Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190597

    Validate with FDA (Live)

    Date Cleared
    2019-06-21

    (106 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BRILLIANT EverGlow Flow Universal and Translucent shades are suitable for:

    • · Direct class V fillings (cervical caries, root erosion, wedge-shaped defects)
    • · Fillings in the anterior tooth region (Class III and IV)
    • · Small fillings of all cavity classes (not occlusion-bearing)
    • · Blocking out of undercuts
    • · Adhesive luting of indirect composite and ceramic restorations in as far as accuracy of fit and light permeability are given
    • · Repairs of direct and indirect composite restorations
    • · Preventive resin restorations
    • · Cavity lining

    BRILLIANT EverGlow Flow opaque shade is also suitable for:

    • · Aesthetic corrections (e.g. deviations in chroma)
    • · Masking of dark areas
    Device Description

    BRILLIANT EverGlow Flow is a flowable radiopaque submicron hybrid composite and available in several shades that allows for high aesthetic restorations with a clear spectrum of shades. It is applied in the increment technique and available in 7 universal, 2 translucent and 3 opaque shades.

    BRILLIANT EverGlow Flow is chemically characterized by its composition, which can be categorized in three groups based on their chemical and functional similarity: resin components (matrix monomers based on methacrylates), fillers, and additives. The largest part of BRILLIANT EverGlow Flow's composition is represented by fillers (>58% w/w), followed by resin components (<40% w/w), and additives (<2% w/w). BRILLIANT EverGlow Flow is, after curing, a crosslinked methacrylate-based composite material with a three-dimensional polymer network, containing embedded filler particles and incorporated additives.

    BRILLIANT EverGlow Flow has long term contact (>30 days) with tissue/bone/dentin.

    BRILLIANT EverGlow Flow is the supplemental flowable, radiopaque submicron hybrid material version of the universal composite BRILLIANT EverGlow. BRILLIANT EverGlow Flow is, due to its lower viscosity, considered an enhancement of BRILLIANT EverGlow for applications with non-easily accessible areas in the oral cavity (e.g. blocking out of undercuts) and/or extended fissure-sealing. BRILLIANT EverGlow Flow combines a low viscous consistency with high stability and thixotropy, allowing a controlled and comfortable application. BRILLIANT EverGlow Flow is available in different shades and gradations of translucency. BRILLIANT EverGlow Flow is delivered with two application needle sizes (standard size [Ø 0.8mm] and fine size [Ø 0.4mm]). The fine size needle is to improve the accessory range and to allow an application with higher precision.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a dental material, BRILLIANT EverGlow Flow, and its substantial equivalence to a predicate device, BRILLIANT EverGlow. Dental materials are not medical devices in the typical sense of AI, so there is no AI acceptance criteria or study that proves AI meets the acceptance criteria.

    However, I can extract information regarding the product's acceptance criteria and the study conducted to prove it meets those criteria:

    1. Table of acceptance criteria and the reported device performance

    CriteriaPredicate Device (BRILLIANT EverGlow)Subject Device (BRILLIANT EverGlow Flow)Unit
    Flexural strength11796MPa
    Compressive strength390415MPa
    Inorganic filler content7437Vol-%
    Radio-opacity2.02.2mm of Al
    Water sorption1523μg/mm³
    Water Solubility0.72.0μg/mm³
    BiocompatibilityConforms with ISO 10993-1:2018Conforms with ISO 10993-1:2018N/A
    PerformanceConforms with ISO 4049:2009Conforms with ISO 4049:2009N/A

    Acceptance Criteria Reasoning: The acceptance criteria are implicitly established by demonstrating substantial equivalence to the predicate device. The subject device's performance values fall within ranges that are considered acceptable for dental composite materials as per relevant ISO standards and comparison to the predicate device. For example, while the flexural strength is lower and water sorption/solubility are higher for the new device compared to the predicate, these differences are presented in the context of showing clinical equivalence and compliance with established standards like ISO 4049:2009 for dental polymer-based restorative materials.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each non-clinical test (e.g., flexural strength, compressive strength). The data provenance is Coltène/Whaledent AG, Switzerland, a prospective study is implied as part of the new device's development and testing for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a non-clinical assessment of material properties, not an evaluation of an AI algorithm based on human expert ground truth.

    4. Adjudication method for the test set

    Not applicable. This is a non-clinical assessment of material properties.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a non-clinical assessment of material properties. There is no AI component mentioned in the document.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a non-clinical assessment of material properties. There is no AI component mentioned in the document.

    7. The type of ground truth used

    The "ground truth" for the non-clinical performance data is based on:

    • Physical and mechanical property measurements: Direct laboratory measurements using standardized testing methods (e.g., ISO 4049:2009 for dental polymer-based restorative materials).
    • Biocompatibility: Adherence to established biological evaluation standards (ISO 10993-1:2018).
    • Risk Management: Application of risk management principles (EN ISO 14971:2012).

    8. The sample size for the training set

    Not applicable. There is no AI algorithm and therefore no training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no AI algorithm and therefore no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1