510(k) Data Aggregation

    K Number
    K183670

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-09-13

    (260 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vexillum Zephyr (olifilcon A) with Tangible Polymers Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    Eye care practitioners may prescribe the lens for either single-use daily disposable wear, or for frequent/ planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for frequent replacement, the lens may be disinfected using a chemical disinfection system only. The Vexillum Zephyr (olifilcon A) with Tangible Polymers Silicone Hydrogel Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    The OxyPure A (olifilcon A) Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    Eye care practitioners may prescribe the single-use daily disposable wear, or for frequent/ planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for frequent replacement, the lens may be disinfected using a chemical disinfection system only. The OxyPure A (olifilcon A) Silicone Hydrogel Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    Device Description

    The Vexillum Zephyr (olifilcon A) with Tangible Polymers Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material, olifilcon A, with UV blocker available as spherical lens. The composition of the lens is 53% olifilcon A and 47% water. A light blue color tinted with "reactive Blue19" listed in 21 CFR Part 73.3121is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The UV transmission (the thinnest lens measured by spectrophotometry as stated in ISO 18369) is less than 50% in the UVA range of 316 - 380 mm and less than 5% in the range of UVB range of 280-315 nm.

    Lenses are supplied sterile in sealed blister packs containing Tangible Polymers (coating on the lens surface during sterilization process) with isotonic buffered saline solution.

    The OxyPure A (olifilcon A) Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material, olifilcon A, with UV blocker available as spherical lens. The composition of the lens is 53% olifilcon A and 47% water. A light blue color tinted with "reactive Blue19" listed in 21 CFR Part 73.3121is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The UV transmission (the thinnest lens measured by spectrophotometry as stated in ISO 18369) is less than 50% in the UVA range of 316 - 380 nm and less than 5% in the range of UVB range of 280-315 nm.

    OxyPure A (olifilcon A) Silicone Hydrogel Soft Contact Lenses are supplied sterile in sealed blister packs containing Sodium Hyaluronate and Sodium Alginate with isotonic buffered saline solution.

    AI/ML Overview

    This document is a 510(k) summary for contact lenses, which focuses on demonstrating substantial equivalence to a predicate device rather than proving performance against specific acceptance criteria for an AI device. Therefore, much of the requested information regarding AI device testing (e.g., sample sizes for test/training sets, ground truth establishment methods, expert qualifications, MRMC studies) is not applicable to this document.

    However, I can extract information related to the device attributes and the comparisons made to establish equivalence, which can be seen as meeting "acceptance criteria" for chemical/physical properties rather than AI performance.

    Here's a breakdown of what can be inferred and what is explicitly stated:

    Acceptance Criteria and Device Performance (Analogous for Contact Lenses)

    In the context of contact lenses, "acceptance criteria" are not thresholds for AI performance, but rather specifications and physical/chemical properties that demonstrate the new devices (Vexillum Zephyr and OxyPure A) are substantially equivalent to the predicate device (OxyPure Color). The table below summarizes these comparisons.

