Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182825

    Validate with FDA (Live)

    Date Cleared
    2019-02-22

    (141 days)

    Product Code
    Regulation Number
    874.4680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EB-530US is intended for the observation, diagnosis, and endoscopic treatment of the trachea, bronchial tree and surrounding organs using ultrasonic images. It is used with a FUJIFILM ultrasonic processor, video processor, light source, other peripheral equipment and endoscopic accessories. It is not intended for use on children and infants.

    Device Description

    FUJIFILM Ultrasonic Endoscope EB-530US is an ultrasonic bronchoscope that emits ultrasound waves and scans the reflected signals to provide ultrasonic images when used in combination with an ultrasonic processor.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/ParameterReported Device PerformanceStudy Type
    Electrical SafetyANSI/AAMI ES 60601-1:2012Met pre-defined criteriaNon-clinical
    IEC 60601-1-2:2007Met pre-defined criteriaNon-clinical
    IEC 60601-1-6:2013Met pre-defined criteriaNon-clinical
    IEC 60601-2-18:2009Met pre-defined criteriaNon-clinical
    IEC 60601-2-37:2015Met pre-defined criteriaNon-clinical
    BiocompatibilityISO 10993-1:2009Met pre-defined criteriaNon-clinical
    ISO 10993-5:2009Met pre-defined criteriaNon-clinical
    ISO 10993-10:2010Met pre-defined criteriaNon-clinical
    Storage & TransportationExpanded temperature range (-20°C to 60°C)ValidatedNon-clinical
    Expanded humidity range (10 to 85% RH)ValidatedNon-clinical
    Performance TestingField of viewMet pre-defined criteriaNon-clinical
    Forceps channel diameterMet pre-defined criteriaNon-clinical
    Axial resolutionMet pre-defined criteriaNon-clinical
    Bending capabilityMet pre-defined criteriaNon-clinical
    Viewing directionMet pre-defined criteriaNon-clinical
    Lateral resolutionMet pre-defined criteriaNon-clinical
    Suction rateMet pre-defined criteriaNon-clinical
    ResolutionMet pre-defined criteriaNon-clinical
    Penetration depthMet pre-defined criteriaNon-clinical
    Working lengthMet pre-defined criteriaNon-clinical
    LG outputMet pre-defined criteriaNon-clinical

    2. Sample size used for the test set and the data provenance

    The document does not specify human clinical trials or a "test set" in the context of patient data. The performance testing appears to be entirely non-clinical (engineering and laboratory tests), likely using a sample of the manufactured device itself or components. Therefore, information about human data provenance is not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The reported performance testing is non-clinical and does not involve expert evaluation for ground truth in a clinical context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set requiring adjudication by experts is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done, and the device is an ultrasonic endoscope, not an AI-assisted diagnostic tool. Therefore, the effect size of human readers improving with AI assistance is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not an AI algorithm. The device is a medical instrument. Therefore, this question is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance tests, the "ground truth" would be established by objective measurements against engineering specifications, calibrated instruments, and established safety and performance standards (e.g., IEC standards for electrical safety, ISO standards for biocompatibility).

    8. The sample size for the training set

    Not applicable, as this is not an AI/machine learning device requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1