510(k) Data Aggregation

    K Number
    K190071

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-02-25

    (41 days)

    Product Code
    Regulation Number
    888.3075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PERLA® posterior cervico-thoracic fixation system is intended to provide immobilization of spinal segments as an adjunct to following acute and chronic instabilities of the cervical spine (C 1 to C7) and the thoracic spine from Tl-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The PERLA® posterior cervico-thoracic fixation system is also intended to restore the integrity of the spinal column even in the a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the PERLA® posterior cervico-thoracic fixation system may be connected to the ROMEO® Posterior Osteosynthesis System with rod connectors. Transition rods may also be used to connect the PERLA® posterior cervico-thoracic fixation system to the ROMEO® Posterior Osteosynthesis System. Refer to the ROMEO® Posterior Osteosynthesis System package insert for a list of the ROMEO® Posterior Osteosynthesis System indications of use.

    Device Description

    The Spineart Perla® system is a posterior cervico-thoracic fixation system intended to provide stabilization to promote fusion of the cervical spine and the upper thoracic spine. Perla® system consists of a variety of shapes and sizes of rods, hooks, multi-axial screws, set screws, rod connectors and transverse connectors. These connecting components can be rigidly locked to the rod in a variety of configurations to be adapted for the individual case. The multi-axial screws are available with diameters 3.5mm and 4.0 mm with lengths ranging from 08 mm up to 52 mm. The purpose of this Special 510(k) application is to introduce an additional range of multi-axial screw with 4.5 mm diameter and lengths ranging from 08 mm up to 52 mm to the Perla® spinal system.

    AI/ML Overview

    The provided document is a 510(k) summary for the SPINEART Perla® Posterior Cervico-Thoracic Fixation System. It describes a device modification, specifically the introduction of an additional range of multi-axial screws with a 4.5 mm diameter.

    Here's an analysis of the acceptance criteria and study information, based solely on the provided text:

    Key Takeaway: This 510(k) is for a modification to an existing device (Perla® Posterior Cervico-Thoracic Fixation System, cleared under K153386). The modification is the introduction of a new screw diameter (4.5mm). The submission leverages previous testing and an engineering analysis to demonstrate substantial equivalence, rather than conducting new performance studies for the specific modification. As such, the information you requested regarding acceptance criteria and performance data for a new study is largely not present for this specific submission because it was deemed unnecessary due to the nature of the modification.


    1. A table of acceptance criteria and the reported device performance

    Based on the provided document, there are no specific numeric acceptance criteria for device performance reported for this specific 510(k) submission (K190071). Instead, the submission relies on the existing performance data and acceptance for the predicate device (Perla® Posterior Cervico-Thoracic Fixation System, K153386).

    The document states: "An engineering analysis was performed and determined the new 4.5 mm multi-axial screws do not constitute a new worst-case design in the Perla spinal system. No additional testing has been performed for the Perla® 4.5 mm multi-axial screws."

    It then lists non-clinical tests that were conducted on the predicate devices:

    Test TypeStandardAcceptance Criteria (Not explicitly stated for this submission, inferred from standards)Reported Device Performance (Not explicitly stated for this submission)
    Static Compression BendingASTM F1717Device should meet specified strength/integrity requirements of the standardImplied to have met requirements for predicate device
    Static TorsionASTM F1717Device should meet specified strength/integrity requirements of the standardImplied to have met requirements for predicate device
    Dynamic Compression BendingASTM F1717Device should meet specified strength/integrity requirements of the standardImplied to have met requirements for predicate device
    Static flexion-extension testingASTM F1798Device should meet specified strength/integrity requirements of the standardImplied to have met requirements for predicate device
    Static axial grippingASTM F1798Device should meet specified strength/integrity requirements of the standardImplied to have met requirements for predicate device
    Static torsion grippingASTM F1798Device should meet specified strength/integrity requirements of the standardImplied to have met requirements for predicate device
    Axial pullout strengthASTM F543Device should meet specified strength requirements of the standardImplied to have met requirements for predicate device
    Torque to failureASTM F543Device should meet specified strength requirements of the standardImplied to have met requirements for predicate device
    Bacterial endotoxin testing (Pyrogenicity testing)USP standard (for Endotoxin limit of 20 EU / device)Endotoxin limit of 20 EU / deviceImplied to have met this limit for predicate device

    2. Sample size used for the test set and the data provenance

    As no new performance testing was conducted for the 4.5 mm multi-axial screws due to the engineering analysis, there is no "test set" and thus no sample size or data provenance specified for this particular modification. The previous testing for the predicate device would have had its own sample sizes, but those are not detailed here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no new clinical or performance test set (requiring expert ground truth) was performed for this specific 510(k) submission.

    4. Adjudication method for the test set

    Not applicable, as no new clinical or performance test set (requiring adjudication) was performed for this specific 510(k) submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a mechanical implant (spinal fixation system), not an AI-driven diagnostic or assistive device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a mechanical implant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no new performance studies requiring ground truth were conducted for this submission. The "ground truth" for mechanical testing is the physical measurement against the standard.

    8. The sample size for the training set

    Not applicable, as no AI/algorithm is involved, and no new training set was mentioned for mechanical testing.

    9. How the ground truth for the training set was established

    Not applicable, as no AI/algorithm is involved, and no new training set or corresponding ground truth establishment was discussed.