Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190318

    Validate with FDA (Live)

    Date Cleared
    2019-03-14

    (29 days)

    Product Code
    Regulation Number
    870.4450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VASOBand™ is a compression device to assist hemostasis of the radial artery after a transradial procedure.

    Device Description

    The VASOBand™ Vascular Compression Device is an external compression device used on a human arm and provides a means for a healthcare professional to apply external compression proximal to and/or over a catheterization puncture site for the purpose of assisting post-procedure hemostasis at a target vessel's puncture site. Each device is a single-use, sterile, individually pouched inflatable band. The VASOBand™ device is available in two different lengths. The strap measures approximately 26 cm or 29 cm in length and is composed of biocompatible multiple-layer PVC film and PVC tubing, and valves made of biocompatible thermoplastics. Hook and loop material is used to secure the strap in position. Only the biocompatible thermoplastic film comes into contact with broken skin. The user can control the compression applied by the device by introducing or removing air in a bladder of the device.

    AI/ML Overview

    This document is a 510(k) summary for the VASOBand™ Vascular Compression Device. The purpose of a 510(k) summary is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device. This process generally relies on non-clinical performance data rather than extensive clinical studies with human participants. Therefore, the information provided below will be based on the type of data typically found in a 510(k) submission for a device of this nature.

    The device being cleared is the VASOBand™ Vascular Compression Device.

    1. A table of acceptance criteria and the reported device performance

    The provided document lists performance data that was submitted to support substantial equivalence. It states that "The VASOBand™ Vascular Compression Device met all specified criteria." However, it does not explicitly state the specific acceptance criteria or the quantitative reported device performance for each test. It only lists the types of tests conducted.

    Based on the information provided, a table would look like this, with the understanding that the specific numerical acceptance criteria and performance are not detailed in this public summary. They would typically be in the full 510(k) submission.

    Acceptance Criterion (Type of Test)Reported Device Performance
    Band clarityMet specified criteria
    Check valve actuationsMet specified criteria
    Bladder inflation/deflation cyclesMet specified criteria
    Band deployment timeMet specified criteria
    Bladder inflation/deflation timeMet specified criteria
    Bladder inflated heightMet specified criteria
    Bladder integrity (2h compressive force)Met specified criteria
    Bladder integrity (underwater air leak)Met specified criteria
    Bladder integrity (simulated use testing)Met specified criteria
    Tubing/check valve tensile strengthMet specified criteria
    Fastener tensile strengthMet specified criteria
    Fastener slippageMet specified criteria
    Cytotoxicity Study (ISO Elution Method)Met specified criteria
    ISO Guinea Pig Skin Sensitization StudyMet specified criteria
    ISO Intracutaneous Irritation StudyMet specified criteria
    USP Rabbit Material-Mediated Pyrogen StudyMet specified criteria
    ISO Acute Systemic Toxicity StudyMet specified criteria
    Endotoxin (LAL) StudyMet specified criteria
    Shelf Life TestingMet specified criteria

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each test. For devices like this, the sample size for non-clinical performance and biocompatibility testing is typically determined by industry standards (e.g., ISO, ASTM) and statistical rationale to ensure representative results.

    The data provenance is described as "performance data provided in support of the substantial equivalence" by the manufacturer, VASOInnovations, Inc. This is non-clinical, in-vitro, and bench testing, as well as biocompatibility studies (which may involve animal testing as indicated for pyrogen and sensitization studies). The country of origin for the data generation is not explicitly stated, but it would have been conducted by or for VASOInnovations, Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of information is generally not applicable to a 510(k) submission for a physical device like a vascular compression band. The "ground truth" for the performance tests listed are objective, measurable physical and chemical properties and biological responses (e.g., tensile strength, air leak, cytotoxicity). These are assessed against pre-defined engineering specifications and relevant ISO/USP standards, not by expert consensus in a clinical setting.

    4. Adjudication method for the test set

    Not applicable. As described above, the "test set" here refers to the physical device samples undergoing engineering and biological safety tests, not clinical data requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No. An MRMC comparative effectiveness study is typically relevant for diagnostic imaging AI algorithms or similar technologies where human readers interpret medical images or data. The VASOBand™ is a physical medical device (a vascular compression band), and its substantial equivalence is demonstrated through non-clinical performance and safety testing, not human reader studies.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device does not involve an algorithm.

    7. The type of ground truth used

    The "ground truth" for the numerous performance data points (e.g., tensile strength, bladder integrity, biocompatibility) are established through objective measurements against predetermined scientific and engineering specifications, and relevant international standards (e.g., ISO, USP). For biocompatibility, this involves specific lab tests following protocols to assess cytotoxicity, irritation, sensitization, pyrogenicity, and systemic toxicity.

    8. The sample size for the training set

    Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1