510(k) Data Aggregation

    K Number
    K190871

    Validate with FDA (Live)

    Date Cleared
    2020-01-17

    (289 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bio-TackS Orthodontic Mini Implant is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only.

    Device Description

    Fixed on jawbone the Bio-TackS Orthodontic Mini Implant is a screw-shaped product used as a fixed point for orthodontic treatment and it is applied in order to secure sufficient physical and mechanical fixing source for treatment.

    AI/ML Overview

    The provided text describes the Bio-TackS Orthodontic Mini Implant and its substantial equivalence to a predicate device, focusing on non-clinical performance and biocompatibility. It explicitly states that clinical testing was not required (Section 9, page 8). Therefore, the information requested about acceptance criteria and studies proving device performance with human data (e.g., sample size for test set, experts for ground truth, MRMC study, standalone performance, ground truth type for test/training sets) is not available within this document.

    The document primarily focuses on demonstrating substantial equivalence through bench testing (non-clinical performance tests) and biocompatibility testing.

    Here's the available information relevant to the acceptance criteria and the study that proves the device meets them, based on non-clinical data:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state quantitative acceptance criteria for the "performance tests" in a table format. Instead, it describes performing tests according to a standard (ASTM F543) and comparing the subject device's results to the predicate device, concluding "similar performance."

    Reported Device Performance (from Section 8, page 6-7):

    TestAcceptance Criteria (Implicit: Similar to Predicate)Reported Device Performance
    Sterilization and Shelf LifeSterilization validation per ISO 11137-1, 11137-2, 11137-3 (VDmax25 method). Shelf life demonstrated by accelerated aging with evaluations for seal peeling (ASTM F88), dye penetration (ASTM F1929), and sterility testing.Sterilization validation performed (ISO 11137-1, -2, -3, VDmax25 method). Shelf-life demonstrated by accelerated aging with seal peeling, dye penetration, and sterility testing.
    Insertion-Removal Torque TestMechanical stability similar to predicate device (evaluated at 1440 degrees at 1 rpm).Both samples (BioTackS BSH1810S and Osstem OSSH1810) demonstrated "almost equivalent torque values" when inserted and removed.
    Rotational Fracture Torque Strength TestMaterial yield strength similar to predicate device."The tapered implant fracture occurred in the same tread section." Both implants made from the same raw material (titanium alloy-ELI grade 23), indicating "material yield strengths of both products are the same." All screw-head types tested, results similar.
    Axial Pull-Out Strength TestingEffectiveness of mechanical design and elimination similar to predicate device."The test data demonstrated similar results due to the similar pitch tread thickness of the samples" (BioTackS BSH1406S and Osstem OSSH1406).
    Biocompatibility - CytotoxicityNon-cytotoxic (per ISO 10993-5)Non-cytotoxic
    Biocompatibility - Acute Systemic ToxicityNon-acute systemic toxicity (per ISO 10993-11)Non-acute systemic toxicity
    Biocompatibility - Intracutaneous (Intradermal) ReactivityNo irritation (per ISO 10993-10)None Irritation
    Biocompatibility - Local Lymph Node Assay (LLNA)No hypersensitivity (per ISO 10993-10)Do not show any hypersensitivity

    2. Sample size used for the test set and the data provenance:

    • Bench Tests (Performance Tests): For each specific test (Insertion-Removal torque, Rotational fracture torque, Axial pull-out), the document mentions testing specific models of the Bio-TackS and Osstem implants (e.g., BSH1810S, OSSH1810 for insertion-removal torque). It does not specify the exact number of samples tested for each condition but implies at least one sample of the "worst-case" or specific diameter/length combinations.
    • Biocompatibility Tests: Sample sizes are not explicitly stated for these tests (e.g., number of cells for cytotoxicity, number of animals for toxicity/reactivity tests).
    • Data Provenance: The tests are explicitly non-clinical bench and laboratory tests. No country of origin for the data is specified, but the applicant (BIO CETEC CO., LTD.) is from the Republic of Korea.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable as no clinical test set requiring human expert ground truth was conducted. The "ground truth" for the non-clinical tests is based on objective measurements against established standards (e.g., ASTM F543, ISO 11137, ISO 10993 series).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable, as no clinical test set requiring adjudication was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable, as no clinical study (MRMC or otherwise) involving human readers or AI assistance was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable, as this device (an orthodontic mini implant) is a physical medical device, not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the non-clinical performance tests, the "ground truth" is measured physical and mechanical properties against established engineering and material science standards (ASTM F543), and for biocompatibility, it is biological response as assessed by ISO 10993 standards and laboratory assays.

    8. The sample size for the training set:

    Not applicable. This is a physical device, and the evaluation did not involve machine learning or AI models with training sets.

    9. How the ground truth for the training set was established:

    Not applicable (as above).