Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182833

    Validate with FDA (Live)

    Date Cleared
    2019-02-08

    (122 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ArgenZ Liquids are intended to be used by trained dental technicians as an accessory for shading ArgenZ allzirconia, monolithic restorations for anterior and posterior dental prosthetics.

    Device Description

    ArqenZ Zirconia Coloring Liguids are dilute aqueous solutions of transition and lanthanoid metal salts used in combination with ArgenZ Zirconia material (K071410 and K150919) to properly shade teeth. Elements, which may include (but are not limited to) Co, Cr, Er, Fe, Mn, Pr, Ni, Bi, and Nd are combined with water to result in the finished product.

    AI/ML Overview

    The provided text describes Argen Corporation's ArgenZ Liquids, which are dental coloring liquids for zirconia restorations. The document is a 510(k) premarket notification for the device, seeking substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Product Standard BS EN ISO 6872:2015)CriteriaReported Performance (ArgenZ Liquids)
    Flexural Strength
    ArgenZ HT+>= 800 MPa1230 MPa
    ArgenZ HT+ Shaded with ArgenZ HT Liquid (A4)>= 800 MPa1223 MPa
    ArgenZ Anterior>= 500 MPa868 MPa
    ArgenZ Anterior Shaded with ArgenZ ST Liquid (A3)>= 500 MPa765 MPa
    ArgenZ Anterior Shaded with ArgenZ ST Liquid (C4)>= 500 MPa963 MPa
    Chemical Solubility
    ArgenZ HT+< 100 ug/cm²4.16 ug/cm²
    ArgenZ HT+ Shaded with ArgenZ HT Liquid (A4)< 100 ug/cm²4.16 ug/cm²
    ArgenZ Anterior< 100 ug/cm²20.80 ug/cm²
    ArgenZ Anterior Shaded with ArgenZ ST Liquid (A3)< 100 ug/cm²10.32 ug/cm²
    ArgenZ Anterior Shaded with ArgenZ ST Liquid (C4)< 100 ug/cm²8.30 ug/cm²
    Cytotoxicity (ISO 10993-5)Reactivity grade <= 2 (mild reactivity)All test articles showed no evidence of cell lysis or toxicity (i.e. grade <= 2).

    2. Sample Size for Test Set and Data Provenance

    • Mechanical & Chemical Properties: The document does not specify the exact sample size for each test (flexural strength and chemical solubility). It mentions "test records of parent materials and representative accessories infused into the parent materials," and lists several variations (e.g., ArgenZ HT+, ArgenZ HT+ Shaded with ArgenZ HT Liquid (A4)). The provenance is not explicitly stated, but these are laboratory tests conducted by the manufacturer.
    • Cytotoxicity: The document states that "test articles for the cytotoxicity testing were selected as representative samples for each product family, and determined to be worst case scenarios for potential biocompatibility issues." Four specific liquid/zirconia combinations were selected as representative samples (ST on K150919 (A3), ST on K150919 (C4), HT on K071410 (A4), HT on K071410 (B4)). The provenance is not explicitly stated, but these are laboratory tests conducted by the manufacturer.

    3. Number of Experts and Qualifications

    This document describes a pre-market notification for a dental material, not a diagnostic device involving human interpretation. Therefore, there are no "experts" in the context of establishing ground truth for diagnostic accuracy, nor qualifications for such experts. The evaluation relies on standardized material testing.

    4. Adjudication Method

    Not applicable, as this is related to material properties testing, not human interpretation of diagnostic results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This device is a dental coloring liquid used by trained dental technicians, not a diagnostic tool where human readers assess cases. The study focuses on material properties, not reader performance.

    6. Standalone (Algorithm Only) Performance

    Not applicable. This device is a material, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the performance evaluation is based on product standards (BS EN ISO 6872:2015 for mechanical and chemical properties, and ISO 10993-5 for cytotoxicity). These standards define the acceptable limits for various material properties.

    8. Sample Size for Training Set

    Not applicable. This is a material, not a machine learning model, so there is no training set in the conventional sense. The "training" in manufacturing refers to established processes and quality control.

    9. How Ground Truth for Training Set was Established

    Not applicable, as there is no training set for an AI/algorithm. For materials, the "ground truth" for manufacturing involves adherence to established specifications, quality control procedures, and compliance with relevant industry standards and regulations.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1