510(k) Data Aggregation

    K Number
    K190306

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-01-17

    (339 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AMD Ritmed AssureWear™ VersaGown is intended to be worn by healthcare personnel to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. AssureWear™ VersaGown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70). AMD Ritmed AssureWear™ VersaGown is a single use, non-sterile disposable medical device and not intended for use in operating rooms. The medical device will be available in 18 models in large and Xlarge sizes.

    Device Description

    AMD Ritmed AssureWear™ VersaGown isolation gown is intended to be worn by healthcare personnel to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material. AMD Ritmed AssureWear™ VersaGown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown ger ANSI/AAMI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70).

    AMD Ritmed AssureWear™ VersaGown is a single use, disposable medical device provided nonsterile and non-intended for use in operating rooms and is constructed of polypropylene Spunbond--Meltblown-Spunbond Nonwovens (SMS), non-woven and coated with polyethylene. The isolation gown consists of one critical zone throughout the entire gown including the seams and belt attachments but excluding cuffs, hems and bindings.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the AMD Ritmed AssureWear™ VersaGown. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study of an AI/human-in-the-loop diagnostic device.

    Therefore, many of the requested criteria regarding AI device performance, clinical studies, expert consensus, and ground truth establishment are not applicable to this document. The document describes the acceptance criteria and performance of a physical medical device (isolation gown) based on non-clinical performance testing against established industry standards.

    Here's the information as it can be extracted from the document:


    Acceptance Criteria and Device Performance for AMD Ritmed AssureWear™ VersaGown

    The device is an isolation gown intended to provide barrier protection. The "acceptance criteria" are essentially the performance requirements for an AAMI Level 3 barrier protection gown, as defined by ANSI/AAMI PB70:2012, along with other physical and biological safety standards. The "reported device performance" is the result of the tests conducted on the VersaGown.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance ParameterAcceptance Criteria (for AAMI Level 3 Gown, where applicable & standards listed)Reported Device Performance (AMD Ritmed AssureWear™ VersaGown)
    Level of barrier protection (AAMI PB70)Level 3Level 3
    Basic weight (ASTM D3776)Predicate: 41 g/m² (1.21 oz/yd²)39.97 ± 1.61 g/m² (1.17 oz/yd²±0.05)
    Liquid barrier performance
    - Hydrostatic Pressure (AATCC 127)Predicate: Body/sleeve mean: 69 cmH2OChest: 109.34 ± 0.34 cmH2OSleeve seams: 110.67 ± 3.84 cmH2OBelt attachments: 104 ± 5.19 cmH2OBody/sleeve/belt mean: 108 ± 3.1 cm H₂O
    - Impact penetration (AATCC 42)Predicate: Body/sleeve mean: 0.08 gChest: <0.1 gSleeve seams: <0.1 gBelt attachments: < 0.1 gBody/sleeve/belt mean: < 0.1 g
    Flammability (16 CFR Part 1610.7)Class IClass I
    Breaking strength (ASTM D5034)Predicate MD: 22.23 lbf; Predicate CD: 14.18 lbfMD: 18.17 ± 0.31 lbfCD: 11.78 ± 0.33 lbf
    Tearing strength (ASTM D5587)Predicate MD: 4.40 lbf; Predicate CD: 7.99 lbfMD: 11.01 ± 0.64 lbfCD: 5.30 ± 0.35 lbf
    Linting (ISO 9073-10)No predicate data provided for comparisonParticulate size range (µm): 1 to 25Outside: Total linting >0.3: 2.07; >0.5: 1.97Index for Particulate Matter (IPM): 1.50Inside: Total linting >0.3: 2.16; >0.5: 2.00Index for Particulate Matter (IPM): 1.35
    BiocompatibilityNon-cytotoxic, non-irritant, non-sensitizingIrritation (ISO 10993-10): Under the condition of the study, not an irritant.Sensitization (ISO 10993-10): Under the condition of the study, not a sensitizer.Cytotoxicity (ISO 10993-5): Under the condition of the study, non-cytotoxic.

    Note: For several physical properties (e.g., basic weight, breaking strength, tearing strength), the "acceptance criteria" are implicitly set by comparing to the performance of the legally marketed predicate device. The values for the predicate device are listed in the table if available for comparison.

    2. Sample sized used for the test set and the data provenance

    The document does not specify the exact sample sizes for each non-clinical performance test. The results are presented as means and standard deviations, which implies multiple measurements were taken for each test. Data provenance is not described in terms of country of origin or retrospective/prospective, as this is laboratory testing of a manufactured product, not human-subject data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is for a physical medical device (isolation gown), not a diagnostic algorithm requiring expert interpretation or ground truth establishment in a clinical context. The "ground truth" here is determined by the results of standardized physical and biological tests.

    4. Adjudication method for the test set

    Not applicable. No human adjudication is involved in the performance testing of a physical medical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done

    No. This type of study is typically done for diagnostic imaging AI devices, not for physical medical devices like isolation gowns.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm. The performance stated is intrinsic to the physical device.

    7. The type of ground truth used

    The ground truth for this device's performance is established by standardized laboratory testing protocols (e.g., AATCC 127, AATCC 42, ASTM D3776, ISO 9073-10, ISO 10993 series) against the requirements for AAMI Level 3 barrier protection. It is a physical/chemical ground truth rather than an expert consensus, pathology, or outcomes data ground truth in the context of diagnostic AI.

    8. The sample size for the training set

    Not applicable. This is not an AI device that involves training data.

    9. How the ground truth for the training set was established

    Not applicable. No training set is involved.