510(k) Data Aggregation

    K Number
    K183175

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-01-04

    (49 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BostonSight Scleral daily wear contact lenses are indicated for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic and non aphakic persons. Also, the lenses may be prescribed in eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, keratoglobus, post-LASIK ectasia, or following penetrating keratoplasty or refractive surgery.

    Furthermore, eves suffering from certain ocular surface disorders may benefit from the physical protection, aqueous hydrated environment and the saline bath provided by scleral lens designs. BostonSight Scleral daily wear contact lenses are indicated for therapeutic use for the management of irregular and distorted corneal surfaces where the subject:

    1. cannot be adequately corrected with spectacle lenses
    2. requires a rigid gas permeable contact lens surface to improve vision
    3. is unable to wear a corneal rigid gas permeable lens due to corneal distortion or surface irregularities

    Common causes of corneal distortion include, but are not limited to, corneal infections, trauma, tractions as a result of scar formation secondary to refractive surgery (e.g. LASIK or radial keratotomy) or corneal transplantation. Causes may also include corneal degeneration (e.g. keratoconus, keratoclobus, pellucid marginal degeneration, Salzmann's nodular degeneration) and corneal dystrophy (e.g., lattice dystrophy, granular corneal dystrophy, Reis-Bucklers dystrophy, Cogan's dystrophy).

    The BostonSight Scleral daily wear contact lenses are also indicated for therapeutic use in eyes with ocular surface disease including, but not limited to, ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome and Filamentary Keratitis, limbal stem cell deficiency (e.g. Stevens-Johnson syndrome, chemical radiation and thermal burns), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for a distorted cornea or ocular surface disease, the BostonSight Scleral daily wear contact lenses may concurrently provide correction of refractive error.

    The lenses may be disinfected using a chemical disinfection (not heat) system only.

    Device Description

    The BostonSight Scleral daily wear contact lenses are manufactured with a large diameter RGP lens design that vaults over the cornea and rests on the conjunctiva overlying the sclera. The BostonSight Scleral daily wear contact lenses are lathe cut and fabricated from one of the following fluoro-silicone acrylate rigid gas permeable (RGP) lens materials:

    • roflufocon D supplied by Contamac Ltd.
    • roflufocon E supplied by Contamac Ltd.
    • oprifocon A supplied by Bausch and Lomb, Inc.
    • hexafocon B supplied by Bausch and Lomb, Inc.

    The BostonSight Scleral daily wear contact lenses may be shipped "dry" or "wet" in a polypropylene contact lens case. The primary container for shipping the BostonSight Scleral daily wear contact lenses is the Bausch & Lomb Frequent Replacement Contact Lens Case, with clearance under 510(k) K896685.

    When shipped "wet", The BostonSight Scleral daily wear contact lenses manufactured from material supplied by Bausch & Lomb, inc. may be packaged and shipped in Boston Advance Comfort Formula Conditioning Solution (K974466) or Boston SIMPLUS Multi-Action solution (K024289). The BostonSight Scleral daily wear contact lenses manufactured from material supplied by Contamac, Ltd. may be packaged and shipped "wet" in in the OPTIMUM by Lobob Cleaning and Disinfecting Storage solution, with clearance under 510(k) K014162.

    AI/ML Overview

    The provided document is a 510(k) summary for the BostonSight Scleral daily wear contact lenses. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving the device meets specific acceptance criteria through a clinical study with a defined test set, ground truth, and expert adjudication in the manner typically seen for novel AI/ML-based medical devices.

    Therefore, the information required to populate a table of acceptance criteria and the details of a study proving the device meets these criteria (including sample sizes, expert qualifications, and adjudication methods) are not present in this document. The document explicitly states that the purpose of the application is "to modify the labeling of previously FDA cleared RGP contact lenses/materials to include therapeutic indications for use." It further clarifies that non-clinical and clinical testing to demonstrate safety and effectiveness for the materials used "has been addressed in previous applications."

    Here's a breakdown of why the requested information cannot be extracted from this document, and what is provided instead:

    1. Table of Acceptance Criteria and Reported Device Performance:

    • Not present. The document focuses on demonstrating substantial equivalence based on technological characteristics and intended use compared to predicate devices, not on meeting predefined performance criteria with quantitative results.

    2. Sample size used for the test set and the data provenance:

    • Not present. There is no mention of a dedicated "test set" or a new clinical study conducted for this specific 510(k) submission. The document states that clinical performance data "has been previously addressed" in prior applications for the materials used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not present. Since no new dedicated test set or clinical study is detailed, there's no mention of experts involved in establishing ground truth for such a study.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable/Not present. No test set or corresponding adjudication process is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This document describes a physical medical device (contact lenses), not an AI/ML-based device that would involve human readers or AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm or software device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not present. No new ground truth establishment is described for this submission. The equivalence relies on the established safety and performance of the materials and predicate devices.

    8. The sample size for the training set:

    • Not applicable/Not present. Contact lenses are physical devices, not AI models that require training sets.

    9. How the ground truth for the training set was established:

    • Not applicable/Not present. As above, no training set for an AI model is involved.

    What the document does provide:

    • Device Name: BostonSight Scleral
    • Regulation Number: 21 CFR 886.5916 (Rigid Gas Permeable Contact Lens)
    • Regulatory Class: Class II
    • Indications for Use: (Detailed on pages 2 and 5-6)
      • Correction of refractive ametropia (myopia, hyperopia, astigmatism, presbyopia) in aphakic and non-aphakic persons.
      • Management of irregular corneal conditions (e.g., keratoconus, pellucid marginal degeneration, post-LASIK ectasia, post-keratoplasty/refractive surgery).
      • Therapeutic use for ocular surface disorders (e.g., GvHD, Sjögren's, dry eye, Filamentary Keratitis, limbal stem cell deficiency, disorders of the skin, neurotrophic keratitis, corneal exposure) benefiting from an expanded tear reservoir and protection.
    • Predicate Devices:
      • "Boston XO® and Boston XO2® Rigid Gas Permeable Contact Lens" (K171404)
      • "OPTIMUM GP Daily Wear Contact Lens" (K180616)
      • "BostonSight PD Prosthetic Device" (K161461)
    • Device Materials: roflufocon D, roflufocon E, oprifocon A, hexafocon B.
    • Physical Properties: Refractive Index, Light Transmission, Specific Gravity, Oxygen Permeability (Dk), UV Light Blocking, Dynamic Receding Contact Angle (all listed with values for each material on page 5).
    • Lens Parameters: Base Curve, Center Thickness, Diameter, Spherical Power (with ranges and tolerances on page 5).
    • Basis for Clearance: Substantial equivalence to predicate devices based on intended use, indications for use, actions, classification, FDA material group, USAN materials, production method, and final packaging/shipping. The safety and effectiveness of the materials themselves were addressed in prior applications.