Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190194

    Validate with FDA (Live)

    Date Cleared
    2019-05-31

    (116 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The C2 CryoBalloon Ablation System is intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett's Esophagus with dysplasia.

    Device Description

    The subject and marketed predicate device is a cryosurgical unit with a nitrous oxide cooled balloon that is compatible with commercially available endoscopes with a minimum working channel inner diameter of 3.7 mm and maximum length of 105 cm. The subject device is a cryosurgical system comprised of four components including a Catheter (sterile, single use), Controller (non-sterile, reusable), Foot Pedal (non-sterile, reusable), and Cartridge (non-sterile, single use).

    The subject and marketed predicate device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the pre-inflated balloon, and the treatment site is selected is by adjusting the endoscope and Catheter The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Controller.

    AI/ML Overview

    The PENTAX Medical C2 CryoBalloon Ablation System is intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett's Esophagus with dysplasia.

    Here's an analysis of the provided information regarding acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document is a 510(k) summary for a medical device seeking substantial equivalence to a predicate device. It doesn't present a specific table of quantitative acceptance criteria for clinical performance that would typically be found in a clinical trial report. Instead, the "acceptance criteria" are implied to be the successful completion of various design verification and validation tests, ensuring the device performs as intended and is as safe and effective as its predicate.

    CategoryAcceptance Criteria (Implied)Reported Device Performance
    Design Verification & ValidationAll physical, reliability, and safety specifications met."The acceptance criteria have been satisfied for all tests."
    SterilizationSterilization validated in accordance with ISO 11135."both sterilized utilizing an Ethylene Oxide sterilization cycle validated in accordance with ISO 11135."
    BiocompatibilityEvaluation conducted per ISO 10993-1 with passing results for patient-contacting materials."evaluation... conducted in accordance with 'Guidance...ISO 10993-1'... with passing results." (Note: No new testing was required due to no changes in patient-contacting materials from predicate).
    EMC and Electrical SafetyCompliance with relevant Electromagnetic Compatibility (EMC) and electrical safety standards (IEC 60601-1-2:2014 and ES60601-1:2005/(R)2012 And A1:2012)."acceptable level of electromagnetic compatibility (EMC) and electrical safety (ES)... confirmed by the testing to the following standards."
    Technological CharacteristicsSimilar technological characteristics to the legally marketed predicate device (K163684)."Cryoablation is the fundamental technological principle for both... same endoscopic instrumentation... similar technological characteristics... based on the following same technological elements." (Details listed in the device description).
    Safety and EffectivenessAs safe and effective as the legally marketed predicate for the same intended use."The performance data results demonstrate that the subject C2 CryoBalloon™ Ablation System is as safe and effective as the legally marketed predicate that is currently marketed for the same intended use."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes preclinical (bench) performance testing, not a clinical study with a "test set" of patients. Therefore, information regarding human sample size and data provenance (country of origin, retrospective/prospective) is not applicable or provided in this 510(k) summary. The testing performed was related to design verification and validation of the device's physical, reliability, and safety specifications.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable, as this document refers to preclinical engineering and safety testing, not a clinical study requiring expert ground truth for patient data.

    4. Adjudication Method for the Test Set:

    Not applicable for the reasons stated above.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence to a predicate device through preclinical performance testing, not a comparative clinical trial with human readers.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance):

    This device is not an AI algorithm. It is a cryosurgical tool. Therefore, the concept of "standalone performance" for an algorithm is not applicable. The performance discussed relates to the device's physical and functional operation.

    7. Type of Ground Truth Used:

    For the performance data discussed (sterilization, biocompatibility, EMC, electrical safety, and design verification/validation), the "ground truth" is established through:

    • Standardized testing procedures: Adherence to recognized national and international standards (e.g., ISO 11135, ISO 10993-1, IEC 60601-1-2, ES60601-1).
    • Engineering specifications and design requirements: The device's performance is measured against predefined engineering and design specifications.
    • Comparison to the predicate device: The performance is also judged in the context of being similar to the legally marketed predicate device.

    8. Sample Size for the Training Set:

    Not applicable, as this is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as this is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1