Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K190157

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-23

    (266 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diana ChemoClave Transfer Set is a sterile, single-use closed system transfer device used for drug preparation to transfer drug from a drug vial to an IV bag for intravenous drug administration. It prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system.

    Device Description

    Diana ChemoClave Transfer Set is intended for use with the Diana Automated Compounding System. It is a disposable, single use ancillary device for Pharmacy Compounding Devices (PCD).

    The transfer set is comprised of multiple components bonded together to form a single device. These components include the following: stopcock, tubing, ChemoClave, syringe, and cassette handle. When placed in the Diana Automated Compounding System, the Diana ChemoClave Transfer Set allows the transfer of fluids from one container to the other container for reconstitution of lyophilized drug or transfer of stock drug solution to prepare medications. Connections to sourcing/receiving containers are enabled using the dedicated ChemoClave connector.

    The purpose of this submission is to modify the predicate's Closed System Transfer Device (CSTD) from ChemoLock to ChemoClave, as an alternative to be used with Diana Automated Compounding System.

    AI/ML Overview

    The provided text pertains to a 510(k) premarket notification for a medical device called the "Diana ChemoClave Transfer Set." The document focuses on demonstrating substantial equivalence to a predicate device, rather than providing a detailed study report with specific acceptance criteria and performance data in a quantitative manner that would be typical for an AI/ML device.

    However, I can extract the types of tests conducted and list them as criteria that the device must have met to achieve substantial equivalence. The document does not provide numerical performance data against these criteria, but rather states that the tests were conducted and the device meets functional claims.

    Here's an interpretation based on the provided text:

    Acceptance Criteria and Device Performance

    The document describes various performance tests conducted to demonstrate the safety and effectiveness of the Diana ChemoClave Transfer Set and its substantial equivalence to the predicate device. The "acceptance criteria" are implied to be that the device passes these tests, thereby meeting functional claims and ensuring safety and efficacy consistent with the predicate.

    Missing Information: It's important to note that this document does not contain the specific numerical acceptance criteria or the reported device performance values for each test. For instance, it lists "Fluid Flow" as a test but doesn't specify a required flow rate (e.g., X ml/minute) or the measured flow rate of the Diana ChemoClave Transfer Set. It merely states that the functional testing was conducted.

    Acceptance Criteria CategorySpecific Test / RequirementImplied Reported Device Performance (as per document)
    Functional PerformanceFluid Flow"Functional testing was conducted to demonstrate the functionality... The Diana ChemoClave Transfer Set meets the functional claims..." (Implied pass)
    Positive Pressure Leakage"Functional testing was conducted... The Diana ChemoClave Transfer Set meets the functional claims..." (Implied pass)
    Chemical Compatibility"Functional testing was conducted... The Diana ChemoClave Transfer Set meets the functional claims..." (Implied pass)
    Stopcock Handle Torque"Functional testing was conducted... The Diana ChemoClave Transfer Set meets the functional claims..." (Implied pass)
    Component Bond Strength Testing"Functional testing was conducted... The Diana ChemoClave Transfer Set meets the functional claims..." (Implied pass)
    Visual Inspection"Functional testing was conducted... The Diana ChemoClave Transfer Set meets the functional claims..." (Implied pass)
    Accuracy Testing"Functional testing was conducted... The Diana ChemoClave Transfer Set meets the functional claims..." (Implied pass)
    Internal Seal Integrity"Functional testing was conducted... The Diana ChemoClave Transfer Set meets the functional claims..." (Implied pass)
    Environmental ContainmentMicrobial Ingress Test"Microbial ingress testing was conducted to demonstrate that needless access site... maintains physical integrity and mechanically prohibits the transfer of environmental contaminants into the system..." (Implied pass)
    Emission (of drug/vapor)"Emission, Dry Disconnection, and CSTD Hazardous Drug Exposure Evaluation were performed to validate no escape of drug from the subject device outside the closed system." (Implied pass)
    Dry Disconnection (no drug escape)"Emission, Dry Disconnection, and CSTD Hazardous Drug Exposure Evaluation were performed to validate no escape of drug from the subject device outside the closed system." (Implied pass)
    CSTD Hazardous Drug Exposure Evaluation (no drug escape)"Emission, Dry Disconnection, and CSTD Hazardous Drug Exposure Evaluation were performed to validate no escape of drug from the subject device outside the closed system." (Implied pass)
    BiocompatibilityHemocompatibility (Per FDA Guidance and ISO 10993-1)"The biocompatibility evaluation... was conducted... Testing included: Hemocompatibility." (Implied pass, consistent with standards)
    Cytotoxicity (Per FDA Guidance and ISO 10993-1)"The biocompatibility evaluation... was conducted... Testing included: Cytotoxicity." (Implied pass, consistent with standards)
    Sensitization (Per FDA Guidance and ISO 10993-1)"The biocompatibility evaluation... was conducted... Testing included: Sensitization." (Implied pass, consistent with standards)
    Intracutaneous Irritation (Per FDA Guidance and ISO 10993-1)"The biocompatibility evaluation... was conducted... Testing included: Intracutaneous Irritation." (Implied pass, consistent with standards)
    Acute Systemic Toxicity (Per FDA Guidance and ISO 10993-1)"The biocompatibility evaluation... was conducted... Testing included: Acute Systemic Toxicity." (Implied pass, consistent with standards)
    Pyrogenicity (Per FDA Guidance and ISO 10993-1)"The biocompatibility evaluation... was conducted... Testing included: Pyrogenicity." (Implied pass, consistent with standards)
    Sterility & ParticulatesSterility Assurance Level (SAL) of 10-6"E-beam sterilization process validation was conducted according to ISO 11137-1 and ISO11137-2... (Implied achieved SAL of 10-6 as listed in comparison table)"
    Bacterial Endotoxin Testing (USP <85>, AAMI ST72)"Bacterial endotoxin testing was conducted based on AAMI ST72 and USP <85>, and followed FDA Guidance..." (Implied pass)
    Particulate Contamination (USP <788> requirements)"Particulate contamination testing was performed by following USP <788> to demonstrate particulate levels in the subject device meet USP 788 requirements." (Implied pass)
    Packaging Integrity & Shelf LifePackage integrity (ASTM D4169, F1980, F2096, F88)"Package integrity and Shelf Life was conducted according to FDA recognized consensus standards of ASTM D4169... F1980... F2096... and F88..." (Implied pass, meeting standards and demonstrating shelf life)
    Shelf Life (ASTM F1980)"Package integrity and Shelf Life was conducted according to FDA recognized consensus standards... F1980..." (Implied pass, demonstrating shelf life)

