510(k) Data Aggregation

    K Number
    K183572

    Validate with FDA (Live)

    Date Cleared
    2019-07-16

    (207 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FUJIFILM Endoscope Model EG-760CT is intended for the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

    FUJIFILM Endoscope Models EC-760P-V/L and EC-740T/L are intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and large intestine.

    Device Description

    FUJIFILM Endoscope Models EG-760CT, EC-760P-V/L and EC-740T/L are comprised of three general sections: a control portion, an insertion portion and an umbilicus. The controls the angulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion portion contains glass fiber bundles, several channels and either a complementary metal-oxide-semiconductor (CMOS, in 760 series models) or charged-coupled device (CCD, in EC-740T/L) image sensor in its distal end. The channels in the insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CMOS or CCD image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when plugged into the video processor and the light source. The endoscopes are used in combination with FUJIFILM's video processors, light sources and peripheral devices such as monitor, printer, foot switch, and cart. All of these combinations were previously cleared in K172916.

    AI/ML Overview

    This document is a 510(k) summary for FUJIFILM Endoscope Models EG-760CT, EC-760P-V/L, and EC-740T/L. It does not describe a study proving the device meets acceptance criteria related to AI or algorithm performance. Instead, it focuses on demonstrating substantial equivalence to predicate devices through various non-clinical performance tests and adherence to regulatory standards.

    Therefore, many of the requested details about acceptance criteria for AI algorithms, sample sizes for test sets, expert involvement, and MRMC studies are not applicable to the information provided.

    However, I can extract the information related to the performance data presented to support substantial equivalence.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table with specific quantitative acceptance criteria or reported device performance for clinical outcomes. Instead, it lists various types of engineering and safety tests and states that the device "met performance specifications" or "are substantially equivalent in performance."

    Category of TestingAcceptance Criteria (Implied)Reported Device Performance
    Electrical SafetyCompliance with ANSI/AAM ES 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-6:2013, and IEC 60601-2-18:2009.Evaluated and met these standards.
    BiocompatibilityCompliance with ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010, and FDA guidance on ISO 10993-1.Evaluated and successfully conducted in accordance with standards and guidance. Differences in materials from predicate were validated.
    Endoscope Specific TestingCompliance with ISO 8600-1:2015, ISO 8600-3:1997, and ISO 8600-4:2014.Evaluated and met these standards.
    Software Specific TestingCompliance with ANSI/AAM//EC 62304:2006 and FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.Software validation activities performed in accordance with standards and guidance.
    Cleaning, High-level Disinfection, and SterilizationCompliance with AAMI TIR12:2010, AAMI TIR30:2011, and FDA guidance "Reprocessing Medical Devices in Health Care Settings."Validation of cleaning, disinfection, and sterilization instructions according to standards and guidance.
    Bench Testing (Features)Validation of "Advanced Force Transmission" and "Adaptive Bending" features.Conducted and features were validated.
    Bench Testing (Comparison)Substantial equivalence in the "rate of suction" compared to the predicate device.Comparative bench testing conducted, showing substantial equivalence in the rate of suction.
    Additional Performance TestingMeeting performance specifications for: Field of view, Bending capability, Rate of air supply, Rate of water supply, Working length, Diameter of forceps channel, Viewing direction, Resolution, LG output.Device met performance specifications for all listed items.

    2. Sample size used for the test set and the data provenance

    Not applicable. The document describes engineering and bench testing, not clinical studies with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This document pertains to device engineering and safety testing, not clinical evaluation requiring expert ground truth for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically for clinical study outcomes and expert consensus, which are not described here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document does not mention any MRMC studies or AI assistance. The devices are endoscopes, and the evaluation focuses on their physical and functional characteristics.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No. The devices are physical endoscopes, not standalone algorithms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the bench and engineering tests, the "ground truth" was established by technical specifications, consensus standards (e.g., ISO, AAMI, IEC, ANSI/AAMI), and established testing protocols. For comparative tests, the predicate device's performance served as a reference for substantial equivalence.

    8. The sample size for the training set

    Not applicable. There is no mention of a training set as this is not an AI-driven device being evaluated for a machine learning model.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set mentioned, this question is not relevant.