510(k) Data Aggregation

    K Number
    K182264

    Validate with FDA (Live)

    Date Cleared
    2018-10-19

    (59 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONN Revolution Knotless Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications:

    Shoulder: Rotator cuff repair, Bankart repair, Biceps Tenodesis, Acromio-Clavicular separation, Deltoid repair, and Capsular shift or Capsulolabral reconstruction.

    Foot/Ankle: Lateral stabilization, Medial stabilization, Achilles tendon repair, Hallux valgus reconstruction, Mid-foot reconstruction, Metatarsal ligament repair.

    Device Description

    The ICONN Revolution Knotless Suture Anchor (P/N 350000) is a sterile (gamma irradiation), single use implantable device made from Solvay's Zeniva® ZA-500 polyetheretherketone (PEEK) material indicated for soft tissue reattachment procedures in the shoulder, foot, and ankle. The anchor is available in one size (Diameter x Length): 5mm x 23mm. The implants with attached disposable inserter are supplied sterile and the instrument is supplied Non-Sterile and is to be sterilized in the ICONN instrument tray prior to surgery. The disposable inserter is made out of 17-4 stainless steel and the reusable driver is made out of 17-4 stainless steel (shaft) and silicone (handle). All anchors are intended to be used with the supplied #2 sutures and/or 1.5mm ICONN TetherTape, both made from Ultra High Molecular Weight Polyethylene (UHMWPE). The sutures/tape are supplied sterile, are packaged individually, and are single strands measuring 36/38 inches in length.

    AI/ML Overview

    The provided text is a 510(k) summary for the ICONN Revolution Knotless Suture Anchor, a medical device. It describes the device, its intended use, and comparative studies against a predicate device to establish substantial equivalence for FDA clearance. The summary does not provide information about an AI/algorithm-based device as requested in the original prompt. Therefore, I cannot extract the specific details you've asked for regarding acceptance criteria and an AI study.

    However, I can provide the available information related to the device's performance testing and acceptance criteria as described in the document, which are related to its mechanical and biocompatibility properties, not AI performance.

    Here's a breakdown of the information that is available in the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document mentions performance testing but does not present a formal table of "acceptance criteria" with specific numerical targets alongside "reported device performance." Instead, it states the following regarding the outcomes of the performance tests:

    Test TypeAcceptance Criteria (Implicit)Reported Device Performance
    Pullout CharacteristicsCompliance with FDA's Draft Guidance Document for Testing Bone Anchor Devices (April 20, 1996)"performed similarly or better than the predicate or within acceptable ranges for the intended use."
    - Anchor displacement after cyclic loading(Implicitly, within limits acceptable for safe and effective use and comparable to predicate)(Met implicit criteria)
    - Ultimate anchor pull-out(Implicitly, within limits acceptable for safe and effective use and comparable to predicate)(Met implicit criteria)
    Insertion TestingCompliance with ASTM F543-2013 (specifically sections A1 and A2)"demonstrate the device can be safely implanted."
    - Torsional properties(Implicitly, within limits acceptable for safe implantation)(Met implicit criteria)
    - Driving torque(Implicitly, within limits acceptable for safe implantation)(Met implicit criteria)
    Biocompatibility TestingCompliance with ISO 10993/(R)2013"all testing met acceptance criteria."
    - Cytotoxicity(Met ISO 10993/(R)2013 acceptance criteria)(Met acceptance criteria)
    - Irritation(Met ISO 10993/(R)2013 acceptance criteria)(Met acceptance criteria)
    - Sensitization(Met ISO 10993/(R)2013 acceptance criteria)(Met acceptance criteria)
    LAL testing (Endotoxin)In accordance with ANSI/AAMI ST72:2006, result <0.05 EU/device"results demonstrated the test article were <0.05 EU/device."
    Pyrogenicity(Non-pyrogenic status)"device was determined to be non-pyrogenic."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for the mechanical or biological testing described. It also does not provide information about data provenance (e.g., country of origin, retrospective or prospective nature) for these tests. The tests performed are laboratory bench tests and material tests, not clinical studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable. The evaluations described are mechanical and biocompatibility tests conducted in a laboratory setting, not clinical studies requiring expert interpretation of medical data or ground truth establishment by medical professionals.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable, as the tests described are objective laboratory and material tests, not subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC study or any AI component in the provided text. The device is a physical Suture Anchor.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not applicable, as the device is a physical Suture Anchor and does not involve any algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This question is not applicable in the context of the physical device's performance and biocompatibility testing. "Ground truth" in this context refers to objective measurements and established standards (e.g., precise force measurements, chemical analysis, cell viability assays according to ISO standards).

    8. The sample size for the training set

    This question is not applicable, as the device does not involve a "training set" in the context of an AI or algorithm.

    9. How the ground truth for the training set was established

    This question is not applicable, as the device does not involve a "training set" in the context of an AI or algorithm.