    Study Details for Device Performance

    The document describes the various studies performed to demonstrate equivalence, but does not provide details on sample sizes, provenance, expert qualifications, or adjudication methods as these are not typically included in a 510(k) summary for this type of medical device (non-AI/ML).

    Here's an analysis based on your specific questions:

    1. A table of acceptance criteria and the reported device performance:

      • See table above for the types of acceptance criteria (tests) and the implied reported performance ("passed" or "meets functional claims"). Specific numerical acceptance limits and measured performance values are not provided in this document.
    2. Sample size used for the test set and the data provenance:

      • Not specified. The document mentions various tests (Functional, Biocompatibility, Microbial Ingress, etc.) but does not provide details on the number of units tested (sample size) for any of these studies.
      • Data Provenance: Not explicitly stated, but the submission is from ICU Medical, Inc. based in Lake Forest, Illinois, USA. It's typical for such testing to be conducted in-house or by contract labs in the company's country of origin or a region with accredited testing facilities. The document refers to FDA guidance and international standards, implying adherence to regulated testing environments.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable / not specified. This is not an AI/ML device where "ground truth" established by experts is a primary element of performance validation in the way it would be for diagnostic algorithms. Device performance is determined by meeting physical, chemical, and biological specifications through laboratory testing.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable / none specified. Adjudication methods like 2+1 or 3+1 are used for reconciling expert opinions in diagnostic AI/ML studies. This document describes physical, chemical, and biological testing governed by established standards, not expert adjudication.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is not an AI/ML device, so MRMC studies and "human readers improving with AI assistance" are not relevant or performed for this product.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No. This is not an AI/ML algorithm. Its "standalone" performance refers to its physical and functional attributes as a mechanical device. The listed tests (Fluid Flow, Leakage, Biocompatibility, Sterility, etc.) represent its standalone performance.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Benchmarking against established technical standards and specifications. For this type of device, "ground truth" comes from meeting pre-defined physical, chemical, and biological specifications as outlined in harmonized standards (e.g., ISO, ASTM, USP) and FDA guidance documents. For example, "Non-pyrogenic" means it passed specific tests to confirm absence of pyrogens, which is the "ground truth" for that characteristic.
    8. The sample size for the training set:

      • Not applicable. This device does not involve a "training set" in the context of machine learning.
    9. How the ground truth for the training set was established:

      • Not applicable. This device does not involve a "training set" or a corresponding "ground truth" establishment process for machine learning.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1