510(k) Data Aggregation

    K Number
    K183549

    Validate with FDA (Live)

    Date Cleared
    2019-02-16

    (58 days)

    Product Code
    Regulation Number
    862.1665
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GEM Premier ChemSTAT is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of Sodium (Na+), Ionized Calcium (Ca++) and Chloride (C1-) from arterial and venous heparinized whole blood. These parameters. along with derived parameters, aid in the diagnosis of a patient's electrolyte balance.

    Electrolytes in the human body have multiple roles. Nearly all metabolic processes depend on or vary with electrolytes:

    • · Sodium (Na+) measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.
    • · Potassium (K+) measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.
    • Ionized calcium (Ca++) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
    • · Chloride (Cl-) measurements are used in the diagnosis and metabolic and metabolic disorders, such as, cystic fibrosis and diabetic acidosis.
    Device Description

    The GEM Premier ChemSTAT is a portable system that analyzes arterial and venous lithium heparinized whole blood at the point of health care delivery in a clinical setting and in a central laboratory for Na*, K*, Ca** and Cl . All tests are included in a single self-contained, disposable GEM Premier ChemSTAT PAK (cartridge).

    Key Components:
    Analyzer: The GEM Premier ChemSTAT analyzer has the internal logic and processing power necessary to perform analysis. It employs a unique touch-sensitive color screen and a simple set of menus and buttons for user interaction. The analyzer guides operators through the sampling process with simple, clear messages and prompts.
    PAK (Cartridge): The disposable, multi-use GEM Premier ChemSTAT PAK is a completely closed cartridge that houses all components necessary to operate the instrument once the GEM PAK is validated. These components include the sensors, Process Control (PC) Solutions, sampler, and waste bag. The values of all PC Solutions are read from the GEM PAK Electronically Erasable Programmable Read Only Memory (EEPROM) chip. The components and processes used to manufacture the PC Solutions in the GEM PAK are traceable to National Institute of Standards and Technology (NIST) standards, Clinical & Laboratory Standards Institute (CLSI) procedures or other internal standards, where available and appropriate. The GEM Premier ChemSTAT PAK has flexible menus to assist facilities in maximizing efficiency. As part of this program, GEM ChemSTAT CVP (Calibration Valuation Products) are external solutions intended to complete the calibration process and final accuracy assessment of the iQM cartridge calibration following warm-up.
    Intelligent Quality Management (iQM): Intelligent Quality Management (iQM) is used as the quality control and assessment system for the GEM Premier ChemSTAT system. iQM is an active quality process control program designed to provide continuous monitoring of the analytical process before and after sample measurement with real-time, automatic error detection, automatic correction and automatic documentation of all corrective actions. iQM performs 4 types of continuous, quality checks to monitor the performance of the GEM PAK, sensors, and reagents throughout the cartridge use-life. These checks include System, Sensor, Pattern Recognition (PR) and Stability Checks.

    AI/ML Overview

    The provided FDA 510(k) summary for the GEM Premier ChemSTAT describes the device's analytical performance, which is a type of acceptance criteria study. This document is not for an AI/ML device, but rather for an in-vitro diagnostic device that measures specific analytes. Therefore, many of the requested elements pertaining to AI/MRMC studies, expert ground truth, and training sets are not applicable or not provided in this type of submission. However, I can extract information related to the device's analytical performance and the studies conducted to demonstrate its performance against predefined criteria.

    Here's a breakdown of the acceptance criteria and performance study details, focusing on what is available in the provided document:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the analytical performance specifications and the comparison to the predicate device. The performance studies aim to demonstrate that the device meets these standards.

    Table 1: Acceptance Criteria (Implicit from Study Outcomes) and Reported Device Performance for GEM Premier ChemSTAT

    Performance MetricAnalyteAcceptance Criteria (Stated Goal / Implied Target)Reported Device Performance
    Precision (Internal)Na+All results within specification (implied acceptable SD/CV based on clinical utility and state of the art)Ranges from 0.4% to 0.6% Total %CV (Na+); 0.2% to 2.3% Total %CV (K+); 0.8% to 2.8% Total %CV (Ca++); 0.9% to 1.5% Total %CV (Cl-)
    K+See above
    Ca++See above
    Cl-See above
    Reproducibility (Aqueous Controls - POC)Na+All results at all sites within specification (implied acceptable Reproducibility SD/CV)Reproducibility %CV ranges from 0.5% to 0.8% (Na+); 0.0% to 1.2% (K+); 0.7% to 1.7% (Ca++); 0.8% to 3.3% (Cl-)
    K+See above
    Ca++See above
    Cl-See above
    External Precision (Whole Blood - POC)Na+All results at all sites within specification; Measured data partitioned into Fixed Acceptance Range (Constant SD) or Variable Acceptance Range (Constant %CV).Pooled "Within Sample SD of %CV" of 1.1% (Na+); 0.08% (K+); 0.006 (Ca++); 0.6% (Cl-)
    K+See above
    Ca++See above
    Cl-See above
    LoB, LoD, LoQAll AnalytesEstablished values for each analyteNa+: LoB 69, LoD 70, LoQ 88; K+: LoB 0.0, LoD 0.1, LoQ 0.3; Ca++: LoB 0.00, LoD 0.01, LoQ 0.05; Cl-: LoB 4, LoD 4, LoQ 36
    LinearityAll AnalytesR2 close to 1.0, slope close to 1.0, intercept close to 0; Claimed reportable range supported.Na+: Slope 1.023, Intercept -1.189, R2 0.9997; K+: Slope 0.995, Intercept 0.057, R2 0.9998; Ca++: Slope 0.986, Intercept 0.019, R2 0.9984; Cl-: Slope 1.011, Intercept -1.909, R2 0.9998. Claimed ranges are 100-180 mmol/L (Na+), 0.3-19.0 mmol/L (K+), 0.10-4.25 mmol/L (Ca++), 40-158 mmol/L (Cl-).
    Analytical Specificity (Interference)All AnalytesNo significant interference with listed substances at specific concentrations (implied bias < acceptable threshold) and identification of concentrations for observed interference.Many substances showed no interference. Specific interferences identified for Hemoglobin (Potassium), Perchlorate (Chloride), and Triglyceride (Potassium) with quantified bias at lowest impact concentrations.
    Clinical Testing (Method Comparison)All AnalytesSubstantially equivalent performance to predicate device (GEM Premier 4000) for R value, slope, and intercept.Na+: R 0.987, Slope 1.021, Intercept -2.157; K+: R 0.999, Slope 1.000, Intercept 0.100; Ca++: R 0.999, Slope 1.000, Intercept 0.015; Cl-: R 0.994, Slope 1.000, Intercept 1.000.

