510(k) Data Aggregation

    K Number
    K190383

    Validate with FDA (Live)

    Device Name
    GM501 Wash
    Date Cleared
    2019-09-05

    (198 days)

    Product Code
    Regulation Number
    884.6180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GM501 Wash is intended for in vitro procedures involving handling and micromanipulation of human occytes and embryos outside of a CO> incubator. Include oocyte and embryo washing (e.g. after oocyte aspiration, after hyaluronidase treatment to remove cumulus cells, before and after cryopreservation, and before embryo transfer) and micromanipulation procedures (e.g. assisted hatching). GM501 Wash is not intended for use in transferring embryos into the uterine cavity.

    Device Description

    GM501 Wash is a ready-to-use solution providing supporting conditions for human oocytes and embryos during in vitro Assisted Reproduction Technology (ART) procedures taking place outside of a CO2 incubator, including washing and micromanipulation procedures. GM501 Wash is aseptically filled into sterilized bottles (20, 50 and 500 ml) and has a six-month shelf-life when stored as recommended. This product can also be used for up to seven days after bottle opening.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the GM501 Wash device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance CriterionReported Device Performance (Implied Met)
    pH7.2-7.5Met (pH testing performed)
    Osmolality270-290 mOsm/kgMet (Osmolality testing performed)
    SterilityNo growth (per USP <71>)Met (Sterility testing performed)
    Bacterial Endotoxins<0.25 EU/ml (per USP <85>)Met (Bacterial endotoxins testing performed)
    Mouse Embryo Assay (MEA)>80% blastocysts at 96h following a 1h exposure to GM501 WashMet (MEA performed with established protocol)
    Aseptic Filling ValidationPer ANSI/AAMI/ISO 13408-1:2008(R)2011, ANSI/AAMI/ISO 13408-2:2003(R)2013, and FDA GuidanceMet (Performed)
    Shelf-Life (6 months)All above specifications (pH, osmolality, sterility, 1-cell MEA, endotoxin) maintained at 6 monthsMet (Shelf-life testing performed)
    After Bottle Opening (7 days)All above specifications (pH, osmolality, sterility, 1-cell MEA, endotoxin) maintained 7 days after openingMet (Stability testing after bottle opening performed)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a specific "test set" sample size for evaluating the device as a whole. However, it mentions the following for the Mouse Embryo Assay (MEA):

    • Sample Size for MEA: "One-cell mouse embryos were exposed to GM501 Wash for one hour..." The specific number of embryos used is not provided.
    • Data Provenance: The origin of the mouse embryos (e.g., country) is not specified. The study appears to be a prospective evaluation of the new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The ground truth for the MEA and other tests (pH, osmolality, etc.) is based on scientific and regulatory standards, not expert consensus on individual cases.

    4. Adjudication Method for the Test Set

    This information is not applicable as the evaluations are objective laboratory tests with defined criteria, not subjective interpretations requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a reproductive media product, not an AI-powered diagnostic or assistive tool, so such a study would not be relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device. The phrase "standalone" as typically used in AI/software evaluation does not apply here. The device's performance is inherently "standalone" in that its intrinsic physical and biological properties are tested.

    7. The Type of Ground Truth Used

    The ground truth used for each parameter is based on:

    • Established scientific/regulatory standards and specifications:
      • pH, Osmolality, Endotoxin: Standard laboratory measurements against defined numerical ranges.
      • Sterility: Absence of microbial growth, per USP <71>.
      • Aseptic Filling Validation: Compliance with international and FDA guidelines (ANSI/AAMI/ISO 13408 series, FDA Guidance).
      • Mouse Embryo Assay (MEA): The biological response of mouse embryos (% developing to blastocyst stage) against a defined percentage threshold (>80%). This is an established biological assay used in reproductive media testing to assess toxicity and developmental support.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical product (reproductive media), not a machine learning model, so there is no concept of a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for a machine learning model, this question does not apply.