DC power adapter or a 5 V USB C type power adapter. The device is provided non-sterile.

    AI/ML Overview

    The provided text is related to the K191109 510(k) clearance for the Spectra Q breast pump. It describes the device, its intended use, and its substantial equivalence to a predicate device (Spectra 9Plus).

    However, the document does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of AI/ML or diagnostic performance. The non-clinical tests mentioned are primarily related to:

    • Electrical Safety/Electromagnetic Compatibility: Compliance with standards like AAMI / ANSI ES60601-1, IEC 60601-1-2, and IEC 60601-1-11.
    • Software Validation: In accordance with FDA Guidance for software in medical devices (May 11, 2005).
    • Biocompatibility: Evaluation based on ISO-10993 and compliance of milk-contacting components with 21 CFR 177.1520, leveraging information from the predicate device.
    • Bench Performance Testing: Confirming minimum/maximum vacuum levels and cycle rates, and demonstrating sustained specifications over its use life.

    These tests are standard for a breast pump and focus on physical and electrical safety and performance parameters, not on the diagnostic accuracy or classification capabilities that would typically require the kind of "acceptance criteria" table and study details (sample size, ground truth, expert consensus, MRMC studies) you've requested.

    Therefore, I cannot fulfill your request for the specific information regarding AI/ML device acceptance criteria and studies from this document. The document does not describe such a device or study.