510(k) Data Aggregation

    K Number
    K190409

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-05-03

    (71 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DePuy Synthes MatrixRIB Fixation System is indicated for use in skeletally mature patients with normal or osteoporotic bone for chest wall fixation, where:
    DePuy Synthes MatrixRIB pre-contoured plates are indicated for the fixation, stabilization and reconstruction of:

    • Rib fractures, fusions, osteotomies, and/or resections, including spanning gaps and/or defects
    • Pectus Excavatum, Pectus Carinatum, and other chest wall deformities
      DePuy Synthes MatrixRIB straight plates are indicated for the fixation, stabilization and reconstruction of:
    • Rib and sternum fractures, fusions, osteotomies, and/or resections, including spanning gaps and/or defects
    • Pectus Excavatum, Pectus Carinatum, and other chest wall deformities
      DePuy Synthes MatrixRIB sternal plates, 2.8mm thickness, are indicated for fixation, stabilization and reconstruction of:
    • Sternum fractures, fusions, and/or osteotomies
    • Pectus Excavatum, Pectus Carinatum, and other chest wall deformities
      The DePuy Synthes MatrixRIB intramedullary splints and the universal plate are indicated for the fixation and stabilization of ribs.
    Device Description

    The devices subject to this notification are proposed for addition to the MatrixRIB Fixation System.
    The subject devices consist of self-drilling screws and screw guides designed to be used with the previously cleared MatrixRIB 1.5mm plates and the splints.
    The MatrixRIB Self-Drilling Screws are available as
    • locking screws, for permanent fixation,
    • and non-locking screws, for temporary fixation only.
    The subject screws are manufactured from Titanium Alloy (Ti-6Al-7Nb). They are provided in a range of lengths
    • locking: 13 lengths from 8mm to 20mm, with 1mm increments;
    • non-locking: 2 lengths from 10mm to 12 mm, with 2 mm increments).
    The MatrixRIB Screw Guides are to be used with the subject self-drilling screws to facilitate their fixation on patient's ribs and sternum. The subject screw guides are manufactured from Stainless Steel and are provided in 2 designs,
    • one for the 1.5mm MatrixRIB plates,
    • and one for the MatrixRIB IM splints.

    AI/ML Overview

    The provided text does not contain information about the acceptance criteria or study details for a device that involves performance metrics such as reader studies or algorithm performance. The document describes a 510(k) premarket notification for the "MatrixRIB Fixation System" which is a metallic bone fixation appliance. The acceptance criteria and performance data discussed are related to the mechanical properties of the screws within the system, not diagnostic or AI-driven performance.

    Therefore, I cannot fulfill the request for information on acceptance criteria, a table of acceptance criteria and reported device performance, sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth, training set size, or how training set ground truth was established, as this information is not present in the provided document.

    The document does mention:

    • Non-clinical performance data: "Non-clinical testing and analyses conducted as part of this submission include: Screw Pull-Out Testing per ASTM F543-17, Screw Failure Torque Testing per ASTM F543-17, Screw Cantilever Load Testing, Construct Dynamic Testing. The non-clinical performance data demonstrates that the mechanical performance of the proposed MatrixRIB Self-Drilling Screws met all acceptance criteria and is comparable to that of the predicates."
    • Clinical testing: "Clinical testing was not necessary for the determination of substantial equivalence."

    This indicates that the "acceptance criteria" referred to in the document are mechanical and material-specific, not related to diagnostic accuracy or AI performance.