    Details of the Study

    1. Sample sizes used for the test set and the data provenance:

      • Internal Precision Study – Whole Blood: 120 replicates per level for each of 5 levels (N=600 total samples analyzed for each analyte). Data provenance is internal (Instrumentation Laboratory Co.). The study simulates an internal lab environment.
      • Reproducibility Study with Aqueous Controls – POC Setting: 90 pooled replicates per level for each of 7 levels (N=630 samples analyzed for each analyte). The study was performed at three (3) external clinical point-of-care (POC) sites.
      • External Precision – Whole Blood: Variable N per site. For the Na+, K+, Ca++, Cl- main ranges, N=~63-69 samples per site for 3 sites, pooled N=195. For specific lower/higher ranges, N=3 samples per site for 3 sites, pooled N=9. The study was performed at three (3) external clinical point-of-care (POC) sites. Less than 10% of samples were "contrived" (meaning artificially prepared or modified) and the rest were patient samples from the intended use population. Data provenance is implied to be mixed: patient samples are real-world, while contrived samples are laboratory-sourced. The locations are referred to as "external clinical point-of-care (POC) sites", implying real-world clinical settings, likely within the US given FDA submission. It is a prospective study as samples were "run" during the study.
      • LoB, LoD, LoQ: No specific N mentioned, but involved testing with three (3) lots of GEM Premier ChemSTAT PAKs.
      • Linearity: 18 replicates per level (9 levels, total of 162 samples) for each analyte. Whole blood samples, prepared by spiking or diluting, were used.
      • Analytical Specificity (Interference): Not explicitly stated, but various substances were screened at different concentrations. Implied to be laboratory-based evaluation.
      • Clinical Testing (Method Comparison):
        • Na+: N=436 patient samples.
        • K+: N=442 patient samples.
        • Ca++: N=444 patient samples.
        • Cl-: N=435 patient samples.
        • Data provenance: Patient samples from the intended use population, collected at three (3) point-of-care (POC) sites. Less than 10% of samples were contrived. Implied to be prospective as a method comparison study involving running samples.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • For this type of in-vitro diagnostic device, "ground truth" for clinical samples is typically established by comparative measurement against a predicate device or a reference method, rather than by expert consensus (as would be the case for image interpretation, for example).
      • The "ground truth" in the Clinical Testing (Method Comparison) was established by comparing the GEM Premier ChemSTAT's measurements to those of the GEM Premier 4000 (K133407), which is the legally marketed predicate device. Operator qualifications involved "health care professionals" for point-of-care use, and "internal precision studies" would be performed by qualified lab personnel. Specific expert qualifications are not detailed beyond "health care professionals."
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable/Not mentioned. This is a quantitative measurement device, not an interpretation task requiring adjudication. The performance is assessed statistically (SD, CV, R, slope, intercept).
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/ML device or an imaging device requiring human reader interpretation. It is an IVD device providing quantitative measurements.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This is an in-vitro diagnostic instrument. Its primary function is "standalone" measurement of analytes. The performance studies (precision, linearity, interference, method comparison) assess the instrument's inherent analytical capabilities. There is no "human-in-the-loop performance" in the sense of an AI assisting human interpretation; direct measurement is the output.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For LoB, LoD, LoQ, Linearity, Analytical Specificity, and Internal Precision: Ground truth is against known concentrations in prepared samples (standards, calibrators, controls, spiked/diluted samples). This is a metrological approach.
      • For Clinical Testing (Method Comparison) and External Precision (Whole Blood): Ground truth is established by the measurements from the predicate device (GEM Premier 4000) for patient samples. The assumption is that the predicate device provides clinically acceptable and accurate measurements for comparison.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device, so there is no "training set" in the machine learning sense. The device is based on established electrochemical principles (potentiometry).
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no "training set" for an AI/ML model. The device's calibration and internal quality control use known standards and calibrated reference materials.