    CategoryAcceptance Criteria (Predicate Device K171447: OxyPure Color)Reported Device Performance (New Devices: Vexillum Zephyr & OxyPure A)Result of Comparison
    ApplicantVisco Vision IncVisco Vision IncSame
    ClassificationClass IIClass IISame
    Regulation Number886.5925886.5925Same
    Product CodeLPL, MVNLPL, MVNSame
    Intended UseMyopia, Hyperopia, astigmatism, PresbyopiaMyopia, Hyperopia, astigmatism, PresbyopiaSame
    Replacement ScheduleDaily WearDaily WearSame
    USAN Nameolifilcon Aolifilcon ASame
    FDA Category (Group)Group 5C (Nonionic, Water < 50 wt %)Group 5C (Nonionic, Water < 50 wt %)Same
    Manufacturing MethodCast MoldedCast MoldedSame
    Lens DesignSphericalSphericalSame
    Water Content47%47%Same
    Light Transmittance94%94%Same
    Refractive Index1.410 (hydrated)1.410 (hydrated)Same
    Oxygen Permeability (DK, 35°C)150 (Fatt method)150 (Fatt method)Same
    Diameter Range13.0 to 15.0 mm13.0 to 15.0 mmSame
    Power Range-20.00D to +20.00D-20.00D to +20.00DSame
    Center Thickness0.08mm @ -3.00D (Varies with Power)0.08mm @ -3.00D (Varies with Power)Same
    Base Curve8.0 mm to 9.2 mm8.0 mm to 9.2 mmSame
    Blue handling tintNoReactive Blue19 (for both new devices)Different
    Color Additives in DyeIron oxide, Titanium dioxide, Phthalocyanine green, Carbazole violetReactive Blue19 (for both new devices)Different
    Leachabilityno leachable monomers and addictive residuesno leachable monomers and addictive residuesSame
    PackagingBlister PackBlister PackSame
    Packaging SolutionSterile isotonic borate buffered salineVexillum Zephyr: sterile isotonic borate buffered saline with Tangible polymersOxyPure A: sterile isotonic borate buffered salineDifferent
    Coating on the lens surfaceNoVexillum Zephyr: Yes, coating with Tangible Polymer during sterilization processOxyPure A: NoDifferent
    Sterilization MethodSteamSteamSame
    Shelf Life5 years5 yearsSame
    Sterility of DeviceSAL= 10^-6SAL= 10^-6Same
    Tensile Strength (Mpa)0.78 ± 0.20.80 ± 0.2Different
    Modulus (Mpa)0.60 ± 0.10.60 ± 0.1Same
    Elongation at break (%)195 ± 20250 ± 50Different
    Toughness (J/m3)0.87 ± 0.051.10 ± 0.05Different
    Compliance Standard (Biocompatibility)ISO10993-1, ISO10993-5, ISO10993-10, ISO10993-11ISO10993-1, ISO10993-5, ISO10993-10, ISO10993-11Same
    Compliance Standard (Sterilization & Shelf Life)ISO 17665-1, ISO11737-1, ISO 11987ISO 17665-1, ISO11737-1, ISO 11987Same
    Compliance Standard (Performance)ISO18369-3, ISO18369-4, ASTM D792-13ISO18369-3, ISO18369-4, ASTM D792-13Same

    Study Proving Device Meets Criteria:

    The study proving the devices meet the "acceptance criteria" (i.e., are substantially equivalent) is based on non-clinical bench testing and comparison of technical characteristics to the predicate device.

    1. Sample Size and Data Provenance:

      • Test Set Sample Size: Not explicitly stated as a "test set" in the context of AI. The "test" consists of various physical, chemical, and mechanical tests on the contact lenses. The specific number of lenses tested for each parameter (e.g., tensile strength, oxygen permeability) is not provided in this summary but is implied to be sufficient for standard quality control and regulatory submission.
      • Data Provenance: The document does not specify the country of origin for the testing data, but the applicant's address is Taiwan. The studies are non-clinical bench tests, implying they are laboratory-based, not patient-based data. They are presumably prospective in the sense that they were specifically conducted for this submission.
    2. Number of Experts and Qualifications for Ground Truth:

      • Not Applicable. This document is not evaluating AI performance or subjective diagnostic interpretations. Ground truth, in this context, refers to the objective measurements of the lens properties themselves, assessed through established laboratory procedures according to recognized ISO and ASTM standards. No human expert "ground truth" establishment in the diagnostic sense is mentioned.
    3. Adjudication Method for Test Set:

      • Not Applicable. As there are no human-in-the-loop diagnostic interpretations requiring adjudication, this concept does not apply.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No. The document explicitly states: "This 510(k) submission does not utilize clinical study for establishing substantial equivalence therefore this section does not apply." This confirms that no clinical studies, including MRMC studies, were conducted.
    5. Standalone Performance (Algorithm Only):

      • Not Applicable. This is not an AI device.
    6. Type of Ground Truth Used:

      • For the physical and chemical properties, the "ground truth" is established through objective laboratory measurements following recognized international standards (ISO, ASTM). Examples include:
        • Spectrophotometry for UV transmission.
        • Fatt method for oxygen permeability.
        • Standardized tests for tensile strength, modulus, elongation, and toughness.
        • Biocompatibility testing as per ISO 10993 series.
        • Sterility testing.
      • For the overall determination of substantial equivalence, the "ground truth" is the established safety and effectiveness of the predicate device (OxyPure Color Silicone Hydrogel Soft Contact Lenses, K171447). The new devices are compared to this established benchmark.
    7. Sample Size for Training Set & How Ground Truth for Training Set was Established:

      • Not Applicable. There is no training set as this is not an AI/machine learning device. The entire process relies on direct measurement and comparison to a known good (predicate) device, and adherence to established industry standards.