Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K183021

    Validate with FDA (Live)

    Device Name
    ClearEndoclip
    Date Cleared
    2019-06-28

    (239 days)

    Product Code
    Regulation Number
    876.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ClearEndoclip is intended to be used with FDA-cleared endoscopic clip placement within the gastrointestinal tract. It is indicated to be used for

    • (1) Endoscopic marking
    • (2) Hemostasis for
    • (a) Mucosal/sub-mucosal defects < 3cm
    • (b) Bleeding ulcers
    • (c)Arteries < 2 mm
    • (d)Polyps < 1.5 cm in diameter
    • (e)Diverticula in the colon
    • (3) Anchoring to affix jejunal feeding tubes to the wall of the small bowel
    • (4) As a supplementary method, closure of GI tract luminal perforations <20 mm that can be treated conservatively
    Device Description

    ClearEndoclip is a sterile device consisting of a pre-loaded, single-use, endoscopic clipping device with two main components: the (clip-fixing) delivery system and the clip. Clip is pre-loaded in the clip fixing delivery system, connected with an operation wire. Clip will open when the slider of the clip fixing device is pushed, and closed when it is pulled towards the operator. The clip could be closed and re-opened up to five times. When the slider is pulled further, the clip closes completely. The clip will be released when the slider is pushed. The delivery system consists of a handle and delivery catheter. The delivery system is constructed using stainless steel. HDPE outer sheath, and polypropylene stopper materials. The delivery system will allow for the device to rotate at the distal end. The ClearEndoclip is offered in 165cm and 230cm working lengths.

    ClearEndoclip consists of two main components, first, the endoscopic clip, which gets physically deployed and placed as a hemostatic purpose clip in the patient's gastrointestinal tract and second, the delivery system, known as a clip fixing device used to deliver the endoscopic clip under the use of an endoscope.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study information for the ClearEndoclip device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria in a table format, nor does it provide specific quantitative performance metrics like sensitivity, specificity, accuracy, or a pass/fail percentage for a diagnostic or AI-driven device.

    Instead, the performance data section focuses on demonstrating conformance to established standards and comparing the device's characteristics to a predicate device. The general acceptance criterion seems to be "demonstrates equivalent safety and performance profile as the predicate device" and "conformance to the appropriate recognized standards of FDA, ISO 10993-1 (2009)."

    Here's a summary of the performance claims:

    Acceptance Criterion (Implied)Reported Device Performance
    BiocompatibilityFavorable biocompatibility test results drawn in 2018 testing, ensuring FDA's latest consensus standards (ISO 10993-1: 2009) are met. Classified as "Implant Device - Tissue/bone/dentin (Prolonged contact duration)" for the clip and "Externally Communicating Devices (<24 hrs of limited contact duration)" for the delivery system, matching predicate.
    Design Verification & Validation (General Safety & Performance)ClearEndoclip meets pre-defined safety and performance requirements. Design input matched with design outputs. Risk management (ISO 14971:2012) conducted with effective risk control measures.
    Shelf Life PerformanceThree-year accelerated aging performance testing completed to affirm performance profile over indicated shelf life.
    Time-Zero Performance Comparison to PredicateFavorable test results for: - Appearance - Dimension - Rotation - Repeated open/close - Clamping retention time - Tensile strength
    Physicochemical SafetyFavorable test results of physicochemical testing of extractable substance of Delivery System, characterizing reliable physicochemical profile and assurance of safety.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes design verification and validation testing and biocompatibility testing, but these are not clinical studies in the typical sense with patient test sets.

    • Test Set Sample Size: No mention of a "test set" in terms of subject or image count for an AI or diagnostic device. The testing described focuses on device components and their physical/chemical properties.
    • Data Provenance: Not applicable in the context of device component and materials testing. The testing was conducted in 2018 for biocompatibility.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. The described testing involves laboratory and engineering evaluations, not clinical adjudication by experts.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set or expert adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. The document describes a device (ClearEndoclip) rather than an AI system. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant or mentioned. This is a medical device clearance, not an AI software clearance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No. This document pertains to a physical medical device (endoscopic clip) and not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the tests performed appears to be:

    • Biocompatibility: Conformance to ISO 10993-1: 2009 standards and demonstrating a safety profile equivalent to the predicate device.
    • Design Verification & Validation: Meeting pre-defined safety and performance requirements, matching design inputs with outputs, and effectiveness of risk mitigation measures (ISO 14971:2012).
    • Performance Tests (Appearance, Dimension, Rotation, etc.): Presumably established engineering specifications and comparison to the performance of the predicate device.
    • Physicochemical Testing: Established analytical methods and safety profiles for extractable substances.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there's no training set mentioned for an AI/ML device, this question is not relevant.

    In summary: The provided text describes the 510(k) clearance for a physical medical device, the ClearEndoclip, which is an endoscopic clipping device, not an AI or software device. The "study" described consists of engineering and biocompatibility testing to demonstrate substantial equivalence to a predicate device and compliance with recognized standards. It does not involve clinical trials with human subjects or AI performance metrics.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1