Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190909

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-02-13

    (311 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for Use include electrosurgical coagulation, dissection, and grasping of tissue during the performance of laparoscopic and general surgical procedures.

    Device Description

    This product is sterile, single-use, invasive instrument that used in laparoscopic surgery. There are two models, one for ABF01-F and the other for lock functions, ABF01-L. The two models are same except for lock function. This product is a specific component, but not the entire electrosurgical device. The device is not intended to be marketed with multiple components, accessories, and as part of a system.

    AI/ML Overview

    The document describes the ArtiSential Bipolar Fenestrated Forceps, a single-use electrosurgical instrument used in laparoscopic surgery for coagulation, dissection, and grasping of tissue. The device underwent several performance tests to demonstrate its safety and effectiveness and establish substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test Clause and Test RequirementAcceptance Criteria (Test Specification)Reported Device Performance (Results - Remarks)
    1. AppearanceNo defects in the appearance; no problem in use.Visual observationNo crack, stain, or no substances on the surface of the product.
    2. DimensionWithin ± 5% of the indicated value.Measure by Vernier calipers and dial gauge etc.Pass (Refer to [Measurement data] on 65 page at attachment #3).
    3. Operational TestJaw smoothly opened/closed, free from jamming. Jaw and hub bend up, down, left, right ≥ ±80°, 360° rotation.Adjusting open and close the jaw by manipulating the grip and control ring; measure angle at bending and turning.The jaw and hub are bent up, down, left and right within above 80°, and can rotate 360°.
    4. Tensile StrengthJaw and shaft connections not damaged from 20 N pull.Hold the jaw and shaft connections and apply a force of 20 N using a Push pull gauge.No damage to the connection when applying a force of 20N.
    5. Feedthrough TestElectricity transmitted between electrode tip and connector.Electrical conduction between electrode tip and connector tested using DMM.The resistance value between the electrode tip and the connector is less than 1Ω.
    BiocompatibilityAccording to ISO 10993-1, assessed for Cytotoxicity, Sensitization, Intracutaneous reactivity, Acute systemic toxicity.All evaluated endpoints passed.
    Electrical SafetyAccording to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-18, and IEC 60601-2-2.The device passed all performed tests (Active accessory insulation, hf leakage, hf dielectric strength, mains frequency dielectric strength, feedthrough test).
    SterilizationEO-Sterilization in accordance with ISO-11135.The device is provided sterile via EO-Sterilization.
    Shelf LifeProposed expiration date is 2 years from manufacturing date (real-time testing for 3 years).Proposed 2-year shelf life. (Real-time testing for 3 years to be performed).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the performance tests. The data provenance is also not specified; however, given the applicant's address (Republic of Korea) and the FDA submission, it is likely that the testing data was generated in support of regulatory approval in the US. The nature of the tests (physical, electrical, and biocompatibility) suggests that they are laboratory-based and generally not dependent on patient data provenance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The performance tests described are objective, engineering-based evaluations of the device's physical and electrical characteristics, and biological safety. They do not involve expert interpretation or ground truth derived from a clinical dataset (e.g., medical images).

    4. Adjudication Method for the Test Set

    Not applicable, as the tests are objective and do not involve human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. The ArtiSential Bipolar Fenestrated Forceps is a medical instrument, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study related to AI assistance is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable, as this is a medical instrument and not an algorithm. Performance tests mentioned are for the physical device itself.

    7. The Type of Ground Truth Used

    The "ground truth" for the tests performed is based on objective, measurable engineering specifications, established international standards (e.g., ISO 10993-1, IEC 60601 series, ISO-11135), and visual inspection criteria. It is not dependent on expert consensus, pathology, or outcomes data in a clinical sense.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1