    K Number
    K182516

    Validate with FDA (Live)

    Date Cleared
    2019-04-22

    (221 days)

    Product Code
    Regulation Number
    888.3660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Primary total shoulder arthroplasty.
    2. Non-inflammatory degenerative joint disease including osteoarthritis

    Comprehensive Nano Stemless Shoulder humeral components have a porous coated surface coating and are indicated for uncemented biological fixation applications.

    The Comprehensive Modular Hybrid Glenoid is intended to be implanted with bone cement. The porous titanium peg may be inserted without bone cement.

    Device Description

    The Comprehensive Nano Stemless Shoulder is part of the Comprehensive Shoulder System intended for use in total shoulder arthroplasty. The humeral component is used as part of an anatomic configuration with a Versa-Dial® humeral head that articulates against the Comprehensive hybrid glenoid component. The head is connected to the Nano humeral component by a taper adaptor.

    The Comprehensive Nano Stemless Shoulder is manufactured from Ti6Al4V alloy. It consists of a central tapered region and six outer wings. The taper has a machine finish and accepts the taper adaptor of the humeral head component. The bone-contacting outer surface features a porous coating of plasma sprayed titanium alloy for cementless fixation in the proximal humerus. Six sizes are available - 30 mm through 40mm in 2mm increments.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Comprehensive® Nano Stemless Shoulder, based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria Study EndpointAcceptance CriteriaReported Device Performance (Comprehensive Nano Stemless Shoulder)
    Endpoint a: ASES mean scores at two yearsNon-inferiority to the predicate Comprehensive mini stem (p < 0.05, 95% CI lower bound)Mean ASES score for nano group: 92.5 at two years. Mean ASES score for mini stem group: 92.2 at two years. Difference: 0.27. Non-inferiority demonstrated (p = < 0.0001, 95% CI lower bound = -2.81).
    Endpoint a: ASES MCID Improvement at 2 YearsNot explicitly stated as a separate acceptance criterion, but data provided for comparison. "Minimal Clinically Important Difference (MCID) of 20.9 points defined by Tashjian et al."2 subjects (1.79%) in the nano group did not meet MCID improvement. 3 subjects (2.48%) in the control (mini stem) group did not meet MCID improvement. (Fisher Exact Test P Value 1.000000 for "No").
    Endpoint b: Absence of unanticipated device-related adverse events, fracture, perforation, dissociation, or revision/removal of components.Non-inferiority to the predicate Comprehensive mini stem (p < 0.05, 95% CI lower bound) for the outcomes success rate. (The success rate is defined as % of subjects without these events).Outcomes success rate for nano group: 92.2% (107 of 116). Outcomes success rate for mini group: 92.7% (114 of 123). Difference: -0.0044. Non-inferiority demonstrated (p = 0.0063, 95% CI lower bound = -0.0674).
    Endpoint c: Radiographic Success (Humeral Component)1. Subsidence of the humeral component < 5mm.All subjects in both groups successfully passed the radiographic endpoint criteria. (Indicates all nano subjects met this criterion).
    2. Migration of the humeral component < 5mm.All subjects in both groups successfully passed the radiographic endpoint criteria. (Indicates all nano subjects met this criterion).
    3. No progressive lucency around the humeral component > 2mm in two or more contiguous zones.All subjects in both groups successfully passed the radiographic endpoint criteria. (Indicates all nano subjects met this criterion).
    Endpoint c: Radiographic Success (Glenoid Component)1. Migration of the glenoid component < 5mm.All subjects in both groups successfully passed the radiographic endpoint criteria. (Indicates all nano subjects met this criterion).
    2. No progressive lucency > 2mm around the entire glenoid component.All subjects in both groups successfully passed the radiographic endpoint criteria. (Indicates all nano subjects met this criterion).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • Comprehensive Nano Humeral Component (Investigational): 116 subjects (who reached the two-year follow-up time point or were removed from the study).
      • Comprehensive mini stem (Control): 123 subjects.
    • Data Provenance: Prospective, randomized, blinded, multicenter study. The country of origin is not specified in the provided text.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • The document does not specify the number of experts used to establish the ground truth for the test set.
    • It refers to evaluations by Principal Investigators (for adverse events) and implies clinical assessment for outcomes like ASES scores and radiographic success. Their specific qualifications (e.g., "radiologist with 10 years of experience") are not detailed, but they are implicitly qualified medical professionals conducting a clinical study.

    4. Adjudication Method for the Test Set

    • The document does not explicitly describe a formal adjudication method (e.g., 2+1, 3+1) for the outcomes.
    • The adverse events were classified by the principal investigator as definitely related, probably related, or possibly related to the device. This indicates an individual clinical assessment rather than a multi-expert adjudication panel for initial classification of device-related adverse events.
    • For the co-primary endpoints (ASES scores, adverse events/revisions, radiographic success), the collection and analysis appear to be based on study protocols and clinical assessments by the investigators.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done.
    • This was a clinical trial comparing two different medical devices (the Nano Stemless Shoulder vs. the mini stem shoulder prosthesis) in human subjects, not a study evaluating human readers' performance with or without AI assistance. The "readers" here are the patients themselves (their ASES scores and outcomes), and the clinicians assessing their progress and radiographs.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • This is not an AI/algorithm device. The device is a physical shoulder prosthesis (Comprehensive® Nano Stemless Shoulder).
    • Therefore, a standalone (algorithm only) performance study is not applicable. The study evaluated the clinical performance of the physical implant itself.

    7. The Type of Ground Truth Used

    • The "ground truths" in this study are derived from a combination of:
      • Clinical Outcomes Data: ASES (American Shoulder and Elbow Surgeons) scores, which are patient-reported and clinician-evaluated functional assessments.
      • Adverse Event Reporting: Documented unanticipated device-related adverse events, fractures, perforations, dislocations, and revisions/removals, classified by the principal investigator.
      • Radiographic Assessments: Measurable criteria from imaging studies (subsidence, migration, progressive lucency around components) to determine radiographic success.
    • This represents real-world clinical performance data from human patients.

    8. The Sample Size for the Training Set

    • Not applicable. As this is a physical medical device (shoulder prosthesis) and not an AI algorithm, there is no "training set" in the context of machine learning.
    • The study described is a clinical study to evaluate the device's performance in patients, not to train a model.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. See point 8. There is no training set for an AI algorithm.