Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Veloxion System is intended for intrauterine use by trained gynecologists for endoscopically controlled tissue chip resection and coagulation, and removal of intrauterine polyps and myomas via suction channel under continuous flow conditions following resection using a bipolar resectoscope. It is also intended to distend the uterus by filling with saline to facilitate viewing and to monitor the volume differential between fluid flowing into and out of the uterus.

    Device Description

    The Veloxion System consists of the Veloxion Controller (with Integrated Fluid Control), Footswitch, Veloxion Resectoscope, Veloxion Fluid Control Set, Veloxion Video Control Unit, and Veloxion Roll Stand. It also includes Class I accessories for waste handling: Waste Management Tubing, Tissue Catch, and Waste Management Bags. The system provides bipolar resection and coagulation, distends the uterus with saline, controls intrauterine pressure, and monitors fluid deficit.

    AI/ML Overview

    Based on the provided text, here's a detailed breakdown of the acceptance criteria and the study conducted for the Veloxion System:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative "acceptance criteria" in a pass/fail format for specific performance metrics. Instead, it describes performance characteristics that were tested and compared to the predicate device. For the purpose of this request, I will extract the performance characteristics that were "tested" and the reported "performance" from the "Technological Characteristics" table (Table 1) and the "Performance Data" section.

    Acceptance Criteria/Performance CharacteristicReported Device Performance (Subject Veloxion System)Notes / Comparison to Predicate
    Fluid Deficit Monitoring
    Set Fluid Deficit RangeFrom 250 to 2500mlSame as Predicate
    Fluid Deficit Rate ≥ 300ml/min notificationYESSame as Predicate
    Notification when Actual fluid deficit is 1000mlYES (no pause, notification only)Same as Predicate
    Notification & pause when Actual fluid deficit is 2000mlYESDifferent (Predicate: NO)
    Notification when Actual fluid deficit is within 250ml of Set LimitYESSame as Predicate
    Device Stops when Set Limit or 2500ml is reachedYESSame as Predicate
    Pressure Control
    Set Pressure Range35 to 120mmHgSlightly different (Predicate: 35 to 125mmHg)
    Positive Action to Increase Above 100mmHgYESSame as Predicate
    Pressure SensorDual, independent sensing of cavity pressureSame as Predicate
    Over-pressure condition detectionYESSame as Predicate
    Maximum Allowable Actual Cavity Pressure135mmHg for 5 secondsSame as Predicate
    Primary mitigation for risk of over-pressurizationReverse rotation of irrigation wheelSame as Predicate
    Ultimate mitigation for risk of over-pressurizationNon-defeatable, continuous notification tone; Notification displayed: "Remove device from cavity. Check for Clog."Same as Predicate
    Fluid Flow and Management
    Continuous FlowYESSame as Predicate
    Irrigation fluidSalineSame as Predicate
    Maximum # of saline bag(s) hung on saline load cellFour (4)Different (Predicate: One (1))
    Load cellsSeparate load cells to weigh the saline bags and the waste bagsDifferent (Predicate: One load cell for combined saline and waste)
    Pump TypeDual Pump (Irrigation, Aspiration)Same as Predicate
    Irrigation Flow Rate (Programmed for each mode)Steady State = 100ml/min, Cut = 380ml/min, Coag = 100ml/min, ASPIRATE = 650ml/minDifferent (Predicate: ASPIRATE = 400ml/min)
    Suction for aspiration of waste from patient drapeFacility SuctionDifferent (Predicate: Waste Accessory Pump)
    Able to Monitor saline remaining?YESSame as Predicate
    Imaging/Optics
    OpticsIntegrated, Sterile, 10° EndoscopeDifferent (Predicate: Compatible with reusable 30° marketed Endoscope)
    ImagingIntegrated CMOS sensorDifferent (Predicate: External CCD camera)
    Total # Pixels160,000(Predicate: Unknown)
    Pixels/mm2160,000(Predicate: Unknown)
    Light SourceLED lightDifferent (Predicate: Xenon, Halogen, LED)
    Resectoscope/Electrode
    Energy TypeRadiofrequency, BipolarSame as Predicate
    Monopolar or BipolarBipolarSame as Predicate
    Shaft OD (Sheath)With Sheath Assembled: 25Fr (8.3mm)Slightly different (Predicate: 26Fr (8.6mm))
    Electrode OD0.019"Slightly different (Predicate: 0.015")
    Electrode MaterialTungsten (99.95% purity)Same as Predicate
    Electrode InsulationFEPSame as Predicate
    Electrode ActuationLinear Oscillating LoopDifferent (Predicate: Circumferential Oscillating Wire)
    How Supplied (Sterility)Sterile, Single Use (Resectoscope, Fluid Control Set)Same as Predicate

    2. Sample Size for the Test Set and Data Provenance

    The document lists various performance tests and comparative testing to the predicate, but it does not specify the sample sizes (n) used for each test set.

    The data provenance is not explicitly stated regarding country of origin or whether the studies were retrospective or prospective. The studies appear to be benchtop performance and engineering tests conducted in-house by the manufacturer rather than clinical trials involving patient data. This is implied by tests such as "Comparative visualization testing in a simulated model" and "Simulated Use: Tissue resection and spot coagulation of cavity pressure, O imaging."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    The document does not mention the use of experts or their qualifications for establishing ground truth for any of the performance tests described. The "performance data" section primarily describes engineering and benchtop tests, not clinical evaluations requiring expert interpretation.

    4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set

    As the document does not describe studies involving human readers or interpretation of results that would require adjudication (e.g., image reading), no adjudication method is mentioned or implied.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC comparative effectiveness study, nor any evaluation of AI assistance or human reader improvement. The device described is a medical instrument (hysteroscope system), not an AI-powered diagnostic or interpretive tool that would typically involve such studies.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a physical medical instrument, not a standalone algorithm. Its performance is inherent to its mechanical, electrical, and optical functions.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the engineering and performance tests:

    • Physical/Engineering Standards: Measurements against known physical properties or engineering specifications (e.g., pressure, flow rates, dimensions, temperature, electrical safety).
    • Predicate Device Performance: Direct comparison to the performance of the legally marketed predicate device (K180752), establishing its performance as a de facto "ground truth" for equivalence.
    • Simulated Models: For visualization and tissue interaction tests ("Comparative visualization testing in a simulated model," "Simulated Use: Tissue resection and spot coagulation").

    There is no mention of expert consensus, pathology, or outcomes data being used to establish ground truth for the technical performance data provided.

    8. The Sample Size for the Training Set

    This question is not applicable as the document describes a medical device, not a machine learning model that would require a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1