    Device Name
    E-CUBE i7
    Date Cleared
    2019-01-24

    (126 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal (renal & GYN/pelvic); Pediatric, Small Organ (breast, testes, thyroid); Adult Cephalic; Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional); Musculo-skeletal (Superficial); Cardiac (adult& pediatric); Peripheral Vascular (PV); and Urology (including prostate).

    Device Description

    E-CUBE i7 product is an ultrasound imaging system for medical diagnosis. This device is available for portable(only body) and mobile(with system cart). Also, this innovative system platform provides optimal patient diagnosis workflow with the 15.6" wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.

    1. Signal Mode:
      B(2D) mode, M mode, Color Flow(CF) mode, Power Doppler(PD) mode, Pulsed Wave Doppler (PWD) mode, Continuous wave Doppler (CWD) mode, Tissue Doppler Imaging(TDI) mode, 3D, 4D

    B/M, B/CF, B/PD, B/PWD, B/CF/PWD, B/CF/M

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the ALPINION MEDICAL SYSTEMS Co., Ltd. E-CUBE i7 ultrasound system. This submission primarily focuses on establishing substantial equivalence to previously cleared predicate devices rather than presenting a novel device with specific acceptance criteria that require proof through a new study.

    Therefore, there is no specific study mentioned in this document that proves the device meets particular acceptance criteria in the format requested. The document asserts substantial equivalence based on comparisons to predicate devices and adherence to established safety and performance standards.

    Here's an attempt to structure the information based on your request, highlighting what is (and isn't) present in the provided text:

    1. A table of acceptance criteria and the reported device performance

    Since this is a 510(k) submission for substantial equivalence, the "acceptance criteria" are implicitly defined by the safety and performance characteristics of the predicate devices and relevant industry standards. The "reported device performance" is essentially the statement that the E-CUBE i7 is equivalent in performance to these predicates.

    Acceptance Criteria (Implied by Predicate Equivalence & Standards)Reported Device Performance (E-CUBE i7)
    Indications for Use:Identical to primary predicate (K153424)
    Fetal, Abdominal, Pediatric, Small Organ, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional & Superficial), Cardiac (Adult & Pediatric), Peripheral Vessel, Urology (including prostate)Same
    Electrical Power:Voltage: 19V, 10.5A; Frequency: 50/60Hz; Power: 200W Max
    Imaging Modes:2D (B) mode, M mode, Anatomical M, Color Flow Doppler (CF) mode, Power Doppler (PD) mode, Directional PD, Pulsed wave Doppler (PWD) mode, Continuous wave Doppler (CWD) mode, Tissue Doppler imaging (TDI) mode, 3D/4D mode
    Image Processing Technology:Xpeed™, Full SRI™, Spatial Compounding Image (SCI), ECG mode
    Software Features:Panoramic, Needle Vision™/Needle Vision™ Plus, Cube View™, Cube Strain™, Stress Echo
    Thermal, Mechanical, Electrical Safety:Conforms to NEMA UD2, AIUM Medical Ultrasound Safety, IEC60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC61000-3-2, IEC 61000-3-3, IEC 61000-4-2, IEC 61000-4-3, IEC61000-4-4, IEC61000-4-5, IEC61000-4-6, IEC61000-4-8, IEC61000-4-11 (identical to primary predicate)
    Biocompatibility:Transducer materials and other patient contact materials are biocompatible. (Statement from document)

    The document states, "Therefore, the differences between E-CUBE i7 and the predicate devices would not affect the safety, effectiveness and essential performance." This is the core "proof" in a 510(k) context.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document explicitly states: "The subject of this premarket submission, E-CUBE i7, did not require clinical studies to support substantial equivalence."
    Therefore, there is no test set, sample size, or data provenance information of this nature provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical studies with test sets were conducted.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical studies with test sets were conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The E-CUBE i7 is an ultrasound imaging system, and the submission does not mention any AI assistance features or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable, as the device is a diagnostic ultrasound system and not an algorithm-only device. Performance assessment is based on demonstrating equivalence to predicate ultrasound systems through engineering and performance testing against standards, not through standalone algorithm performance studies.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Given that no clinical studies were performed, there's no clinical "ground truth" established in this submission. The "ground truth" for demonstrating substantial equivalence relies on:

    • Engineering principles and design specifications.
    • Performance testing against recognized standards (e.g., NEMA UD2, AIUM Medical Ultrasound Safety, IEC 60601 series).
    • Comparison of features and specifications to legally marketed predicate devices.

    8. The sample size for the training set

    Not applicable, as no machine learning/AI models are discussed that would require a training set in the context of this submission. The device is an ultrasound imaging system.

    9. How the ground truth for the training set was established

    Not applicable.