Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190685

    Validate with FDA (Live)

    Date Cleared
    2019-07-26

    (130 days)

    Product Code
    Regulation Number
    890.5500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hair Growth System (Model: KN-8000A) is a prescription use device intended for the prowth in females with androgenic alopecia who have Ludwig-Savin Classifications I-II, and in males with androgenic alopecia who have Norwood Hamilton Classifications IIa-V; and both genders having Fitzpatrick Classification of Skin Phototypes I to IV.

    Device Description

    The Hair Growth System Model: KN-8000A is a transportable, non-invasive, low-level laser device intended to treat Androgenetic Alopecia (Hair Loss) and to promote hair growth in both Males and Females. It produces a red light, diode lasers operating at a 650-nanometer wavelength (maximum output power of each is 5 mW). The device mainly includes a host machine, light source and lifting bracket. The light source applicator was equipped with the access sensor device, it can power off the light source when the patient's scalp deviates from the work area.

    AI/ML Overview

    The provided text is a 510(k) summary for the "Hair Growth System (Model: KN-8000A)" and focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a performance study against specific acceptance criteria for an AI/algorithm-driven device.

    Therefore, the document does not contain the information requested regarding acceptance criteria related to AI/algorithm performance, sample sizes for test sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set details for an AI model.

    The "Test Summary" section in the document only mentions:

    • Electrical safety test (IEC 60601-1 and IEC 60825-1 standards)
    • Electromagnetic compatibility test (IEC 60601-1-2 standard)
    • Software verification and validation test (according to FDA guidance for software in medical devices)

    These tests are related to the safety and fundamental performance of the device's hardware and software, but not to the clinical effectiveness or performance of an AI component in achieving a specific outcome (like hair growth based on algorithmic assessment). The device itself is described as a "low-level laser device" not explicitly an AI-driven system.

    In summary, none of the requested information regarding AI/algorithm acceptance criteria and study details is present in the provided text. The document's purpose is to establish substantial equivalence for a physical device, not to describe the validation of an AI algorithm.

    Ask a Question

    Ask a specific question about this device