510(k) Data Aggregation

    K Number
    K183323

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-08-05

    (248 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Used to temporarily occlude the ureteropelvic junction to prevents from entering the ureter during percutaneous lithotripsy, and for injection of contrast medium.

    Device Description

    The UPJ Occlusion Balloon Catheter is composed of a balloon catheter, a removable inflation/injection adapter, a 1 mL syringe, and a wire guide assembly. The UPJ Occlusion Balloon Catheter is manufactured from a double lumen polyurethane radiopaque tubing with a working length of 75 centimeters. The smaller lumen is designed for balloon inflation, while the larger lumen fits over the wire guide assembly. The distal balloon is manufactured from silicone and has an approximate inflation volume of 4 mL. The inflation/injector adapter is removable, allowing for a cystoscope to be removed after the catheter is in place. The wire guide assembly is manufactured in 0.028- or 0.038-inch diameter stainless steel coils. The stainless-steel coils are coated with polytetrafluoroethylene and measure 80 centimeters. The wire guide assembly is designed to be placed through the small lumen of the balloon catheter. The UPJ Occlusion Balloon Catheter is supplied sterile, intended for one-time use only, and packaged in a peel-open pouch with a shelf-life of 3 years.

    AI/ML Overview

    This document describes the Cook Incorporated UPJ Occlusion Balloon Catheter (K183323). This is a physical medical device, not an AI/ML powered device. As such, the information you requested about acceptance criteria and studies is not applicable in the context of an AI/ML device.

    Here's a breakdown based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not provide a table with specific acceptance criteria and reported device performance in the manner typically seen for AI/ML models (e.g., sensitivity, specificity, AUC). Instead, it lists performance data in the form of engineering and material testing.

    Acceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityMet requirements
    Tensile StrengthMet requirements
    Kink RadiusMet requirements
    Leakage and Lumen BlockageMet requirements
    RadiopacityMet requirements
    Balloon Fatigue and BurstMet requirements
    Dimensional Measurement and CompatibilityMet requirements

    2. Sample size used for the test set and the data provenance:

    This information is not explicitly provided. The "testing" mentioned is for a physical device, and typically involves a certain number of manufactured units or material samples. The document does not specify quantities or data provenance (e.g., country of origin, retrospective/prospective) because these concepts are not applicable to the engineering tests described.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. For physical devices, "ground truth" is established through engineering and material standards, not expert consensus on medical images or clinical data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable as the testing involves physical device properties, not diagnostic interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable as the device is a physical catheter, not an AI/ML diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable as the device is a physical catheter, not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this physical device is based on established engineering and material standards, and compliance with the device's design specifications and intended use. This is demonstrated through various physical and material performance tests.

    8. The sample size for the training set:

    This information is not applicable as the device is a physical catheter and does not involve an AI/ML training set.

    9. How the ground truth for the training set was established:

    This information is not applicable as the device is a physical catheter and does not involve an AI/ML training set.